We experienced a case of venous vessel wall entrapment between the introducer needle and the guide wire during an attempt to perform right internal jugular vein (IJV) catheterization. The guide wire was introduced with no resistance but could not be withdrawn. We performed ultrasonography and C-arm fluoroscopy to confirm the entrapment location. We assumed the introducer needle penetrated the posterior vessel wall during the puncture and that only the guide wire entered the vein; an attempt to retract the wire pinched the vein wall between the needle tip and the guide wire. Careful examination with various diagnostic tools to determine the exact cause of entrapment is crucial for reducing catastrophic complications and achieving better outcomes during catheterization procedures.

Abstract

We experienced a case of venous vessel wall entrapment between the introducer needle and the guide wire during an attempt to perform right internal jugular vein (IJV) catheterization. The guide wire was introduced with no resistance but could not be withdrawn. We performed ultrasonography and C-arm fluoroscopy to confirm the entrapment location. We assumed the introducer needle penetrated the posterior vessel wall during the puncture and that only the guide wire entered the vein; an attempt to retract the wire pinched the vein wall between the needle tip and the guide wire. Careful examination with various diagnostic tools to determine the exact cause of entrapment is crucial for reducing catastrophic complications and achieving better outcomes during catheterization procedures.

참고문헌 (6)

  1. Koç F, "Tricuspid valve dysfunction associated with entrapment of the guide wire in the tricuspid valve during central venous catheterization", Turk Kardiyol Dern Ars, 38, pp.115~117 (2010)
  2. Mastan M, "Internal jugular venous cannulation complicated by J-tip guide wire entrapment", Br J Anaesth, 86, pp.292~293 (2001)
  3. Reddy S, "Ultrasound diagnosis of central line guidewire entrapment with an inferior vena cava filter", ICU Dir, 2, pp.215~217 (2011)
  4. Kang SH, "J-tip guide wire entrapment within the heart during central venous catheterization", Hemodial Int, 16, pp.438~440 (2012)
  5. Hoda MQ, "Unusual site of guide-wire entrapment during central venous catheterization", J Pak Med Assoc, 56, pp.139~141 (2006)
  6. Marcus HE, "The impact of Trendelenburg position and positive end-expiratory pressure on the internal jugular cross-sectional area", Anesth Analg, 111, pp.432~436 (2010)

관련 논문AI추천

중심정맥도관 삽입 중 발생한 유도철사의 매듭과 꼬임 -증례 보고-고성훈, 김덕규, 김동찬, 손지선, 이지혜, 임형선대한중환자의학회 · Acute and Critical Care2011 · 제26권 제1호Direction of the J-Tip of the Guidewire to Decrease the Malposition Rate of an Internal Jugular Vein Catheter유인술, Byeong jun Ahn, Jin Woong Lee, Seung Ryu, Sung Uk Cho, Won Joon Jeong, Yeon Ho You, Yong chul Cho대한중환자의학회 · Acute and Critical Care2015 · 제30권 제4호Ultrasound-guided supraclavicular internal jugular vein compression to increase internal jugular vein cross-sectional area in hospitalized patients: a prospective observational study in JapanHiroyuki Hond, Kei Nishiyama, Masataka Hiruma, Mitsuhiro Watanabe, Shuichiro Kurita, Shunsuke Nukaga대한중환자의학회 · Acute and Critical Care2025 · 제40권 제4호중심정맥도관삽입술 시 위치이상의 빈도를 감소시키는 방법: 실시간 초음파 유도하 재위치유승, 김건동, 김승환, 안홍준, 유연호, 유인술, 이진웅, 정원준, 조용철, 조별님희대한중환자의학회 · Acute and Critical Care2013 · 제28권 제4호쇄골하정맥을 통하여 중심정맥도관 삽입 실패 후 나타난 혈흉 1예김대영, 김대우, 박상진, 손희원, 이덕희영남대학교 의과대학 · Journal of Yeungnam Medical Science2008 · 제25권 제2호A Rare Case of Massive Hemothorax due to Central Venous Catheterization Treated with Angiographic Stent Implantation배정민대한중환자의학회 · Acute and Critical Care2015 · 제30권 제1호