Noise reduction in low-dose positron emission tomography (PET) is a well-researched topic aimed at reducing patient radiation doses and improving diagnosis. Software-based noise reduction mainly improves the contrast between regions by reducing the variation of the acquired image. However, it should be performed under appropriate parameters to reduce discrimination. We propose a method that derives optimal noise-reduction parameters using the multi-scale structural similarity index measure and visual information fidelity, which are metrics for image quality assessment. Simulation and experimental studies demonstrated the viability of the proposed algorithm. The contrast-to-noise ratio value of the denoised reconstruction slice, which was used as the optimal parameter, increased approximately three times compared to that of the low-dose slice while preserving the resolution. The results indicate that the proposed method successfully predicted the parameters according to the noise-reduction algorithm and PET system conditions in the sinogram domain. The proposed algorithm should help prevent misdiagnosis and provide standardized medical images for clinical application by performing appropriate noise reduction

Abstract

Noise reduction in low-dose positron emission tomography (PET) is a well-researched topic aimed at reducing patient radiation doses and improving diagnosis. Software-based noise reduction mainly improves the contrast between regions by reducing the variation of the acquired image. However, it should be performed under appropriate parameters to reduce discrimination. We propose a method that derives optimal noise-reduction parameters using the multi-scale structural similarity index measure and visual information fidelity, which are metrics for image quality assessment. Simulation and experimental studies demonstrated the viability of the proposed algorithm. The contrast-to-noise ratio value of the denoised reconstruction slice, which was used as the optimal parameter, increased approximately three times compared to that of the low-dose slice while preserving the resolution. The results indicate that the proposed method successfully predicted the parameters according to the noise-reduction algorithm and PET system conditions in the sinogram domain. The proposed algorithm should help prevent misdiagnosis and provide standardized medical images for clinical application by performing appropriate noise reduction

참고문헌 (40)

