자동혈압계 점검을 위한 액추에이터 기반의 혈압 시뮬레이터 개발

Development of An Actuator-Based Blood Pressure Simulator for Automatic Blood Pressure Monitor

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Accurate measurement of blood pressure is essential for classifying an individual's disease, identifying blood pressure-related risks, and managing health. Due to the environmental and health hazards of mercury sphygmomanom- eters, automatic sphygmomanometers using the oscillometric method are widely used in hospitals as well as in gener- al homes, and have established themselves as a practical standard sphygmomanometer. In this study, we developed a blood pressure simulator using an actuator that provides simulated pressure to an automatic blood pressure cuff. The developed blood pressure simulator adopts an arm-shaped cylindrical shape similar to the situation in which a person measures blood pressure with an automatic blood pressure monitor, and implements a method of transmitting pressure to the cuff using a pressure plate. Accuracy was evaluated through the mean and standard deviation of the difference with the commercialized blood pressure simulator BP PUMP 2, and reproducibility was confirmed using two automat- ic blood pressure monitors. The developed blood pressure simulator enables automatic blood pressure monitoring in a simple manner and also meets the evaluation standards for accuracy and reproducibility. In the future, as a standardized blood pressure simulator, it is expected to be of great help in evaluating and verifying the performance of automat- ic blood pressure monitors by supplementing precise hardware and software and building a blood pressure database.

Abstract

Accurate measurement of blood pressure is essential for classifying an individual's disease, identifying blood pressure-related risks, and managing health. Due to the environmental and health hazards of mercury sphygmomanom- eters, automatic sphygmomanometers using the oscillometric method are widely used in hospitals as well as in gener- al homes, and have established themselves as a practical standard sphygmomanometer. In this study, we developed a blood pressure simulator using an actuator that provides simulated pressure to an automatic blood pressure cuff. The developed blood pressure simulator adopts an arm-shaped cylindrical shape similar to the situation in which a person measures blood pressure with an automatic blood pressure monitor, and implements a method of transmitting pressure to the cuff using a pressure plate. Accuracy was evaluated through the mean and standard deviation of the difference with the commercialized blood pressure simulator BP PUMP 2, and reproducibility was confirmed using two automat- ic blood pressure monitors. The developed blood pressure simulator enables automatic blood pressure monitoring in a simple manner and also meets the evaluation standards for accuracy and reproducibility. In the future, as a standardized blood pressure simulator, it is expected to be of great help in evaluating and verifying the performance of automat- ic blood pressure monitors by supplementing precise hardware and software and building a blood pressure database.

참고문헌 (17)

  1. Asada H, "Towards the development of wearable blood pressure sensors : a photo-plethysmograph approach using conducting polymer actuators", Conf Proc IEEE Eng Med Biol Soc, 2005, pp.4156~4159 (2005)
  2. Drzewiecki GM, "The Korotkoffsound", Ann Biomed Eng, 17(4), pp.325~359 (1989)
  3. Roh DK, "Design of the blood pressure measurement system using the inflatable oscillometric method", J Biomed Eng Res, 24(4), pp.281~286 (2003)
  4. Moraes JCTB, "A strategy for determination of systolic, mean and diastolic blood pressures from oscillometric pulse profiles", IEEE Comput Cardiol, 27, pp.211~214 (2000)
  5. Perloff D, "Human blood pressure determination by sphygmomanometry", Circulation, 88, pp.2460~2470 (1993)
  6. Geddes LA, "Characterization of the oscillometric methods for measuring indirect blood pressure", Ann Biomed Eng, 10(6), pp.271~280 (1982)
  7. Information Compilation on Safety and Performance Testing Methods for Automated Electronic Blood Pressure Monitors, "Lee DH", National Institute of Food and Drug Safety Evaluation, 17~24 (2020)
  8. Cho EJ, "A Study on development of a guideline for performance evaluation of automatic sphygmomanometer", Regulatory Research on Food, Drug and Cosmetic, 9(1), pp.33~43 (2014)
  9. Williams B, "2018 ESC", ESH guidelines for the management of arterial hypertension, 2018, pp.39(33), 3021~3104
  10. Kim JE, "A Study on MZ generation(2030s)male body shape comparison and body shape change", J Korea Fashion Costume Design Assoc, 25(1), pp.33~47 (2023)
  11. Kim JE, "A study on changes in body shape of MZ generation(2030s)women for clothing construction", J Korea Fashion Costume Design Assoc, 24(3), pp.111~125 (2022)
  12. Kim SY, "A study of relation between symptoms and low blood pressure", J Yeongnam Med Sci, 11(1), pp.94~100 (1994)
  13. Lee HY, "New definition for hypertension", J Korea Med Assoc, 61(8), pp.485~492 (2018)
  14. Park DH, "The prognostic value of resting heart rate for health status", Exerc Sci, 29(1), pp.24~33 (2020)
  15. Kobalava ZD, "Validation of TM-2655 oscillometric device for blood pressure measurement", Blood Press Monit, 11(2), pp.87~90 (2006)
  16. Kim WK, "The development of a cuff for the accuracy enhancement of sphygmomanometer", J Korean Soc Precis Eng, 22(5), pp.181~188 (2005)
  17. Doh I, "A simulator for the validation of non-invasive blood pressure(NIBP)monitoring devices", J Biomed Eng Res, 38(3), pp.111~115 (2017)

관련 논문AI추천

이 간행물의 다른 논문 (6건)