Effects of thermoforming on the physical and mechanical properties of thermoplastic materials for transparent orthodontic aligners
2
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Objective: The aim of this systematic multiscale analysis was to evaluate the effects of thermoforming on the physical and mechanical properties of thermoplastic materials used to fabricate transparent orthodontic aligners (TOAs). Methods: Specimens were fabricated using four types of thermoplastic materials with different thicknesses under a thermal vacuum. Transparency, water absorption and solubility, surface hardness, and the results of three-point bending and tensile tests were evaluated before and after thermoforming. Data were analyzed using one-way analysis of variance and Student’s t -test. Results: After thermoforming, the transparency of Duran and Essix A+ decreased, while the water absorption ability of all materials; the water solubility of Duran, Essix A+, and Essix ACE; and the surface hardness of Duran and Essix A+ increased. The flexure modulus for the 0.5-mm-thick Duran, Essix A+, and eCligner specimens increased, whereas that for the 0.75-/1.0-mm-thick Duran and eClginer specimens decreased. In addition, the elastic modulus increased for the 0.5-mm-thick Essix A+ specimens and decreased for the 0.75-mm-thick Duran and Essix ACE and the 1.0-mm-thick Essix ACE specimens. Conclusions: Our findings suggest that the physical and mechanical properties of thermoplastic materials used for the fabrication of TOAs should be evaluated after thermoforming in order to characterize their properties for clinical application.
Abstract
Objective: The aim of this systematic multiscale analysis was to evaluate the effects of thermoforming on the physical and mechanical properties of thermoplastic materials used to fabricate transparent orthodontic aligners (TOAs). Methods: Specimens were fabricated using four types of thermoplastic materials with different thicknesses under a thermal vacuum. Transparency, water absorption and solubility, surface hardness, and the results of three-point bending and tensile tests were evaluated before and after thermoforming. Data were analyzed using one-way analysis of variance and Student’s t -test. Results: After thermoforming, the transparency of Duran and Essix A+ decreased, while the water absorption ability of all materials; the water solubility of Duran, Essix A+, and Essix ACE; and the surface hardness of Duran and Essix A+ increased. The flexure modulus for the 0.5-mm-thick Duran, Essix A+, and eCligner specimens increased, whereas that for the 0.75-/1.0-mm-thick Duran and eClginer specimens decreased. In addition, the elastic modulus increased for the 0.5-mm-thick Essix A+ specimens and decreased for the 0.75-mm-thick Duran and Essix ACE and the 1.0-mm-thick Essix ACE specimens. Conclusions: Our findings suggest that the physical and mechanical properties of thermoplastic materials used for the fabrication of TOAs should be evaluated after thermoforming in order to characterize their properties for clinical application.
참고문헌 (35)
- Martorelli M, "A comparison between customized clear and removable orthodontic appliances manufactured using RP and CNC techniques", Dent Mater, 29, pp.e1~e10 (2013)
- Ogaard B, "Orthodontic appliances and enamel demineralization. Part 1. Lesion development", Am J Orthod Dentofacial Orthop, 94, pp.68~73 (1988)[DOI]
- Featherstone J, "Dental caries: a dynamic disease process", Aust Dent J, 53, pp.286~291 (2008)[DOI]
- Joffe L, "Invisalign: early experiences", J Orthod, 30, pp.348~352 (2003)[DOI]
- Shalish M, "Adult patients' adjustability to orthodontic appliances. Part I: a comparison between Labial, Lingual, and <TEX>$Invisalign^{TM}$<", TEX>, 2012, pp.34, 724~730
- Kesling HD, "The philosophy of the tooth positioning appliance", Am J Orthod Oral Surg, 31, pp.297~304 (1945)[DOI]
- Sheridan JJ, "The Readers' Corner. 2. What percentage of your patients are being treated with Invisalign appliances?", J Clin Orthod, 38, pp.544~545 (2004)
