임피던스 변환 회로를 이용한 건식능동뇌파전극 개발

Development of an Active Dry EEG Electrode Using an Impedance-Converting Circuit

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류신경과학

초록

Background: A dry-type electrode is an alternative to the conventional wet-type electrode, because it can be applied without any skin preparation, such as a conductive electrolyte. However, because a dry-type electrode without electrolyte has high electrode-to-skin impedance, an impedance-converting amplifier is typically used to minimize the distortion of the bioelectric signal. In this study, we developed an active dry electroencephalography (EEG) electrode using an impedance converter, and compared its performance with a conventional Ag/AgCl EEG electrode. Methods: We developed an active dry electrode with an impedance converter using a chopper-stabilized operational amplifier. Two electrodes, a conventional Ag/AgCl electrode and our active electrode, were used to acquire EEG signals simultaneously, and the performance was tested in terms of (1)the electrode impedance, (2) raw data quality, and (3) the robustness of any artifacts. Results: The contact impedance of the developed electrode was lower than that of the Ag/AgCl electrode (0.3±0.1 vs. 2.7±0.7 kΩ, respectively). The EEG signal and power spectrum were similar for both electrodes. Additionally, our electrode had a lower 60-Hz component than the Ag/AgCl electrode (16.64 vs. 24.33 dB, respectively). The change in potential of the developed electrode with a physical stimulus was lower than for the Ag/AgCl electrode (58.7±30.6 vs. 81.0±19.1 μV, respectively), and the difference was close to statistical significance (P=0.07). Conclusions: Our electrode can be used to replace Ag/AgCl electrodes, when EEG recording is emergently required, such as in emergency rooms or in intensive care units.

Abstract

Background: A dry-type electrode is an alternative to the conventional wet-type electrode, because it can be applied without any skin preparation, such as a conductive electrolyte. However, because a dry-type electrode without electrolyte has high electrode-to-skin impedance, an impedance-converting amplifier is typically used to minimize the distortion of the bioelectric signal. In this study, we developed an active dry electroencephalography (EEG) electrode using an impedance converter, and compared its performance with a conventional Ag/AgCl EEG electrode. Methods: We developed an active dry electrode with an impedance converter using a chopper-stabilized operational amplifier. Two electrodes, a conventional Ag/AgCl electrode and our active electrode, were used to acquire EEG signals simultaneously, and the performance was tested in terms of (1)the electrode impedance, (2) raw data quality, and (3) the robustness of any artifacts. Results: The contact impedance of the developed electrode was lower than that of the Ag/AgCl electrode (0.3±0.1 vs. 2.7±0.7 kΩ, respectively). The EEG signal and power spectrum were similar for both electrodes. Additionally, our electrode had a lower 60-Hz component than the Ag/AgCl electrode (16.64 vs. 24.33 dB, respectively). The change in potential of the developed electrode with a physical stimulus was lower than for the Ag/AgCl electrode (58.7±30.6 vs. 81.0±19.1 μV, respectively), and the difference was close to statistical significance (P=0.07). Conclusions: Our electrode can be used to replace Ag/AgCl electrodes, when EEG recording is emergently required, such as in emergency rooms or in intensive care units.

참고문헌 (26)

  1. Kwon OY, "Artifact in Electroencephalography", Korean J Clin Neurophysiol, 5, pp.157~169 (2003)
  2. Niedermeyer E, "Electroencephalography: basic principles, clinical applications, and related fields", pp.110~121 (1999), Williams & Wilkins
  3. Huigen E, "Investigation into the origin of the noise of surface electrodes", Med Biol Eng Comput, 40, pp.332~338 (2002)
  4. Carter SJ, "Comparison of impedance at the microelectrode-saline and microelectrodeculture medium interface", IEEE Trans Biomed Eng, 39, pp.1123~1129 (1992)
  5. Rosell J, "Skin impedance from 1 Hz to 1 MHz", IEEE Trans Biomed Eng, 35, pp.649~651 (1988)
  6. Padmadinata FZ, "Microelectronic skin electrode", Sensor Actuat B-Chem, 1, pp.491~494 (1990)
  7. Webster JG, "Medical instrumentation: application and design", pp.248~290 (1992), Houghton Mifflin
  8. "Handbook of chemistry and physics" (1974), CRC Press
  9. Ferree TC, "Scalp electrode impedance, infection risk, and EEG data quality", Clin Neurophysiol, 112, pp.536~544 (2001)
  10. Nishimura S, "Clinical-application of an active electrode using an operational-amplifier", IEEE Trans Biomed Eng, 39, pp.1096~1099 (1992)
  11. Searle A, "A direct comparison of wet, dry and insulating bioelectric recording electrodes", Physiol Meas, 21, pp.271~283 (2000)
  12. Bergey GE, "Electrocardiogram recording with pasteless electrodes", IEEE Trans Biomed Eng, BME-18, pp.206~211 (1971)
  13. Mclaughlin JA, "Novel dry electrode ECG sensor system", Proceedings of the 16th annual international conference: IEEE EMBS, pp.804~ (1994)
  14. Geddes LA, "Dry electrodes and holder for electro-oculography", Med Biol Eng, 11, pp.69~72 (1973)
  15. Fujisawa M, "Surface electromyographic electrode pair with built-in buffer-amplifiers", J Prosthet Dent, 63, pp.350~352 (1990)
  16. Degen T, "Enhancing interference rejection of preamplified electrodes by automated gain adaption", IEEE Trans Biomed Eng, 51, pp.2031~2039 (2004)
  17. Valverde E, "Effect of electrode impedance in improved buffer amplifier for bioelectric recordings", J Med Eng Technol, 28, pp.217~222 (2004)
  18. Taheri BA, "A dry electrode for EEG recording", Electroencephalogr Clin Neurophysiol, 90, pp.376~383 (1994)
  19. "TLC2652, TLC2652A, TLC2652Y Advanced LinCMOS PRECISION CHOPPER-STABILIZED OPERATIONAL AMPLIFIERS", Texas Instruments, ~ (2001)
  20. Hsieh KC, "A low-noise chopper-stabilized differential switched-capacitor filtering technique", IEEE J Solid-St Circ, 16, pp.708~715 (1981)
  21. Hanasusanto GA, "A chopper stabilized pre amplifier for biomedical signal acquisition", International symposium on integrated circuits: IEEE, pp.200~203 (2007)
  22. Miller HA, "Biomedical electrode technology: theory and practice" (1974), Academic Press
  23. Ko D, "Development of PDMS and silver ball based dry-type flexible EEG electrode for long-term EEG recording. 13th Conference of the Korean Society of Clinical Neurophysiology", Korean J Clin Neurophysiol, 12, pp.114~ (2010)
  24. Ruffini G, "ENOBIO dry electrophysiology electrode; first human trial plus wireless electrode system", 2007 Annual International Conference: IEEE EMBS, pp.6690~6694 (2007)
  25. Casson A, "Wearable electroencephalography", IEEE Eng Med Biol Mag, IEEE EMBS, 29, pp.44~56 (2010)
  26. Shin JH, "The trends of BCI based Entertainic, Technology", The Magazine of the IEEK, 34, pp.71~82 (2007)

관련 논문AI추천

이 간행물의 다른 논문 (8건)