A Study of Design of Hollow Fiber Membrane Modules for using in Artificial Lung by the PZT actuator
0
조회
0
다운로드
1
인용
0
참고문헌
초록
Abstract
The purpose of this work was to assess and quantify the beneficial effects of gas exchange, while testingto the various frequencies of the sinusoidal wave that was excited by the PZT actuator, for patients suffering from acute respiratory distress syndrome (ARDS) or chronic respiratory problems. Also, this paper considered a simulator to design a hollow type artificial lung, and a mathematical model was used to predict a behavior of blood. This simulation was carried out according to the Montecarno's simulation method, anda fourth order Runge-Kutta method was used to solve the equation. The experimental design and procedure are then applied to the construction of a new device to assess the effectiveness of the membrane vibrations. As a result, the vibration method is very effective in the increase of gas transport. The gas exchange efficiency for the vibrating intravascular lung assist device can be increased by emphasizing the following design features: consistent and reproducible fiber geometry, and most importantly, an active means of enhancing convective mixing of water around the hollow fiber membranes. The experimental results showed the effective performance of the vibrating intravascular lung assist device. Also, we concluded that important design parameters were blood flow rates, fiber outer diameter and oxygen pressure drop. Based on the present results, it was believed that the optimal level of blood flow rates was 200cm3/min.
참고문헌 (25)
- "Development of a low flowresistance intravenous oxygenator", 43, pp.725~730 (1997)
- "Principles of Pulmonary Medicine. Philadelphia" (1992)
- "Treatment of acute respiratory failure with low-frequency positive-pressure ventilation and extracorporeal removal of CO2", 2, pp.292~294 (1980)
- "Extracorporeal circulation in adult respiratory failure", 34, pp.43~47 (1988)
- "Venovenous perfusion of adult and children with severe acute respiratory distress syndrome", 34, pp.1014~1020 (1988)
- "Changing criteria for the artificial lung:Historic controls on the technology of ECMO", 40, pp.109~120 (1994)
- "Current status of extracorporeal membrane oxygenation for severe respiratory failure", 20, pp.120~123 (1996)
- "A mathematical model of gas exchange in an intravenous membrane oxygenator", 26, pp.166~178 (1998)
- "Intravascular oxygenator:A new alternative method for augmenting blood gas transfer in patient with acute respiratory failure", 16, pp.75~82 (1992)
- "Major findings from the clinical trials of the intravascular oxygenator", 18, pp.846~863 (1994)
- "Management of pneumonia", 86, pp.459~469 (1992)
- "Significant enhancement of carbon dioxide removal by a new prototype IVOX", 39, pp.719~724 (1993)
- "Use of a mathematical model to predict oxygen transfer rates in hollow fiber membrane oxygenators", 40, pp.990~996 (1994)
- "Gas transfer to blood flowing in coiled circular tubes", 94, pp.857~872 (1968)
- "Tightly wound coils of microporous tubing:progress with secondary flow blood oxygenator design", 21, pp.216~222 (1975)
- "Gas Transfer and Hemolysis Characteristic of a New Type Intravenous Lung Assist Devices", 24(2), pp.121~126 (2003)
- "Design of an Intravenous Oxygenator", 9(1), pp.34~41 (2006)
- "Bedford, Fluid Mechanics (9th ed.)" (1998), McGrew-Hill Inc.
- "Enhancement of Oxygen Transfer in Hollow Fiber Membrane by the Vibration Method", 22(4), pp.521~527 (2005)
- "Basic Principles of Membrane Technology", pp.418~422 (1996), Kluwer Academic Publishers
- "Clinical trails of an intravenous oxygenator in patients with adult respiratory distress syndrome", 77, pp.855~863 (1972)
- "Transport Phenomena", (mena), pp.262~267 (1999), John Wiley & Sons Inc.
- "Implantable artificial lung", 74, pp.130~136 (1977)
- "The Need for Standardizing the Index of Hemolysis", 18, pp.7~10 (1994)
- "Recommended practice for assessment of hemolysis in continuous flow blood pump west conshohoken", pp.40~41 (1998)