- Nahoum H, "The vacuum formed dental contour appliance", N Y State Dent J, 9, pp.385~390 (1964)
- Ponitz RJ, "Invisible retainers", Am J Orthod, 59, pp.266~272 (1971)[DOI]
- Melkos AB, "Advances in digital technology and orthodontics: a reference to the Invisalign method", Med Sci Monit, 11, pp.PI39~PI42 (2005)
- Michele Cassetta, "The combined use of computer-guided, minimally invasive, flapless corticotomy and clear aligners as a novel approach to moderate crowding: A case report", 대한치과교정학회지, 47(2), pp.130~141 (2017), 대한치과교정학회
- Boyd RL, "The Invisalign system in adult orthodontics: mild crowding and space closure cases", J Clin Orthod, 34, pp.203~212 (2000)
- Kravitz ND, "How well does Invisalign work? A prospective clinical study evaluating the efficacy of tooth movement with Invisalign", Am J Orthod Dentofacial Orthop, 135, pp.27~35 (2009)[DOI]
- Lagravere MO, "The treatment effects of Invisalign orthodontic aligners: a systematic review", J Am Dent Assoc, 136, pp.1724~1729 (2005)[DOI]
- Barone S, "Computer aided modelling to simulate the biomechanical behaviour of customised orthodontic removable appliances", IJIDeM, 10, pp.387~400 (2016)
- Hahn W, "Torquing an upper central incisor with aligners--acting forces and biomechanical principles", Eur J Orthod, 32, pp.607~613 (2010)[DOI]
- Simon M, "Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization", Am J Orthod Dentofacial Orthop, 145, pp.728~736 (2014)[DOI]
- Xiaowei Li, "Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor", 대한치과교정학회지, 46(2), pp.65~72 (2016), 대한치과교정학회
- Rossini G, "Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review", Angle Orthod, 85, pp.881~889 (2015)[DOI]
- Dupaix RB, "Finite strain behavior of poly (ethylene terephthalate)(PET) and poly (ethylene terephthalate)-glycol (PETG)", Polymer, 46, pp.4827~4838 (2005)[DOI]
- Fang D, "Dynamic stress relaxation of orthodontic thermoplastic materials in a simulated oral environment", Dent Mater J, 32, pp.946~951 (2013)[DOI]
- Lombardo L, "Stress relaxation properties of four orthodontic aligner materials: A 24-hour in vitro study", Angle Orthod, 87, pp.11~18 (2017)[DOI]
- Ryokawa H, "The mechanical properties of dental thermoplastic materials in a simulated intraoral environment", Orthod Waves, 65, pp.64~72 (2006)[DOI]
- 민샘, "교정용 열가소성 재료의 두께와 변형량이 재료의 물리적 특성에 미치는 영향", 대한치과교정학회지, 40(1), pp.16~26 (2010), 대한치과교정학회
- Kwon JS, "Force delivery properties of thermoplastic orthodontic materials", Am J Orthod Dentofac Orthop, 133, pp.228~234 (2008)[DOI]
- 안효원, "A new type of clear orthodontic retainer incorporating multi-layer hybrid materials", 대한치과교정학회지, 45(5), pp.268~272 (2015), 대한치과교정학회
- International Organization for Standardization (ISO), "ISO 20795-2:2013 Dentistry-base polymers--Part 2: orthodontic base polymers", ISO, ~ (2013)
- Azhikannickal E, "Laser light transmission through thermoplastics as a function of thickness and laser incidence angle: experimental and modeling", J Manuf Sci Eng, 134, pp.061007~ (2012)[DOI]
- Kattan M, "Straininduced crystallization in uniaxially drawn PETG plates", J Appl Polym Sci, 81, pp.3405~3412 (2001)[DOI]
- Zhang N, "Preparation and characterization of thermoplastic materials for invisible orthodontics", Dent Mater J, 30, pp.954~959 (2011)[DOI]
- Gardner GD, "Wear comparison of thermoplastic materials used for orthodontic retainers", Am J Orthod Dentofacial Orthop, 124, pp.294~297 (2003)[DOI]
- Alexandropoulos A, "Chemical and mechanical characteristics of contemporary thermoplastic orthodontic materials", Aust Orthod J, 31, pp.165~170 (2015)
- Kohda N, "Effects of mechanical properties of thermoplastic materials on the initial force of thermoplastic appliances", Angle Orthod, 83, pp.476~483 (2013)[DOI]
- Brazel CS, "Fundamental principles of polymeric materials", pp.53~ (2010), John Wiley
- Elkholy F, "Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in-vitro study", Am J Orthod Dentofacial Orthop, 151, pp.407~415 (2017)[DOI]