잡견에서 조영제 혈관외유출 예방을 위한 스트레인 게이지기반의 EDA 시스템 성능 평가를 위한 실험적 연구

An Experimental Study for Performance Evaluation in Dogs of Preventive Contrast Media Extravasation with a Strain Gage Based Prototype Extravasation Detection Accessory System

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

The major risk associated with the use of automated power injectors is the well known complication of contrast material extravasation at the injection site. Automated injection of computed tomography (CT) contrast media can produce the compartment syndrome. The purpose of this study was to assess the ability of this device during clinically important episodes of extravasation. The extravasation detection accessory (EDA) system was composed of a strain gage, an amplifier and a computer based system. A strain gage pliable adhesive patch was applied to the skin over the intravenous catheter and the catheter was connected to the power injector with a cable to monitor the resolution data. If the programmed monitoring, which was developed with MS Visual C++, at the extravasation occurred, then the injection was interrupted the auto injector. CT was used to demonstrate the clinically important extravasation. This study was a prospective, observational study in which the EDA system was used to monitor the automated mechanical injection of contrast material in 7 dogs. There were two true-positive cases (range of extravasation volumes: 18 ∼ 22 ml), twenty three true-negative cases, three false-positive cases and no false-negative cases. The EDA system had a sensitivity of 100% and a specificity of 88% for the detection of clinically important extravasation. The EDA system had good sensitivity for the detection of clinically important extravasation and the EDA system has the clinical potential for the early detection of extravasation of the contrast medium that is administered with power injectors. The EDA system is easy to use safe and accurate for the monitoring extravasation of the intravenous injections, and this system may prove especially useful in CT applications.

Abstract

The major risk associated with the use of automated power injectors is the well known complication of contrast material extravasation at the injection site. Automated injection of computed tomography (CT) contrast media can produce the compartment syndrome. The purpose of this study was to assess the ability of this device during clinically important episodes of extravasation. The extravasation detection accessory (EDA) system was composed of a strain gage, an amplifier and a computer based system. A strain gage pliable adhesive patch was applied to the skin over the intravenous catheter and the catheter was connected to the power injector with a cable to monitor the resolution data. If the programmed monitoring, which was developed with MS Visual C++, at the extravasation occurred, then the injection was interrupted the auto injector. CT was used to demonstrate the clinically important extravasation. This study was a prospective, observational study in which the EDA system was used to monitor the automated mechanical injection of contrast material in 7 dogs. There were two true-positive cases (range of extravasation volumes: 18 ∼ 22 ml), twenty three true-negative cases, three false-positive cases and no false-negative cases. The EDA system had a sensitivity of 100% and a specificity of 88% for the detection of clinically important extravasation. The EDA system had good sensitivity for the detection of clinically important extravasation and the EDA system has the clinical potential for the early detection of extravasation of the contrast medium that is administered with power injectors. The EDA system is easy to use safe and accurate for the monitoring extravasation of the intravenous injections, and this system may prove especially useful in CT applications.

참고문헌 (17)

  1. D. A. Stein, "Compartment syndrome of the hand caused by computed tomography contrast infiltration", Orthopedics, 26(3), pp.333~334 (2003)
  2. 권대철, "전산화단층검사에서 조영제의 피하 정맥 혈관외유출 환자의 3D영상", 의학물리, 16(1), pp.47~51 (2005), 한국의학물리학회
  3. R. C. Nelson, "Contrast media extravasation during dynamic CT: detection with an extravasation detection accessory", Radiology, 209, pp.837~843 (1998)
  4. J. G. Amaral, "Safety of power injector use in children as measured by incidence of extravasation", Am. J. Roentgenol, 187, pp.580~583 (2006)
  5. 권대철, "CT 검사에서 조영제의 혈관외유출 예방을 위한 EDA 시스템 개발의 예비 보고", 의학물리, 17(1), pp.32~39 (2006), 한국의학물리학회
  6. W. H. Bush, "Acute reaction to intravascular contrast media. types, risk factors, recognition, and specific treatment", Am. J. Roentgenol, 157, pp.1153~1161 (1991)
  7. R. K. Daniel, "The free transfer of skin flaps by microvascular anastomoses. An experimental study and a reappraisal", Plast. Reconstr. Surg, 52(1), pp.16~31 (1973)
  8. S. H. Kim, "Experimental tissue damage after subcutaneous injection of water-soluble contrast media", Invest. Radiol, 25(2), pp.171~183 (1989)
  9. S. L. Sistrom, "Extravsastion of iopamidol and iohexhol during contrast-enhancement CT: report of 28 cases", Radiology, 180, pp.707~710 (1991)
  10. M. P. Federle, "Frequency and effects of extravasation of ionic and nonionic CT contrast media during rapid bolus injection", Radiology, 637, pp.637~640 (1998)
  11. J. E. Jacobs, "Contrast media reactions and extravasation: relationship to intravenous injection rates", Radiology, 209, pp.11~416 (1998)
  12. S. T. Cochran, "Trends in adverse events after IV administration of contrast media", Am. J. Roentgenol, 176(6), pp.1385~1388 (2001)
  13. S. G. Miles, "Safe use of an intraveous power injector for CT: experience and protocol", Radiology, 190, pp.69~70 (1990)
  14. E. Vandeweyer, "Extravasation injuries and emergency suction as treatment", Plast. Reconstr. Surg, 105, pp.109~110 (2000)
  15. B. A. Birnbaum, "Extravasation detection accessory: clinical evaluation in 500 patients", Radiology, (212), pp.431~438 (1999)
  16. M. F. Bellin, "Contrast Media Safety Committee Of The European Society Of Urogenital Radiology", Eur. Radiol, 12(11), pp.2807~2812 (2002)
  17. 권대철, "가토 정맥에서 CT 조영제의 혈관외유출 예방을 위한 EDA 시스템의 실험적 연구", 의학물리, 17(4), pp.238~245 (2006), 한국의학물리학회

관련 논문AI추천

이 간행물의 다른 논문 (11건)