이중 박동형 생명구조장치의 혈류, 혈압, 박동성의 분석
Analysis on the Blood Flow, Pressure and Pulsatility of a Dual Pulsatile Extracorporeal Life Support
1
조회
0
다운로드
1
인용
0
참고문헌
초록
Abstract
Previous trials for development of a pulsatile-Extracorporeal Life Support had some defects such as insufficient blood flow, high pressure at its membrane oxygenator and the high risk of blood cell damage. To solve those problems of previous pulsatile-ECLSs, we suggest dual pulsatile blood pump structure for the new pulsatile-ECLS. Two pulsatile pumps areconnected in a parallel manner and this new structure raises the inflow capacity and efficiency and it decreases the high blood pressure at membrane oxygenator. In in-vitro experiments, The Energy Equivalent Pressure Increment (EEP inc.) was 10%, and it showed that its pulsatilty was 5~10 times higher than other commercial ECLSIn in-vivo experiments, we had applied a new pulsatile-ECLS to 30 Kg pigs and a new pulsatile-ECLS couldsupport high blood flow and pulsatility above 2 L/min, 10 % EEP inc.
참고문헌 (33)
- J.K. Boucher, "Organ blood flow during pulsatile cardio pulmonary bypass", 36(1), pp.86~90 (1974)
- A.H. Harken, "The influence of pulsatile perfusion on oxygen uptake by the isolated canine hind limb", 70(2), pp.237~241 (1975)
- B.Y. Chiang, "Effect of pulsatile reperfusion on myocardial high energy phosphates following global ischemia", 37, pp.485~487 (1991)
- M.J. Buckley, "Pulsatile and nonpulsatile caridopulmonary bypass review of a counterproductive controversy", 36(6), pp.720~737 (1983)
- M. Tedder, "Pulsatile versus nonpulsatile reperfusion improves cerebral blood flow after cardiac Arrest", 56, pp.453~461 (1993)
- R.B. Shepard, "Relation of pulsatile flow to oxygen consumption and other variables during cardiopulmonary bypass", 58(5), pp.694~702 (1969)
- J.M. Sanderson, "Brain damage in dogs immediately following pulsatile and non-pulsatile blood flows in extracorporeal circulation", 27, pp.275~286 (1972)
- J. Dunn, "and hematologic effects of pulsatile cardiopulmonary bypass", 68(1), pp.138~147 (1974)
- E. Dapper, "Effects of pulsatile and nonpulsatile perfusion mode during extracorporeal circulation - a comparative clinical study", 40, pp.345~351 (1992)
- K. Minami, "Role of the carotid sinus in response of integrated venous system to pulsatile and nonpulsatile perfusion", 104(6), pp.1639~1646 (1992)
- K. Nakayama, "High-amplitude pulsatile pump in extracorporeal circulation with particular reference to hemodynamics", 54, pp.798~809 (1963)
- J.K. Trinkle, "Metabolic comparison of a new pulsatile pump and a roller pump for cardiopulmonary bypass", 58(4), pp.562~569 (1969)
- T. Eiki, "Hemolysis test of a centrifugal pump in a pulsatile mode The effect of pulse rate an RPM variance", 21(12), pp.1284~1287 (1997)
- A. ndar, "Pulsatile and nonpulsatile flows can be quantified in terms of energy equivalent pressure during cardiopulmonary bypass for direct comparisons", 45, pp.610~614 (1999)
- A. ndar, "Defining pulsatile perfusion: quantification in terms of energy equivalent pressure", 23(8), pp.712~716 (1999)
- R.H. Paul, "review of a counterproducitve controversy", pulsatile and 36(npulsatile cardiopulmonary bypass 6), pp.720~737 (1983)
- A. ndar, "Pulsatile perfusion improves regional myocardial blood flow during and after hypothermic cardiopulmonary bypass in a neonatal piglet model", 48, pp.90~95 (2002)
- A. ndar, "Pediatric physiologic pulsatile pump enhances cerebral and renal blood flow during and after cardiopulmonary bypass", 26(11), pp.919~923 (2002)
- Y. Nos, "Can we develop a nonpulsatile permanet rotary blood pump? yes we can", 20(6), pp.467~474 (1996)
- T.P. Green, "Comparison of pressure-volume-flow relationships in centrifugal and roller pump extracorporeal membrane oxygenation systems for neonates", 37(4), pp.572~576 (1991)
- J.B. Atkinson, "A simplified method for autoregulation of blood flow in the extracorporeal membrane oxygenation circuit", 24(3), pp.251~252 (1989)
- K. Setz, "Comparison of a new venous control device with a bladder box system for use in ECMO", 38(4), pp.835~840 (1992)
- R.H. Steinhorn, "Hemolysis during ling-term extracorporeal membrane oxygenation", 115(4), pp.625~30 (1989)
- H. Schima, "Effect of stationary guiding vanes on improvement of the washout behind the rotor in centrifugal blood pumps", 38(3), pp.220~ (1992)
- H.S. Lee, "In-vivo evaluation of the pulsatile ECLS system", 6, pp.25~29 (2003)
- Y.R. Rho, "Applications of the pulsatile flow versatile ECLS: In vivo Studies", 26, pp.428~435 (2003)
- J. Chung, "Assessment and improvement of the system efficiency for the moving-actuator type biventrifucular assist devices", 28(6), pp.549~555 (2004)
- H.S. Son, "The effect of pulsatile versus non-pulsatile extracorporeal circulation on the pattern of coronary artery blood flow during cardiac arrest", 28(6), pp.609~616 (2005)
- T.S. Kim, "Application of a pressure-relieving air compliance chamber in a single-pulsatile extracorporeal life support system: an experimental study", 28(12), pp.1106~1109 (2004)
- H.M. Kim, "Evaluation of the pulsatile ECLS system in heart failure model An animal experiment", pp.384~ (2001)
- J. Sands, "Difference between delivered and prescribed blood flow in hemodialysis", 42, pp.74~ (1996)
- C.T. Ducko, "The effect of valve type and drive line dP", dt on hemolysis in the pneumatic ventricular assist device, 1994., pp.18(6), 454~460
- S.W. Choi, "Development of a new pulsatile extracorporeal life support device incorporating a dual pulsatile blood pump", 29(6), pp.583~590 (2006)