혈압측정시 가압 단계에서 목표압력 및 측정 종료압력 추정
Estimation of Target and Completion Pressure during the Cuff Inflation Phase in Blood Pressure Measurement
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
In blood pressure measurement, the oscillometric method detects and analyzes the pulse pressure oscillation while deflating the cuff around the arm. For its principle, one has to inflate cuff pressure above the subject’s systolic pressure and deflate below the diastolic pressure. Most of the commercialized devices inflate until the fixed target pressure and deflate until the fixed completion pressure because there is no way to know the systolic and diastolic pressure before measurement. Too high target pressure makes stress to the subject and too low target pressure makes big error or long measurement time because of re-inflation. There are similar problems for inadequate completion pressure. In this study, we suggest new algorithm to set proper target and completion pressure for each subject by analyzing pressure waveform while inflating period. We compared our proposed method and auscultation method to see the errors of estimation. The differences between the two measurements were -4.02 ± 4.80mmHg, -10.50 ± 10.57mmHg and -0.78 ± 5.17mmHg for mean arterial pressure, systolic pressure and diastolic pressure respectively. Consequently, we could set the target pressure by 30 mmHg higher than our estimation and we could stop at 20mmHg lower than our estimated diastolic pressure. Using this method, we could reduce the measurement time.
Abstract
In blood pressure measurement, the oscillometric method detects and analyzes the pulse pressure oscillation while deflating the cuff around the arm. For its principle, one has to inflate cuff pressure above the subject’s systolic pressure and deflate below the diastolic pressure. Most of the commercialized devices inflate until the fixed target pressure and deflate until the fixed completion pressure because there is no way to know the systolic and diastolic pressure before measurement. Too high target pressure makes stress to the subject and too low target pressure makes big error or long measurement time because of re-inflation. There are similar problems for inadequate completion pressure. In this study, we suggest new algorithm to set proper target and completion pressure for each subject by analyzing pressure waveform while inflating period. We compared our proposed method and auscultation method to see the errors of estimation. The differences between the two measurements were -4.02 ± 4.80mmHg, -10.50 ± 10.57mmHg and -0.78 ± 5.17mmHg for mean arterial pressure, systolic pressure and diastolic pressure respectively. Consequently, we could set the target pressure by 30 mmHg higher than our estimation and we could stop at 20mmHg lower than our estimated diastolic pressure. Using this method, we could reduce the measurement time.
참고문헌 (12)
- Geddes, L.A., "Indirect mean blood pressure on the anesthetized pony", Am. J. Vet. Res, 28, pp.2055~2057 (1977), 2055-2057
- Ramsey, M, "Noninvasive automatic determination of mean arterial pressure", Med. Biol. Eng. Comput, 17, pp.17~18 (1979)
- Geddes, L.A, "The efficient detection of Korotkoff sounds", Med. Biol. Eng, 6161, pp.603~609 (1968)
- 노동곤, "가압식 오실로메트릭 방법을 사용한 혈압측정 시스템의 설계", 의공학회지, 24(4), pp.281~286 (2003), 대한의용생체공학회
- National High Blood Pressure Education Program, "The Seventh Report of the Joint National Committe on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure", ~
- N. Morioka, "Measurement of Non-Invasive Oscillometric Blood PRessure During Cuff Inflation, Rather than Deflation", Anesthesiology, 87, pp.A431~ (1997), pp.A431
- L.A.Geddes, "Characterization of the oscillometric method for measuring indirect blood pressure", Ann Biomed Eng, 10, pp.271~280 (1892)
- Alan M. Gross, "Confidence Intervals for Bisquare Regression Estimates", J. American statistical Association, 72(358), pp.341~354 (1977)
- Chemla D, "A new formula for estimating mean aortic pressure", Lancet, 353, pp.1069~1070 (1999)
- Eoin O’Brien, "validation and Reliability of Blood Pressure Monitors", Clinical Hypertension and Vascular Disease, Blood Pressure Monitoring in Cardiovascular Medicine and Therapeutics 2nd Edition
- Jiri Jilek, "Oscillometric Blood Pressure Measurement : The Methodology, Some Observations, and Suggestions", Biomedical Instrumentation & Technology, 39, pp.237~241 (2005)
- Razminia M, "Validation of a new formula for mean arterial pressure calculation: The new formula is superior to the standard formula", Catheter Cardiovasc Interv, 63(4), pp.419~425 (2004)