오실로메트릭 혈압 측정에서 커패시턴스 센서와 적응필터를 이용한 새로운 잡음제거방법에 관한 연구

A New Method for Artifact Reduction Based on Capacitive Sensor and Adaptive Filter in Oscillometric Blood Pressure Measurement

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In this study, a new method using a capacitive sensor and an adaptive filter was proposed to deal with artifacts contaminating an oscillation signal in oscilometric blood pressure measurement. The proposed method makes use of a variation of the capacitance between an electrode fixed to a cuff and an external object to detect artifacts caused by the external object bumping into the cuff. The proposed method utilizes the adaptive filter based on linear prediction to remove the detected artifacts. The conventional method using linear interpolation and the proposed method using the adaptive filter were applied to three types of the artifact-contaminated oscillation signals(no overlap, non-consecutive overlap, and consecutive overlap between artifacts and oscillations) to compare them in terms of the artifact reduction performance. The proposed method was more robust than the conventional method in the case of consecutive overlap between artifacts and oscillations. The proposed method could be useful for measuring blood pressure in such a noisy environment that the subject is being transported.

Abstract

In this study, a new method using a capacitive sensor and an adaptive filter was proposed to deal with artifacts contaminating an oscillation signal in oscilometric blood pressure measurement. The proposed method makes use of a variation of the capacitance between an electrode fixed to a cuff and an external object to detect artifacts caused by the external object bumping into the cuff. The proposed method utilizes the adaptive filter based on linear prediction to remove the detected artifacts. The conventional method using linear interpolation and the proposed method using the adaptive filter were applied to three types of the artifact-contaminated oscillation signals(no overlap, non-consecutive overlap, and consecutive overlap between artifacts and oscillations) to compare them in terms of the artifact reduction performance. The proposed method was more robust than the conventional method in the case of consecutive overlap between artifacts and oscillations. The proposed method could be useful for measuring blood pressure in such a noisy environment that the subject is being transported.

참고문헌 (8)

  1. 노동곤, "가압식 오실로메트릭 방법을 사용한 혈압측정 시스템의 설계", 의공학회지, 24(4), pp.281~286 (2003), 대한의용생체공학회
  2. E.K.Park, "Reduction of interference in oscillometric arterial blood pressure measurement using fuzzy logic", J. Biomed. Eng. Res, 25(6), pp.605~609 (2004)
  3. C. T. Lin, "Reduction of interference in oscillometric arterial blood pressure measurement using fuzzy logic", IEEE Trans. Biomed. Eng, 50(4), pp.432~441 (2003)
  4. L.T.Hersh, "Method for oscillometric blood pressure determination employing curve fitting", U.S.Patent 5704362, ~, pp.Patent (1998)
  5. M.Ramsey III, "Automatic mean blood pressure reading device", U.S.Patent 4360029, ~ (1982)
  6. R.Medero, "Detection of blood pressure complexes in automated vital signs monitors", U.S.Patent 4546775, ~ (1985)
  7. C.H.Nelson, "Method and apparatus for indirect blood pressure measurement", U.S.Patent 4799492, ~ (1989)
  8. J. G. Webster, "Encyclopedia of Medical Devices and Instrum- entation - 2nd edition", John Wiley & Sons, 1, pp.488~ (1988)

관련 논문AI추천

이 간행물의 다른 논문 (10건)