Design Criteria to Miniaturize the Single Use Functional Respiratory Air Flow Tube
Design Criteria to Miniaturize the Single Use Functional Respiratory Air Flow Tube
0
조회
0
다운로드
1
인용
0
참고문헌
초록
Abstract
Respiratory tubes with a length of 35mm and diameters of 10, 15, and 20mm were made for experimental purpose, and both the static(PS) and dynamic(PD) pressures were simultaneously measured for steady flow rates ranging 1-12l/sec. Least squares analysis resulted successful fitting of PS and PD data with quadratic equations with correlation coefficients higher than 0.99(P<0.0001). The spirometric measurement standards of the American Thoracic Society(ATS) were applied to PS data, which demonstrated the smallest tube diameter of 15mm to satisfy the ATS standards. The maximum PD value of the velocity type transducer(the functional single use respiratory air flow tube) with the diameter of 15mm was estimated to be approximately 75cmH2O, implying more than 7 times larger sensitivity than the widely used pneumotachometers. These results showed that the velocity type respiratory air flow transducer is a unique device accomplishing miniaturization with the sensitivity increased, thus would be of great advantage to develop portable medical devices.
참고문헌 (12)
- "E. J. Cha. Measurements of the respiratory system Translated by the BME education consortium", pp.509~598 (1993)
- "MicroLab Operating Manual", pp.2~4 (1998)
- "Non-uniform velocity sampling technique for respiratory air flow measurement", pp.77~ (2001)
- "Functional disposable use flow tube converting the respiratory air flow rate into averaged dynamic pressure", 11(3), pp.125~131 (2002)
- "General principles", pp.331~337 (2001)
- "U. S. Patent No. 5", pp.487~ (1993)
- "Essentials of engineering fluid mechanics. 4th. ed.", pp.452~482 (1980), Maple Press Company Ohio University
- "Standardization of spirometry", 152, pp.1107~1136 (1995)
- "U. S. Patent No. 5", pp.655~ (1991)
- "Flow resistance of air flow transducer" (1999)
- "Expiratory flow limitation", 55, pp.1~8 (1983)
- "Unequl distance sampling technique to design velocity-type respiratory air flow transducer", 25(5), pp.351~359 (2004)