  1. M. N. Wernick, "Aarsvold, Emission tomography : the fundamentals of PET and SPECT" (2004), ELSEVIER
  2. J. Tsuchiya, "Deep learning-based image quality improvement of 18F-fluorodeoxyglucose positron emission tomography : a retrospective observational study", EJNMMI Phys, 8(1), pp.31~ (2021)[DOI]
  3. A. K. Shukla, "Positron emission tomography : an overview", J. Med. Phys, 31(1), pp.13~21 (2006)[DOI]
  4. S. Vandenberghe, "Recent developments in time-of-flight PET", EJNMMI Phys, 3, pp.3~ (2016)[DOI]
  5. P. J. Slomka, "Recent advances and future progress in PET instrumentation", Semin. Nucl. Med, 46(1), pp.5~19 (2016)[DOI]
  6. Y. -R. J. Wang, "Low-count whole-body PET", MRI restoration : and evaluation of dose reduction spectrum and five state-of-the-art artificial intelligence models, 2023, pp.50(5), 1337~1350
  7. K. K. M. Doss, "Utilizing deep learning techniques to improve image quality and noise reduction in preclinical low-dose PET images in the sonogram domain", Med. Phys, 51(1), pp.209~223 (2024)[DOI]
  8. Z. Bian, "Dynamic positron emission tomography image restoration via a kinetics-induced bilateral filter", PLoS One, 9(2), pp.e89282~ (2014)[DOI]
  9. L. I. Rudin, "Nonlinear total variation based noise removal algorithms", Phys. Nonlinear Phenom, 60(1–4), pp.259~268 (1992)[DOI]
  10. K. Dabov, "Image restoration by sparse 3D transform-domain collaborative filtering", Proc. SPIE 6812, Image Processing:Algorithms and Systems VI (2008)
  11. J. Dutta, "Non-local means denoising of dynamic PET images", PLoS One, 8(12), pp.e81390~ (2013)[DOI]
  12. S. Peltonen, "PET sonogram denoising by block-matching and 3D filtering", 2011 IEEE Nuclear Science Symposium Conference Record (2011)
  13. A. Le Pogam, "Denoising of PET images by combining wavelets and curvelets for improved preservation of resolution and quantitation", Med. Image Anal, 17(8), pp.877~891 (2013)[DOI]
  14. A. Bal, "An efficient wavelet and curvelet-based PET image denoising technique", Med. Biol. Eng. Comput, 57(12), pp.2567~2598 (2019)[DOI]
  15. S. Guo, "Graph filtering approach to PET image denoising", 2019 1st International Conference on Industrial Artificial Intelligence (IAI) (2019)
  16. A. Sanaat, "Projection space implementation of deep learning-guided low-dose brain PET imaging improves performance over implementation in image space", J. Nucl. Med, 61(9), pp.1388~1396 (2020)[DOI]
  17. A. Sano, "Denoising PET images for proton therapy using a residual U-net, Biomed", Phys. Eng. Express, 7(2), pp.025014~ (2021)[DOI]
  18. M. Zhu, "A generative adversarial network with"zeroshot"learning for positron image denoising", Sci. Rep, 13, pp.1051~ (2023)[DOI]
  19. K. G. Zeimpekis, "NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET", CT, 2022, pp.49(9), 3023~3032
  20. T. Sartoretti, "Fully automated computational measurement of noise in positron emission tomography", Eur. Radiol (2023)[DOI]
  21. M. Makitalo, "Optimal inversion of the generalized anscombe transformation for Poisson-Gaussian noise", IEEE Trans. Image Process, 22(1), pp.91~103 (2013)[DOI]
  22. C. Riddell, "Noise reduction in oncology FDG PET images by iterative reconstruction : a quantitative assessment", J. Nucl. Med, 42(9), pp.1316~1323 (2001)
  23. K. Kim, "Newly-designed adaptive non-blind deconvolution with structural similarity index in single-photon emission computed tomography", Nucl. Eng. Technol, 55, pp.4591~4596 (2023)[DOI]
  24. P. Wadhwa, "PET image reconstruction using physical and mathematical modelling for time of flight PET-MR scanners in the STIR library", Methods, 185(1), pp.110~119 (2021)[DOI]
  25. B. K. Cha, "Optimization method to predict optimal noise reduction parameters for the non-local means algorithm based on the scintillator thickness in radiography", Sensors, 23(24), pp.9803~ (2023)[DOI]
  26. A. A. Yahya, "BM3D image denoising algorithm based on an adaptive filtering", Multimed. Tool. Appl, 79(4), pp.20391~20427 (2020)[DOI]
  27. K. Kim, "Block-matching and 3D filtering algorithm in X-ray image with photon counting detector using the improved K-edge subtraction method", Nucl. Eng. Technol, 01, pp.013~ (2024)[DOI]
  28. K. Imamura, "Image denoising using non-local means for Poisson noise", 2016 International Symposium on Intelligent Signal Processing and Communication Systems (ISPACS) (2016)
  29. M. Makitalo, "A closed-form approximation of the exact unbiased inverse of the anscombe variance-stabilizing transformation", IEEE Trans. Image Process, 20(9), pp.2697~2698 (2011)[DOI]
  30. L. A. Shepp, "Maximum likelihood reconstruction for emission tomography", IEEE Trans. Med. Imag, 1, pp.113~122 (1982)[DOI]
  31. C. Kamphuis, "Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithm", IEEE Trans. Med. Imag, 17, pp.1101~1105 (1998)[DOI]
  32. Z. Wang, "Multiscale structural similarity for image quality assessment", The Thrity-Seventh Asilomar Conference on Signals, Systems & Computers (2003)
  33. H. R. Sheikh, "Image information and visual quality", IEEE Trans. Signal Process, 15(2), pp.430~444 (2006)[DOI]
  34. S. Jan, "Gate-Geant4 application for tomographic emission : a simulation toolkit for PET and SPECT", Phys. Med. Biol, 49(19), pp.4543~4561 (2004)[DOI]
  35. V. -V. Wettenhovi, "OMEGA-open-source emission tomography software", Phys. Med. Biol, 66(6), pp.065010~ (2021)[DOI]
  36. A. Skretting, "A method for on-site measurements of the effective spatial resolution in PET image volumes reconstructed with OSEM and Gaussian post-filters", Radiat. Protect. Dosim, 139(1–3), pp.195~198 (2010)[DOI]
  37. C. M. Michail, "Assessment of the contrast to noise ratio in PET scanners with Monte Carlo methods", J. Phys. : Conf. Ser, 637, pp.012019~ (2015)[DOI]
  38. E. Kang, "A deep convolutional neural network using directional wavelets for low-dose X-ray CT reconstruction", Med. Phys, 44(10), pp.e360~e375 (2017)[DOI]
  39. S. -K. Kang, "Noise2Noise improved by trainable wavelet coefficients for PET denoising", Electronics, 10(13), pp.1529~ (2021)[DOI]
  40. M. S. Tahaei, "MR-guided PET image denoising", 2016IEEE Nuclear Science Symposium, Medical Imaging Conference and RoomTemperature Semiconductor Detector Workshop (NSS, RTSD), pp.2016[DOI]

관련 논문AI추천

Newly-designed adaptive non-blind deconvolution with structural similarity index in single-photon emission computed tomography김규석, 이영진한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제12호Experimental study of noise level optimization in brain single-photon emission computed tomography images using non-local means approach with various reconstruction methods강성현, 이승완, 이영진한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제5호자기 지도 학습훈련 기반의 Noise2Void 네트워크를 이용한 PET 영상의 잡음 제거 평가: 팬텀 실험박찬록, 윤석환대한방사선과학회 · Journal of Radiological Science and Technology2021 · 제44권 제6호양전자 방출 단층 촬영을 위한 3차원 최대사후 영상재구성에서 평활 파라미터의 적응형 미세 조정엄소연, 이수진, 이재훈대한의용생체공학회 · 의공학회지2024 · 제45권 제5호Feasibility study of improved median filtering in PET/MR fusion images with parallel imaging using generalized autocalibrating partially parallel acquisition김재영, 박찬록, 안창현, 이영진한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제1호노이즈 레벨 및 유사도 평가 기반 저선량 조건의 전산화 단층 검사 영상에서의 비지역적 평균 알고리즘의 최적화김이준, 박수빈, 박수연, 서강현, 오윤지, 이영진, 이우석, 정하선대한방사선과학회 · Journal of Radiological Science and Technology2024 · 제47권 제1호

이 간행물의 다른 논문 (49건)