Purpose: To investigate the reliability of intra-atrial electrocardiogram (ECG) use for external jugular vein (EJV) catheterization. Materials and Methods: Patients undergoing open heart surgery in Suleyman Demirel University Hospital between February and June 2006 were included in the study. Using a sterile Seldinger technique, a triple lumen polyurethane central venous catheter was introduced (Certofix® Trio V 720, length 20 cm, 7 French) under intra-atrial ECG guidance. The presence of an increase in P-wave size was recorded. Just after the surgery, a portable chest X-ray was taken. The method was considered to be successful when a change in P-wave could be seen and the catheter was in the superior vena cava, as well as when there was no change in P-wave and the catheter was not in the superior vena cava. Results: In six patients (12%), we were not able to advance the guidewire. In the remaining 44 patients, the catheter was inserted without problem. Eight of these 44 catheters were positioned in the innominate vein, with a malposition ratio of 18%. The success rate of external jugular vein cannulation with intra-atrial ECG was 95%. No complications occured related to the EJV cannulation. Conclusion: Considering that it is easily accessed without complication, and the malposition is successfully detected by intra-atrial ECG, EJV is a suitable access for central venous cannulation when internal jugular vein (IJV) is not usable.
Abstract
Purpose: To investigate the reliability of intra-atrial electrocardiogram (ECG) use for external jugular vein (EJV) catheterization. Materials and Methods: Patients undergoing open heart surgery in Suleyman Demirel University Hospital between February and June 2006 were included in the study. Using a sterile Seldinger technique, a triple lumen polyurethane central venous catheter was introduced (Certofix® Trio V 720, length 20 cm, 7 French) under intra-atrial ECG guidance. The presence of an increase in P-wave size was recorded. Just after the surgery, a portable chest X-ray was taken. The method was considered to be successful when a change in P-wave could be seen and the catheter was in the superior vena cava, as well as when there was no change in P-wave and the catheter was not in the superior vena cava. Results: In six patients (12%), we were not able to advance the guidewire. In the remaining 44 patients, the catheter was inserted without problem. Eight of these 44 catheters were positioned in the innominate vein, with a malposition ratio of 18%. The success rate of external jugular vein cannulation with intra-atrial ECG was 95%. No complications occured related to the EJV cannulation. Conclusion: Considering that it is easily accessed without complication, and the malposition is successfully detected by intra-atrial ECG, EJV is a suitable access for central venous cannulation when internal jugular vein (IJV) is not usable.
참고문헌 (26)
- Weeks SM, "Unconventional venous access", Tech Vasc Interv Radiol, 5, pp.114~120 (2002)
- Trerotola SO, "Tunneled infusion catheters: increased incidence of symptomatic venous thrombosis after subclavian versus internal jugular venous access", Radiology, 217, pp.89~93 (2000)
- Trerotola SO, "You are asked to place a dialysis access catheter in a patient. What is your preferred access site, and why?", J Vasc Interv Radiol, 8, pp.75~76 (1997)
- Macdonald S, "Comparison of technical success and outcome of tunneled catheters inserted via the jugular and subclavian approaches", J Vasc Interv Radiol, 11, pp.225~231 (2000)
- Schillinger F, "Post catheterisation vein stenosis in haemodialysis: comparative angiographic study of 50 subclavian and 50 internal jugular accesses", Nephrol Dial Transplant, 6, pp.722~724 (1991)
- Moss AH, "Use of a silicone catheter with a Dacron cuff for dialysis short-term vascular access", Am J Kidney Dis, 12, pp.492~498 (1988)
- Cho SK, "Use of the right external jugular vein as the preferred access site when the right internal jugular vein is not usable", J Vasc Interv Radiol, 17, pp.823~829 (2006)
- Nicolson SC, "Comparison of internal and external jugular cannulation of central circulation in the pediatric patient", Crit Care Med, 13, pp.747~749 (1985)
- Chu KS, "Accurate central venous port-A catheter placement: intravenous electrocardiography and surface landmark techniques compared by using transesophageal echocardiography", Anesth Analg, 98, pp.910~914 (2004)
- Watters VA, "Use of electrocardiogram to position right atrial catheters during surgery", Ann Surg, 225, pp.165~171 (1997)
- Schummer W, "Intra-atrial ECG is not a reliable method for positioning left internal jugular vein catheters", Br J Anaesth, 91, pp.481~486 (2003)
- Mickley V, "Central venous catheters: many questions, few answers", Nephrol Dial Transplant, 17, pp.1368~1373 (2002)
- McCowan TC, "Use of external jugular vein as a route for percutaneous inferior vena caval filter placement", Radiology, 176, pp.527~530 (1990)
- Stickle BR, "Prediction of a small internal jugular vein by external jugular vein diameter", Anaesthesia, 52, pp.220~222 (1997)
- Belani KG, "Percutaneous cervical central venous line placement: a comparison of the internal and external jugular vein routes", Anesth Analg, 59, pp.40~44 (1980)
- Jobes DR, "Safer jugular vein cannulation: recognition of arterial puncture and preferential use of the external jugular route", Anesthesiology, 59, pp.353~355 (1983)
- Malatinsky J, "Misplacement and loop formation of central venous catheters", Acta Anaesthesiol Scand, 20, pp.237~224 (1976)
- Salmela L, "Verification of the position of a central venous catheter by intra-atrial ECG. When does this method fail?", Acta Anaesthesiol Scand, 37, pp.26~28 (1993)
- "Illustrated encyclopedia of human anatomic variation"[URL]
- McGee WT, "Accurate placement of central venous catheters: a prospective, randomized, multicenter trial", Crit Care Med, 21, pp.1118~1123 (1993)
- Vesely TM, "Central venous catheter tip position: a continuing controversy", J Vasc Interv Radiol, 14, pp.527~534 (2003)
- Fletcher SJ, "Safe placement of central venous catheters: where should the tip of the catheter lie?", Br J Anaesth, 85, pp.188~190 (2000)
- Peres PW, "Positioning central venous catheters-a prospective survey", Anaesth Intensive Care, 18, pp.536~539 (1990)
- Aslamy Z, "MRI of central venous anatomy: implications for central venous catheter insertio", Chest, 114, pp.820~826 (1998)
- Hsu JH, "Comparison of radiographic landmarks and the echocardiographic SVC", RA junction in the positioning of long-term central venous catheters, 2006, pp.50, 731~735
- Schummer W, "Central venous catheters-the inability of ‘intra--atrial ECG’ to prove adequate positioning", Br J Anaesth, 93, pp.193~198 (2004)