Micro-computed tomography evaluation of general trends in aligner thickness and gap width after thermoforming procedures involving six commercial clear aligners: An in vitro study

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Objective: To assess the effects of thermoforming on aligner thickness and gap width in six aligner systems with the same nominal thickness. Methods: Six passive upper aligners of different brands were adapted to a single printed cast. Each sample was evaluated with high-resolution micro-computed tomography. To investigate aligner thickness and gap width, two-dimensional (2D) analysis was conducted assessing the effects of the following variables: tooth type (central incisor, canine, and first molar), 2D reference points, and aligner type. Data were analyzed and compared using analysis of variance and Tukey’s post-hoc tests (p < 0.05). Results: Tooth type, dental region, and aligner type affected both the gap width and aligner thickness. The aligner thickness remained moderately stable across the arch only in the F22. Conclusions: All thermoformed samples displayed smaller aligner thickness and gap width at anterior teeth and both gingival and coronal centers than at posterior teeth and occlusal surfaces.

Abstract

Objective: To assess the effects of thermoforming on aligner thickness and gap width in six aligner systems with the same nominal thickness. Methods: Six passive upper aligners of different brands were adapted to a single printed cast. Each sample was evaluated with high-resolution micro-computed tomography. To investigate aligner thickness and gap width, two-dimensional (2D) analysis was conducted assessing the effects of the following variables: tooth type (central incisor, canine, and first molar), 2D reference points, and aligner type. Data were analyzed and compared using analysis of variance and Tukey’s post-hoc tests (p < 0.05). Results: Tooth type, dental region, and aligner type affected both the gap width and aligner thickness. The aligner thickness remained moderately stable across the arch only in the F22. Conclusions: All thermoformed samples displayed smaller aligner thickness and gap width at anterior teeth and both gingival and coronal centers than at posterior teeth and occlusal surfaces.

참고문헌 (21)

  1. Lombardo L, "Optical properties of orthodontic aligners--spectrophotometry analysis of three types before and after aging", Prog Orthod, 16, pp.41~ (2015)
  2. 류정현, "Effects of thermoforming on the physical and mechanical properties of thermoplastic materials for transparent orthodontic aligners", 대한치과교정학회지, 48(5), pp.316~325 (2018), 대한치과교정학회
  3. Gao L, "Forces and moments delivered by the PET-G aligner to a maxillary central incisor for palatal tipping and intrusion", Angle Orthod, 87, pp.534~541 (2017)
  4. Elkholy F, "Forces and moments applied during derotation of a maxillary central incisor with thinner aligners : an invitro study", Am J Orthod Dentofacial Orthop, 151, pp.407~415 (2017)
  5. Cowley DP, "The effect of gingival-margin design on the retention of thermoformed aligners", J Clin Orthod, 46, pp.697~702 (2012)
  6. Lombardo L, "Predictability of orthodontic movement with orthodontic aligners : a retrospective study", Prog Orthod, 18, pp.35~ (2017)
  7. Lombardo L, "The effect of temperature on the mechanical behavior of nickel-titanium orthodontic initial archwires", Angle Orthod, 83, pp.298~305 (2013)
  8. Martina S, "In vitro cytotoxicity of different thermoplastic materials for clear aligners", Angle Orthod, 89, pp.942~945 (2019)
  9. Alexandropoulos A, "Chemical and mechanical characteristics of contemporary thermoplastic orthodontic materials", Aust Orthod J, 31, pp.165~170 (2015)
  10. Jindal P, "Mechanical and geometric properties of thermoformed and 3D printed clear dental aligners", Am J Orthod Dentofacial Orthop, 156, pp.694~701 (2019)
  11. Bucci R, "Thickness of orthodontic clear aligners after thermoforming and after 10 days of intraoral exposure : a prospective clinical study", Prog Orthod, 20, pp.36~ (2019)
  12. Mantovani E, "Scanning electron microscopy evaluation of aligner fit on teeth", Angle Orthod, 88, pp.596~601 (2018)
  13. Mantovani E, "Scanning electron microscopy analysis of aligner fitting on anchorage attachments", J Orofac Orthop, 80, pp.79~87 (2019)
  14. Ryokawa H, "The mechanical properties of dental thermoplastic materials in a simulated intraoral environment", Orthod Waves, 65, pp.64~72 (2006)
  15. Lombardo L, "MicroCT X-ray comparison of aligner gap and thickness of six brands of aligners : an in-vitro study", Prog Orthod, 21, pp.12~ (2020)
  16. Gierie WV, "Clear aligner therapy : an overview", J Clin Orthod, 52, pp.665~674 (2018)
  17. Dahlberg G, "Statistical methods for medical and biological students" (1940), Interscience Publishers Inc
  18. "A language and environment for statistical computing" (2011)[URL]
  19. Dorow C, "In vivo experimental investigation of tooth mobility in humans", Biomed Tech, 47, pp.20~25 (2002)
  20. Moshiri S, "Cephalometric evaluation of adult anterior open bite non-extraction treatment with Invisalign", Dental Press J Orthod, 22, pp.30~38 (2017)
  21. Ojima K, "The biomechanics of aligner orthodontics in openbite cases", J Clin Orthod, 53, pp.699~712 (2019)

관련 논문AI추천

Comparison of dimensional accuracy between direct-printed and thermoformed alignersAndrew Hayes, Jin-Young Choi, Julie McCray, Ki Beom Kim, Nickolas Koenig, Patricia Schneider대한치과교정학회 · 대한치과교정학회지2022 · 제52권 제4호Evaluation of different designs of 3D printed clear aligners on mandibular premolar extrusion using force/moment measurement devices and digital image correlation methodJong-Chan Baik, Ki Beom Kim, Seong-Sik Kim, Soo-Byung Park, Sung-Hun Kim, Yonghun Cho, Yong-Il Kim, Youn-Kyung Choi, Yunju Baek대한치과교정학회 · 대한치과교정학회지2024 · 제54권 제6호Force and moment analysis of clear aligners: Impact of material properties and design on premolar rotationDong-Woo Kim, Hyun-Jun Lee, Ki Beom Kim, Seong-Sik Kim, Soo-Byung Park, Sung-Hun Kim, Yong-Il Kim, Youn-Kyung Choi대한치과교정학회 · 대한치과교정학회지2025 · 제55권 제3호Effects of thermoforming on the physical and mechanical properties of thermoplastic materials for transparent orthodontic aligners권재성, 김광만, 류정현, 차정열, Heng Bo Jiang대한치과교정학회 · 대한치과교정학회지2018 · 제48권 제5호Pattern of change in force and moment according to amounts of labial tipping and material types of clear aligner: In vitro experiment with a six-axis mechanical sensor apparatusIl-Hyung Yang, Keun-Chul Cho, Ki Beom Kim, Michael Lee, Seung-Hak Baek대한치과교정학회 · 대한치과교정학회지2025 · 제55권 제4호Accuracy of orthodontic movements with 3D printed aligners: A prospective observational pilot studyAlessandra Campobasso, Anna De Mari, Filippo Forin Valvecchi, Giorgio Gastaldi, Giovanni Battista, Marco Migliorati, Paolo Pesce, Sara Drago, Tommaso Castroflorio대한치과교정학회 · 대한치과교정학회지2024 · 제54권 제3호

이 간행물의 다른 논문 (7건)