본문 바로가기
KERD
논문KCI 등재

새로운 방식의 나사형 인공디스크 고정체 해석

Finite Element Analysis of a Newly Designed Screw Type Fixture for an Artificial Intervertebral Disc

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

The various total replacement artificial discs have developed because spinal fusion has shown a lesser mobility of an operated segment and an accelerated degeneration at adjacent discs. But almost artificial discs have not yet been reached on the substitute surgery of fusion because many problems such as those clinical success rates were not more than them of fusion have not solved. In this paper, vertically inserted assemble-screw fixture in vertebrae was proposed to improve the fixed capability of artificial disc. And also, to evaluate the design suitability of newly designed screw-type, including fixtures of commercial discs such as wedge and plate type, the 1/4 finite element model with a vertebra and various implanted fixtures were generated, and next, 3 bending motions such as flexion, bending and twisting under the moment of 10Nm and compression under the force of 1000N were considered, respectively and finally, FE analyses were performed. Results of three fixture types were compared, such as Range of Motion and maximal stress, and so on. For ROM, the screw type was average 58% less than the wedge type and was average 42% less than the plate type under all loading conditions. For average stress ratio at closer nodes between vertebra and each fixture, the wedge type was the lowest as minimum 0.02 in twisting, screw types were the highest as maximum 0.28 in compression. As the results of using cement material, it was predicted that the instability problem of the wedge type was better solved. The screw type which could be increased by implanting depth according to the number of assembling mid screws, showed that the decreased tendency of ROMs and maximal cancellous bone stresses. In further study, controlling the number of assembling screws that was suitable for a patient’s bone quality, development of surgical tools and keeping on design supplementations, which will be able to develop the competitive artificial disc.

Abstract

The various total replacement artificial discs have developed because spinal fusion has shown a lesser mobility of an operated segment and an accelerated degeneration at adjacent discs. But almost artificial discs have not yet been reached on the substitute surgery of fusion because many problems such as those clinical success rates were not more than them of fusion have not solved. In this paper, vertically inserted assemble-screw fixture in vertebrae was proposed to improve the fixed capability of artificial disc. And also, to evaluate the design suitability of newly designed screw-type, including fixtures of commercial discs such as wedge and plate type, the 1/4 finite element model with a vertebra and various implanted fixtures were generated, and next, 3 bending motions such as flexion, bending and twisting under the moment of 10Nm and compression under the force of 1000N were considered, respectively and finally, FE analyses were performed. Results of three fixture types were compared, such as Range of Motion and maximal stress, and so on. For ROM, the screw type was average 58% less than the wedge type and was average 42% less than the plate type under all loading conditions. For average stress ratio at closer nodes between vertebra and each fixture, the wedge type was the lowest as minimum 0.02 in twisting, screw types were the highest as maximum 0.28 in compression. As the results of using cement material, it was predicted that the instability problem of the wedge type was better solved. The screw type which could be increased by implanting depth according to the number of assembling mid screws, showed that the decreased tendency of ROMs and maximal cancellous bone stresses. In further study, controlling the number of assembling screws that was suitable for a patient’s bone quality, development of surgical tools and keeping on design supplementations, which will be able to develop the competitive artificial disc.

참고문헌 (14)

  1. Lim, J.W, "Effects of artificial discs on spinal system", 박사, pp.2008
  2. Andrew, P.D, "Experimental and theological investigation into the effects of artificial disc implantation on the lumbar spine", 박사, pp.2001
  3. Hedman, T.P, "Design of an intervertebral disc prosthesis", Spine, 16(6 Suppl), pp.S256~260 (1991)
  4. Bushelow, M., "Comparison of wear rates:metal", UHMWPE and metal on metal total disc arthroplasty, 2007, S31~S32
  5. Murrey, D, "Results of the prospective,randomized,controlled multicenter Food and Drug Administration investiga-tional device exemption study of the ProDisc-C total disc replacement versus anterior discectomy and fusion for the treatment of 1-level symptomatic cervical disc disease" (2008), Spine J
  6. Bertagnoli, R, "Treatment of symptomatic adjacent-segment degeneration after lumbar fusion with total disc arthroplasty by using the prodisc prosthesis:a prospective study with 2-year minimum follow up", J Neurosurg Spine, 4(2), pp.91~97 (2006)
  7. Chung, S.S, "Lumbar total disc replacement using ProDisc II:a prospective study with a 2-year minimum follow-up", J Spinal Disord Tech, 19(6), pp.411~415 (2006)
  8. Shim, C.S, "Posterior avulsion fracture at adjacent vertebral body during cervical disc replacement with ProDisc-C:a case report", J Spinal Disord Tech, 20(6), pp.468~472 (2007)
  9. Park, C.K, "Degenerative changes of discs and facet joints in lumbar total disc replacement using ProDisc II:minimum two-year follow-up", Spine, 33(16), pp.1755~1761 (2008)
  10. Bertagnoli, R, "The treatment of disabling single-level lumbar discogenic low back pain with total disc arthroplasty utilizing the Prodisc prosthesis:a prospective study with 2-year minimum follow-up", Spine, 30(19), pp.2230~2236 (2005)
  11. McAfee, P, "SB Charite disc replacement: Biologic ingrowth using a non-human promate model, in The Artificial Disc", pp.63~71 (2003), Springer-Verlag
  12. Taylor, R.L, "The patch test-A condition for assessing FEM convergence", Int.J.Num.Mech.Engng., 22(1), pp.39~62 (1986)
  13. Siegele,D., "Numerical Invesigation of the influence of implant shape on stress distribution in the jaw bone", The international Journal of Oral & Maxillofacial Implant, 4(4), pp.333~340 (1989)
  14. Overaker,D.W., "Finite element analysis of vertebral body mechanics with a nonlinear microstructural model for the trabecular core", J Biomech Eng, 121(5), pp.542~550 (1999)

관련 논문AI추천