2단 절개 형태를 가지는 악교정 수술 장치 연구

A Study of Orthognathic Surgical Guides with Two-stage Split Path

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.

Abstract

In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.

참고문헌 (14)

  1. Hashemi S, "Computational mod-elling of process–structure–property–performance relation-ships in metal additive manufacturing : a review", International materials reviews, pp.1~46 (2022)
  2. Ministry of Food and Drug Safety., "Understanding Medical Device 3D Printing Technology" (2018)
  3. Schouman T, "Accuracy Evaluation of CAD", CAM Generated Splints in Orthognathic Surgery : A Cadaveric Study, 2015, pp.11(1), 11~24
  4. Rinaldo Z, "Clinical Accuracy of Splintless Maxillary Positioning with Aid of CAD", CAM Fabricated Surgical Cutting Guides and Titanium Plates, 2022, pp.12(10), 147~157
  5. 김민욱, "3D 융합 기술을 활용한 악교정 수술장치 연구", 한국정밀공학회지, 36(3), pp.301~309 (2019), 한국정밀공학회[DOI]
  6. Kang SH, "Maxillary Cutting Guide for Executing a Simulated Osteotomy and Removing the Bony Interference During Orthognathic Surgery", Journal of Medical Devices, pp.44~50 (2015)
  7. Kim YS, "Computer Aided Design and Applications" (2018), Sigma Press
  8. Vale F, "3D Virtual Planning in Orthognathic Surgery and CAD", CAM Surgical Splints Generation in One Patient with Craniofacial Microsomia : A Case Report, 2016, pp.21(1), 89~100
  9. Li B, "A Novel Method of Computer Aided Orthognathic Surgery Using Individual CAD", CAM Templates : A Combination of Osteotomy and Reposi-tioning Guides, 2013, pp.51(8), 239~244

관련 논문AI추천

이 간행물의 다른 논문 (9건)