Use of Reference Ear Plug to improve accuracy of lateral cephalograms generated from cone-beam computed tomography scans

2

조회

0

다운로드

6

인용

0

참고문헌

키워드

주제분류치의학

초록

Objective: The purpose of this study was to evaluate the effectiveness of the use of Reference Ear Plug (REP) during cone-beam computed tomography (CBCT)scan for the generation of lateral cephalograms from CBCT scan data. Methods:Two CBCT scans were obtained from 33 adults. One CBCT scan was acquired using conventional methods, and the other scan was acquired with the use of REP. Virtual lateral cephalograms created from each CBCT image were traced and compared with tracings of the real cephalograms obtained from the same subject. Results: CBCT scan with REP resulted in a smaller discrepancy between real and virtual cephalograms. In comparing the real and virtual cephalograms,no measurements significantly differed from real cephalogram values in case of CBCT scan with REP, whereas many measurements significantly differed in the case of CBCT scan without REP. Conclusion: Measurements from CBCTgenerated cephalograms are more similar to those from real cephalograms when REP are used during CBCT scan. Thus, the use of REP is suggested during CBCT scan to generate accurate virtual cephalograms from CBCT scan data.

Abstract

Objective: The purpose of this study was to evaluate the effectiveness of the use of Reference Ear Plug (REP) during cone-beam computed tomography (CBCT)scan for the generation of lateral cephalograms from CBCT scan data. Methods:Two CBCT scans were obtained from 33 adults. One CBCT scan was acquired using conventional methods, and the other scan was acquired with the use of REP. Virtual lateral cephalograms created from each CBCT image were traced and compared with tracings of the real cephalograms obtained from the same subject. Results: CBCT scan with REP resulted in a smaller discrepancy between real and virtual cephalograms. In comparing the real and virtual cephalograms,no measurements significantly differed from real cephalogram values in case of CBCT scan with REP, whereas many measurements significantly differed in the case of CBCT scan without REP. Conclusion: Measurements from CBCTgenerated cephalograms are more similar to those from real cephalograms when REP are used during CBCT scan. Thus, the use of REP is suggested during CBCT scan to generate accurate virtual cephalograms from CBCT scan data.

참고문헌 (20)

  1. Danforth RA, "3-D volume imaging for dentistry: a new dimension", J Calif Dent Assoc, 31, pp.817~823 (2003)
  2. Mozzo P, "A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results", Eur Radiol, 8, pp.1558~1564 (1998)
  3. Proffit WR, "Contemporary orthodontics", pp.201~202 (2007), Mosby
  4. Farman AG, "ALARA still applies", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 100, pp.395~397 (2005)
  5. Farman AG, "Development of imaging selection criteria and procedures should precede cephalometric assessment with cone-beam computed tomography", Am J Orthod Dentofacial Orthop, 130, pp.257~265 (2006)
  6. Moshiri M, "Accuracy of linear measurements from imaging plate and lateral cephalometric images derived from cone-beam computed tomography", Am J Orthod Dentofacial Orthop, 132, pp.550~560 (2007)
  7. Kumar V, "Com parison of conventional and cone beam CT syn the sized cephalograms", Dentomaxillofac Radiol, 36, pp.263~269 (2007)
  8. van Vlijmen OJ, "Comparison of cephalometric radiographs obtained from conebeam computed tomography scans and conventional radiographs", J Oral Maxillofac Surg, 67, pp.92~97 (2009)
  9. Kumar V, "In vivo comparison of conventional and cone beam CT synthesized cephalograms", Angle Orthod, 78, pp.873~879 (2008)
  10. Cattaneo PM, "Comparison between conventional and cone-beam computed tomography-generated cepha lograms", Am J Orthod Dentofacial Orthop, 134, pp.798~802 (2008)
  11. Cevidanes L, "Head orientation in CBCT-generated cephalograms", Angle Orthod, 79, pp.971~977 (2009)
  12. Hwang HS, "Ear plug for forming reference axis in CT taking", Korean Patent 10-0884354, ~ (2009)
  13. Dahlberg G, "Statistical methods for medical and biological students", pp.122~132 (1940), George Allen & Unwin Ltd
  14. Hassan B, "Accuracy of three-dimensional measurements obtained from cone beam computed tomography surfacerendered images for cephalometric analysis: influence of patient scanning position", Eur J Orthod, 31, pp.129~134 (2009)
  15. Ludlow JB, "Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 106, pp.106~114 (2008)
  16. White SC, "Update on the biological effects of ionizing radiation, relative dose factors and radiation hygiene", Aust Dent J, 57(Suppl 1), pp.2~8 (2012)
  17. Grauer D, "Accuracy and landmark error calculation using cone-beam computed tomo graphy- generated cephalograms", Angle Orthod, 80, pp.286~294 (2010)
  18. Ahlqvist J, "The effect of projection errors on angular measurements in cepha lometry", Eur J Orthod, 10, pp.353~361 (1988)
  19. Yoon YJ, "Effect of head rotation on lateral cephalometric radiographs", Angle Orthod, 71, pp.396~403 (2001)
  20. Mah J, "Three-dimensional craniofacial imaging", Am J Orthod Dentofacial Orthop, 126, pp.308~309 (2004)

관련 논문AI추천

Accuracy of three-dimensional cephalograms generated using a biplanar imaging system박하연, 이경민, 이재서, 조진형, 황현식대한치과교정학회 · 대한치과교정학회지2018 · 제48권 제5호A comparative study of the reproducibility of landmark identification on posteroanterior and anteroposterior cephalograms generated from cone-beam computed tomography scansAL JAWAD HUSSEIN MUSTAFA SADIQ, Eui-Ri Na, Hyeon-Shik Hwang, Kyung-Min Lee대한치과교정학회 · 대한치과교정학회지2019 · 제49권 제1호Application of Cone-Beam Computed Tomography-Generated Cephalograms in Children and Adolescents김지민, 마연주, 박범희, 소우주대한소아치과학회 · 대한소아치과학회지2024 · 제51권 제3호Cone beam computed tomography로 합성된 두부규격 방사선사진에서의 각 계측점의 정확도와 재현성에 관한 연구권대근, 민승기, 전인철, 팽준영대한구강악안면외과학회 · 대한구강악안면외과학회지2010 · 제36권 제2호Comparison of cone-beam computed tomography cephalometric measurements using a midsagittal projection and conventional two-dimensional cephalometric measurements문철현, Gung-Chol Lee, Pil-Kyo Jung대한치과교정학회 · 대한치과교정학회지2015 · 제45권 제6호Comparison of landmark position between conventional cephalometric radiography and CT scans projected to midsagittal plane김남국, 박재우, 장영일대한치과교정학회 · 대한치과교정학회지2008 · 제38권 제6호

이 간행물의 다른 논문 (5건)