자동혈압계 성능평가를 위한 인체혈압 시뮬레이터 개발

A Simulator for the Validation of Non-invasive Blood Pressure(NIBP) Monitoring Devices

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

Blood pressure is one of the important vital signs for monitoring the medical condition of a patient. Automated NIBP(non-invasive blood pressure) monitoring devices calculate systolic and diastolic blood pressures from the oscillation in cuff pressure caused by a pulsation of an artery. To validate the NIBP devices, we developed a simulator to supply the oscillometric waveforms obtained from human subjects. The simulator provided pressure pulses to device-under-test and device readings were compared to the auscultatory references. Fully automated simulation system including OCR(optical character recognition) were developed and used for NIBP monitoring devices. The validation results using the simulator agreed well with previous clinical validation. More validation studies using the standardized oscillometric waveforms would be required for the replacement of clinical trials to validate a new automated NIBP monitoring device.

Abstract

Blood pressure is one of the important vital signs for monitoring the medical condition of a patient. Automated NIBP(non-invasive blood pressure) monitoring devices calculate systolic and diastolic blood pressures from the oscillation in cuff pressure caused by a pulsation of an artery. To validate the NIBP devices, we developed a simulator to supply the oscillometric waveforms obtained from human subjects. The simulator provided pressure pulses to device-under-test and device readings were compared to the auscultatory references. Fully automated simulation system including OCR(optical character recognition) were developed and used for NIBP monitoring devices. The validation results using the simulator agreed well with previous clinical validation. More validation studies using the standardized oscillometric waveforms would be required for the replacement of clinical trials to validate a new automated NIBP monitoring device.

참고문헌 (11)

  1. M. Turner, "Sphygmomanometer calibration-why, how and how often?", Aust. Fam. Physician, 36(10), pp.834~837 (2007)
  2. I. Doh, "Calibration of oscillometric non-invasive devices for monitoring blood pressure", Metrologia, 52, pp.291~296 (2015)
  3. C. R. Jones, "Validation of the Welch Allyn ‘Vital Signs’ oscillometric blood pressure monitor.", J. Hum. Hypertens, 15(3), pp.191~195 (2001)
  4. E. O’Brien, "European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults.", Blood Press. Monit, 15(1), pp.23~38 (2010)
  5. R. Braam, "Accuracy of the Welch Allyn Vital Signs Monitor 52000 automatic blood pressure measuring device according to a modified British Hypertension Society", Blood Press. Monit, 7(3), pp.185~189 (2002)
  6. I. Doh, "Development of a simulator for the validation of noninvasive blood pressure-monitoring devices", Blood Press. Monit, 21(3), pp.189~191 (2016)
  7. J. Lee, "Digital recording system of sphygmomanometry", Blood Press. Monit, 14(2), pp.77~81 (2009)
  8. S. Song, "Does the accuracy of blood pressure measurement correlate with hearing loss of the observer?", Blood Press Monit, 19(1), pp.14~18 (2014)
  9. Alpert Bruce S, "Validation of the Welch Allyn ProBP 3400 : a device for modern medical practice", Blood Press. Monit, 16(3), pp.156~158 (2011)
  10. Z. D. Kobalava, "Validation of TM-2655 oscillometric device for blood pressure measurement.", Blood Press. Monit, 11(2), pp.87~90 (2006)
  11. J. Belghazi, "Validation of four automatic devices for selfmeasurement of blood pressure according to the International Protocol of the European Society of Hypertension", Vasc. Health Risk Manag, 3(4), pp.389~400 (2007)