압력 측정에 의한 요량 계측의 유용성 분석
Validation of Urine Volume Evaluation by Hydraulic Pressure Measurement
2
조회
0
다운로드
2
인용
0
참고문헌
초록
Abstract
Uroflowmetry is non-invasive and easily performed to diagnose benign prostatic hypertrophy(BPH) frequent in aged men. Weight change during urination is usually measured to estimate the urinary flow rate by a load cell, but sensitive to any impacts against the bottom of the container, leading to unnecessary noise generation. Moreover, load cells are relatively expensive raising the production cost. The present study proposed a new technique, measuring hydraulic pressure on the bottom of the urine container to evaluate the urinary volume. Low cost pressure transducer enabled almost perfectly linear relationship between the urine volume and the hydraulic pressure. During both the simulated and human urination experiments, variance of the pressure signal was more than 50% smaller than the weight signal acquired by a load cell, which demonstrated that the impact noise was decreased to a great degree by pressure compared to weight measurement.
참고문헌 (6)
- Y. G. Lee, "Molecular Biology of Prostate", pp.16~27 (2004), Ilchokak Press
- C. H. Noh, "Medical Examination and Treatment Guidlines in American Urological Association", pp.119~126 (2004), Ilchokak Press
- Y. K. Na, "Epidemiology and Natural Progress of Benign Prostatic Hyperplasia", pp.321~327 (2003), Ilchokak Press
- M. W. Kim, "Highly Recommended Test and Important test", pp.127~146 (2004), Ilchokak Press
- CAS KOREA, "Loadcell" (2006)
- D. U. Jeong, "Estimation and implementation of the uroflowmetry using load cell", 13(6), pp.436~445 (2004)