다채널 미세전극칩 임피던스 분석을 위한 자동 스위칭 시스템: 한계점 및 개선 방안
Automatic Switching System for The Impedance Analysis of Multichannel Microelectrode Arrays: Limits and Improvement Scheme
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Electrode impedances are measured to quantitatively characterize the electrode-electrolyte or cell-electrode interfaces. In the case of high-density microelectrode arrays(MEAs) that have been developed for brainmachine interface applications, the characterization process becomes a repeating and time-consuming task; a system that can perform the measurement and analysis in an automated fashion with accuracy and speed is required. However,due to the large number of channels, parasitic capacitance and off-capacitance components of the switching system become the major factors that decreased the accuracy for the measurement of high impedance microelectrodes. Here we investigated the implementation of automatic impedance measurement system with analyzing the causes of possible measurement-related problems in multichannel switching configuration. Based on our multi-channel measurement circuit model, we suggest solutions to the problems and introduce a novel impedance measurement scheme using electro-mechanical relays. The implemented measurement system could measure |Z| < 700 kΩ of impedance in −10% errors, which can be widely applicable to high density neural recording MEAs.
Abstract
Electrode impedances are measured to quantitatively characterize the electrode-electrolyte or cell-electrode interfaces. In the case of high-density microelectrode arrays(MEAs) that have been developed for brainmachine interface applications, the characterization process becomes a repeating and time-consuming task; a system that can perform the measurement and analysis in an automated fashion with accuracy and speed is required. However,due to the large number of channels, parasitic capacitance and off-capacitance components of the switching system become the major factors that decreased the accuracy for the measurement of high impedance microelectrodes. Here we investigated the implementation of automatic impedance measurement system with analyzing the causes of possible measurement-related problems in multichannel switching configuration. Based on our multi-channel measurement circuit model, we suggest solutions to the problems and introduce a novel impedance measurement scheme using electro-mechanical relays. The implemented measurement system could measure |Z| < 700 kΩ of impedance in −10% errors, which can be widely applicable to high density neural recording MEAs.
참고문헌 (15)
- K.C.Cheung, "Implantable microscale neural interfaces", Biomed. Microdevices, 9(6), pp.923~938 (2007)
- C.A.J. Thomas, "A miniature microelectrode array to monitor the bioelectric activity of cultured cells", Exp. Cel. Res, 74, pp.61~66 (1972)
- G.W. Gross, "Simultaneous single unit recording in vitro with a photoetched laser deinsulated gold multimicroelectrode surface", IEEE Trans. Biomed. Eng, 26, pp.273~279 (1979)
- J.Pine, "Recording action potentials from cultured neurons with extracellular microcircuit electrodes", J. Neurosci. Methods, 2, pp.19~31 (1980)
- 남윤기, "뉴런온칩 기술: 미세전극칩시스템과 신경세포 패터닝 기술", 의공학회지, 30(2), pp.103~112 (2009), 대한의용생체공학회
- R. Ehret, "Monitoring of cellular behaviour by impedance measurements on interdigitated electrode structures", Biosensors Bioelectron, 12(1), pp.29~41 (1997)
- D.R. Merrill, "Impedance characterization of microarray recording electrodes in vitro", IEEE trans. Biomed. Eng, 52(11), pp.1960~1965 (2005)
- K. Gunalan, "An automated system for measuring tip impedance and among-electrode shunting in high-electrode count microelectrode arrays", J. Neurosci. Methods, 178(2), pp.263~269 (2009)
- Q. Liu, "Impedance studies of bio-behavior and chemosensitivity of cancer cells by micro-electrode arrays", Biosensors Bioelectron, 24(5), pp.1305~1310 (2009)
- C.K. Liang, "A multifunctional online measurement system for neuron-microelectrode interface study", the 26th Annual International Conference of the IEEE Engineering In Medicine And Biology Society (2004)
- G.T.A. Kovacs, "Microelectrode models for neural interfaces in Enabling technologies for cultured neural networks", pp.121~165 (1994), Academic Press
- D.A. Borkholder, "Cell based biosensors using microelectrodes", 박사, pp.1998
- D.A.Robinson, "The electrical properties of metal microelectrodes", Proceedings of the IEEE, 56(6), pp.1065~1071 (1968)
- C.H. Chen, "Micro-multi-probe electrode array to measure neural signals", Biosensors Bioelectron, 24(7), pp.1911~1917 (2009)
- D.T. Price, "Design rule for optimization of microelectrodes used in electric cell-substrate impedance sensing(ECIS)", Biosensors Bioelectron, 24(7), pp.2071~2076 (2009)