본문 바로가기
KERD
논문KCI 등재

좌심실보조장치의 출구 캐뉼라의 삽관 위치에 따른 혈액관류에 대한 시뮬레이션 연구

Simulation Study of Blood Perfusion according to Outflow Cannulation Site of Left Ventricular Assist Device

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Outflow cannulation site of left ventricular assist device(LVAD) chosen by considering anatomical structure of thoracic cavity and vascular system. Though outflow cannulation site influences blood perfusion at each branch, there is no standard rule or quantitative data. In this study, we computed the amount of blood perfusion at each arterial branch numerically according to outflow cannulation sites(ascending aorta, aortic arch, descending aorta). We generated computational meshes to the three-dimensionally reconstructed arterial system. Clinically measured arterial pressure were used for inlet boundary condition, porous media were applied to mimic blood flow resistance. Blood perfusion through left common carotid artery was 2.5 times higher than other cases, and that through right common carotid artery was 1.1 times higher than other branches. Although this is simulation study, will be useful reference data for the clinical study of LVAD which considers blood perfusion efficiency.

Abstract

Outflow cannulation site of left ventricular assist device(LVAD) chosen by considering anatomical structure of thoracic cavity and vascular system. Though outflow cannulation site influences blood perfusion at each branch, there is no standard rule or quantitative data. In this study, we computed the amount of blood perfusion at each arterial branch numerically according to outflow cannulation sites(ascending aorta, aortic arch, descending aorta). We generated computational meshes to the three-dimensionally reconstructed arterial system. Clinically measured arterial pressure were used for inlet boundary condition, porous media were applied to mimic blood flow resistance. Blood perfusion through left common carotid artery was 2.5 times higher than other cases, and that through right common carotid artery was 1.1 times higher than other branches. Although this is simulation study, will be useful reference data for the clinical study of LVAD which considers blood perfusion efficiency.

참고문헌 (11)

  1. 김은혜, "심실의 부하감소 측면에서 좌심실 보조장치의 최적 치료시기 예측을 위한 시뮬레이션 연구", 한국기계가공학회지, 12(4), pp.22~28 (2013), 한국기계가공학회
  2. Casarotto D., "The last to die is hope: prolonged mechanical circulatory support with a Novacor left ventricular assist device as a bridge to transplantation", J Thorac Cardiovasc Surg, 125(2), pp.417~418 (2003)
  3. Goldstein D. J., "Worldwide experience with the MicroMed DeBakey Ventricular Assist Device as a bridge to transplantation", Circulation, 108(Suppl 1), pp.II272~II277 (2003)
  4. Timms D., "Atrial versus ventricular cannulation for a rotary ventricular assist device", Artif Organs, 34(9), pp.714~720 (2010)
  5. 임기무, "Theoretical Estimation of Cannulation Methods for Left Ventricular Assist Device Support as a Bridge to Recovery", Journal of Korean Medical Science, 26(12), pp.1591~1598 (2011), 대한의학회
  6. Lim K.M., "Computational analysis of the effect of the type of LVAD flow on coronary perfusion and ventricular afterload", J Physiol Sci, 59(4), pp.307~316 (2009)
  7. Park J.Y., "Pseudo-organ boundary conditions applied to a computational fluid dynamics model of the human aorta", Comput Biol Med, 37(8), pp.1063~1072 (2007)
  8. Biswajit Kar, "The Effect of LVAD Aortic Outflow-Graft Placement on Hemodynamics and Flow Implantation Technique and Computer Flow Modeling", Tex Heart Inst J, 32(3), pp.294~298 (2005)
  9. George, Stephanie M., "Computational Analyses of an In-Vitro Aneurysm Model Based on Three-Dimensional Angiography With Comparison to Phase Contrast Magnetic Resonance Imaging and Dye Injection Studies", ASME 2010 Summer Bioengineering Conference. American Society of Mechanical Engineers, pp.163~164 (2010)
  10. Nichols W.W., "McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles" (1998), Edward Arnold
  11. Torii Ryo, "Numerical investigation of the effect of hypertensive blood pressure on cerebral aneurysm dependence of the effect on the aneurysm shape", International Journal for Numerical Methods in Fluids, 54(68), pp.995~1009 (2007)

관련 논문AI추천