Improvement of Convection by Pulsatile Blood Pump in High Flux Hemodialysis System
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
High-flux dialysis treatment removes various toxins via diffusion as well as convection, which is induced by ultrafiltration and backfiltration. In this study, in vitro (Using the distilled water and the bovine’s blood) comparison test was performed to determine whether utilization of a high flux dialyzer paired with different pumps would increase the efficiency of convection. At the same blood flow rates, a pulsatile pump and a roller pump were employed to propel the distilled water and bovine whole blood to a high flux dialyzer. Pressures at the dialyzer inlet and outlet in the blood circuit and in the dialysate circuit were measured, respectively. From these data, we calculated the transmembrane pressure and predicted the ultrafiltration and backfiltration rates developed by both pumps. Using the bovine’s blood experiment, ultrafiltration and backfiltration rates were 1.6 times higher with the pulsatile pump than with the roller pump. We conclude that utilization of a pulsatile pump in high flux hemodialysis treatments increases ultrafiltration volume, compared with a roller pump under conditions of the same blood flow rate.
Abstract
High-flux dialysis treatment removes various toxins via diffusion as well as convection, which is induced by ultrafiltration and backfiltration. In this study, in vitro (Using the distilled water and the bovine’s blood) comparison test was performed to determine whether utilization of a high flux dialyzer paired with different pumps would increase the efficiency of convection. At the same blood flow rates, a pulsatile pump and a roller pump were employed to propel the distilled water and bovine whole blood to a high flux dialyzer. Pressures at the dialyzer inlet and outlet in the blood circuit and in the dialysate circuit were measured, respectively. From these data, we calculated the transmembrane pressure and predicted the ultrafiltration and backfiltration rates developed by both pumps. Using the bovine’s blood experiment, ultrafiltration and backfiltration rates were 1.6 times higher with the pulsatile pump than with the roller pump. We conclude that utilization of a pulsatile pump in high flux hemodialysis treatments increases ultrafiltration volume, compared with a roller pump under conditions of the same blood flow rate.
참고문헌 (16)
- John F. MAHER, "Replacement of Renal Function by Dialysis", pp.300~346 (2004), Holland, Nederland : Kluwer Academic Publishers
- Juan P. Bosch, "Hemodialysis(High-Efficiency treatments)", pp.105~133 (1993), New York, USA: Churchill Livingstone Inc.
- Young-Suk Yoon, "Principle of Hemodialysis", The Korean Society ofNephrology, 7(3), pp.1~7 (1988)
- TM. Runge, "Hemodialysis: evidence of enhanced molecular clearance and ultrafiltration volume by using pulsatile flow", Int J Artif Organs, 16(9), pp.645~652 (1993)
- K.M. Lim, "Quantitative Analysis of Pulsatile Flow Contribution to Ultrafiltration" (2008), Artificial Organs
- K. Lee, "The Effect of Pulsatile Flow on Ultrafiltration: In-Vitro Study and Comparison with Roller Pump", 의공학회지, 26(5), pp.337~341 (2005), 대한의용생체공학회
- C. Ronco, "The haemodialysis system: basic mechanisms of water and solute transport in extracorporeal renal replacement therapies", Nephrology Dialysis Transplantation, 13(13), pp.3~9 (1998)
- C. Ronco, "Enhancement of convective transport by internal filtration in a modified experimental hemodialyzer: technical note", Kidney Int, 54(3), pp.979~985 (1998)
- M. Mineshima, "Effects of internal filtration on the solute removal efficiency of a dialyzer", ASAIO Journal, 46(4), pp.456~460 (2000)
- J.C. Lee, "Clinical applications and recent progress of therapeutic artificial organ using moving- actuator mechanisms", International Journal of Applied Electro- magnetics and Mechanics, 27(3), pp.205~214 (2008)
- M. Ruperez, "Continuous venovenous renal replacement therapy using a pulsatile blood pump", Pediatr Nephrol, 18, pp.29~32 (2003)
- M. Ruperez, "Comparison of a tubular pulsatile pump and a volumetric pump for continuous venovenous renal replacement therapy in a pediatric animal model", Asaio J, 51, pp.372~375 (2005)
- JL. Herce, "Continuous venovenous renal replacement therapy with a pulsatile tubular blood pump: analysis of efficacy parameters", Artif Organs, 30, pp.64~69 (2006)
- S. Troyanov, "Solute clearances during continuous venovenous haemofiltration at various ultrafiltration flow rates using Multiflow-100 and HF1000 filters", Nephrology Dialysis Transplantation, 18, pp.961~966 (2003)
- Z. Ricci, "Solute removal during continuous renal replacement therapy in critically ill patients: convection versus diffusion", Critical Care, 10, pp.1~7 (2006)
- S. Brunet, "Diffusive and convective solute clearances during continuous renal replacement therapy at various dialysate and ultrafiltration flow rates", American Journal of Kidney Diseases, 34, pp.486~492 (1999)