주사기의 마찰 특성이 다단 연결된 시린지펌프의 구동 중 하중부하와 카테터 내부 유체압력에 미치는 영향 분석

Effect of the Frictional Characteristics of Syringe on the Driving Workload and the in-line Hydraulic Pressure of Simultaneously Operating Syringe Pumps

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Purpose: Accuracy and stability during drug administration via syringe pumps are vital in intensive care units. However, the effect of frictional characteristic of the pump-equipped syringe and connection protocol of syringe pumps remain elusive. Materials and Methods: Here, these effects, pump arrangement in multiple con- nection on workload, and hydraulic pressure at the outlet of each syringe pump were evaluated. Results: Statistical analysis demonstrated that the driving workload and hydraulic pressure were substantially affected by variations in the mechanical quality of syringe and arrangement of the connected syringe pumps. Conclusion: Therefore, to increase accuracy and stability, clinics should use high-quality syringes and arrange syringe pumps to minimize positional variations due to different infusion rates in intensive care units.

Abstract

Purpose: Accuracy and stability during drug administration via syringe pumps are vital in intensive care units. However, the effect of frictional characteristic of the pump-equipped syringe and connection protocol of syringe pumps remain elusive. Materials and Methods: Here, these effects, pump arrangement in multiple con- nection on workload, and hydraulic pressure at the outlet of each syringe pump were evaluated. Results: Statistical analysis demonstrated that the driving workload and hydraulic pressure were substantially affected by variations in the mechanical quality of syringe and arrangement of the connected syringe pumps. Conclusion: Therefore, to increase accuracy and stability, clinics should use high-quality syringes and arrange syringe pumps to minimize positional variations due to different infusion rates in intensive care units.

참고문헌 (12)

  1. Hayes C, "Medication errors in hospitals : a literature review of disruptions to nursing practice during medication administration", J Clin Nurs, 24, pp.3063~3076 (2015)
  2. Kim GH, "A new injection rate estimation technique for on-site screening test of medication infusion pump by nurses", PI Mech Eng H, 234(4), pp.370~376 (2020)
  3. Batliner M, "Evaluation of a novel flow-controlled infusion pump for precise and continuous drug delivery at low flow rates : a laboratory study", Anaesthesia, 74(11), pp.1425~1431 (2019)
  4. Yun SU, "Experimental study about the effect of the mechanical characteristics of a syringe on the erroneous occlusion alarm generation of syringe pump", J Biomed Eng Res, 42(2), pp.55~61 (2021)
  5. Kim JH, "A study on the cognizance of hospital and oriental hospital workers for medical device safety information monitoring", J Biomed Eng Res, 42(3), pp.86~93 (2021)
  6. Park SM, "ISO14971 : 2019 Detailed analysis and periodic safety update report establishment method for the single use medical device – focusing on medical device regulation 2017", 745 requirements, 2023, pp.44(1), 1~10
  7. Weiss M, "Impact of infusion line compliance on syringe pump performance", Paediatr Anaesth, 10(6), pp.595~599 (2000)
  8. Kim DW, "The effect of syringe size on the performance of an infusion pump", Paediatr Anaesth, 9(4), pp.335~337 (1999)
  9. Schmidt N, "Impact of syringe size on the performance of infusion pumps at low flow rates", Pediatr Crit Care Med, 11(2), pp.282~286 (2010)
  10. Neff SB, "Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps", Eur J Anaesthesiol, 24(7), pp.602~608 (2007)
  11. Weiss M, "The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps", Anaesthesia, 55(11), pp.1094~1098 (2000)
  12. Kim GH, "Real-time monitoring of the actual infusion rate of syringe pump using 2D image marker tracking", J Biomed Eng Res, 44(1), pp.92~98 (2023)

관련 논문AI추천

이 간행물의 다른 논문 (13건)