Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Objectives: This study compared the ability of several techniques to remove calcium hydroxide (CH) from the root canal and determined the influence of CH residues on the accuracy of the electronic apex locator. Materials and Methods: Root canals of 90 human maxillary lateral incisors with confirmed true working length (TWL) were prepared and filled with CH. The teeth were randomly assigned to one of the experimental groups according to the CH removal technique (n = 14): 0.9% saline; 0.9% saline + master apical file (MAF); 17% ethylenediamine tetraacetic acid (EDTA); 17% EDTA + MAF; 5.25% sodium hypochlorite (NaOCl); 5.25% NaOCl + MAF. Six teeth were used as negative control. After CH removal, the electronic working length was measured using Root-ZX (Morita Corp.) and compared with TWL to evaluate Root-ZX accuracy. All specimens were sectioned longitudinally, and the area of remaining CH (CH) and total canal area were measured using imaging software. Results: The EDTA + MAF and NaOCl + MAF groups showed better CH removal than other groups (p < 0.05). Root- ZX reliability to prevent overestimated working length to be > 85% within a tolerance of ± 1.0 mm (p < 0.05). There was strong negative correlation between amount of CH residues and EAL accuracy (r = -0.800 for ± 0.5 mm; r = -0.940 for ± 1.0 mm). Conclusions: The mechanical instrumentation improves the CH removal of irrigation solutions although none of the techniques removed the dressing completely. Residues of CH medication in root canals affected the accuracy of Root-ZX adversely.
Abstract
Objectives: This study compared the ability of several techniques to remove calcium hydroxide (CH) from the root canal and determined the influence of CH residues on the accuracy of the electronic apex locator. Materials and Methods: Root canals of 90 human maxillary lateral incisors with confirmed true working length (TWL) were prepared and filled with CH. The teeth were randomly assigned to one of the experimental groups according to the CH removal technique (n = 14): 0.9% saline; 0.9% saline + master apical file (MAF); 17% ethylenediamine tetraacetic acid (EDTA); 17% EDTA + MAF; 5.25% sodium hypochlorite (NaOCl); 5.25% NaOCl + MAF. Six teeth were used as negative control. After CH removal, the electronic working length was measured using Root-ZX (Morita Corp.) and compared with TWL to evaluate Root-ZX accuracy. All specimens were sectioned longitudinally, and the area of remaining CH (CH) and total canal area were measured using imaging software. Results: The EDTA + MAF and NaOCl + MAF groups showed better CH removal than other groups (p < 0.05). Root- ZX reliability to prevent overestimated working length to be > 85% within a tolerance of ± 1.0 mm (p < 0.05). There was strong negative correlation between amount of CH residues and EAL accuracy (r = -0.800 for ± 0.5 mm; r = -0.940 for ± 1.0 mm). Conclusions: The mechanical instrumentation improves the CH removal of irrigation solutions although none of the techniques removed the dressing completely. Residues of CH medication in root canals affected the accuracy of Root-ZX adversely.
참고문헌 (36)
- Hülsmann M, "Mechanical preparation of root canals: shaping goals, techniques and means", Endod Topics, 10, pp.30~76 (2005)
- Yared G, "Canal preparation using only one Ni-Ti rotary instrument: preliminary observations", Int Endod J, 41, pp.339~344 (2008)
- Vieyra JP, "Comparison of working length determination with radiographs and four electronic apex locators", Int Endod J, 44, pp.510~518 (2011)
- Martins JN, "Clinical efficacy of electronic apex locators: systematic review", J Endod, 40, pp.759~777 (2014)
- de Camargo EJ, "Influence of preflaring on the accuracy of length determination with four electronic apex locators", J Endod, 35, pp.1300~1302 (2009)
- Aggarwal V, "An in vitro evaluation of performance of two electronic root canal length measurement devices during retreatment of different obturating materials", J Endod, 36, pp.1526~1530 (2010)
- Kawashima N, "Root canal medicaments", Int Dent J, 59, pp.5~11 (2009)
- Siqueira JF Jr, "Bacterial reduction in infected root canals treated with 2.5% NaOCl as an irrigant and calcium hydroxide", camphorated paramonochlorophenol paste as an intracanal dressing, 2007, pp.33, 667~672
- Athanassiadis B, "The use of calcium hydroxide, antibiotics and biocides as antimicrobial medicaments in endodontics", Aust Dent J, 52(Supplement 1), pp.S64~S82 (2007)
- Kim SK, "Influence of calcium hydroxide intracanal medication on apical seal", Int Endod J, 35, pp.623~628 (2002)
- Vilela DD, "Evaluation of interference of calcium hydroxide-based intracanal medication in filling root canal systems", J Contemp Dent Pract, 12, pp.368~371 (2011)
- Nandini S, "Removal efficiency of calcium hydroxide intracanal medicament with two calcium chelators: volumetric analysis using spiral CT, an in vitro study", J Endod, 32, pp.1097~1101 (2006)
- Lambrianidis T, "Removal efficacy of various calcium hydroxide", chlorhexidine medicaments from the root canal, 2006, pp.39, 55~61
- Knee DM, "A quantitative assessment of efficacy of various calcium hydroxide removal techniques", J Endod, 32, pp.563~565 (2006)
- Lambrianidis T, "Removal efficiency of calcium hydroxide dressing from the root canal", J Endod, 25, pp.85~88 (1999)
- Wu MK, "The capability of two hand instrumentation techniques to remove the inner layer of dentine in oval canals", Int Endod J, 36, pp.218~224 (2003)
- Thomas AS, "The accuracy of the Root ZX electronic apex locator using stainless-steel and nickel-titanium files", J Endod, 29, pp.662~663 (2003)
- Plotino G, "Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex", Int Endod J, 39, pp.408~414 (2006)
- Hör D, "Ex vivo comparison of two electronic apex locators with different scales and frequencies", Int Endod J, 38, pp.855~859 (2005)
- Stoll R, "Effectiveness of four electronic apex locators to determine distance from the apical foramen", Int Endod J, 43, pp.808~817 (2010)
- Pascon EA, "An ex vivo comparison of working length determination by 3 electronic apex locators", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 108, pp.e147~e151 (2009)
- Erdemir A, "The influence of irrigating solutions on the accuracy of the electronic apex locator facility in the Tri Auto ZX handpiece", Int Endod J, 40, pp.391~397 (2007)
- Calt S, "Dentinal tubule penetration of root canal sealers after root canal dressing with calcium hydroxide", J Endod, 25, pp.431~433 (1999)
- Rödig T, "Efficacy of different irrigants in the removal of calcium hydroxide from root canals", Int Endod J, 43, pp.519~527 (2010)
- Salgado RJ, "Comparison of different irrigants on calcium hydroxide medication removal:microscopic cleanliness evaluation", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 107, pp.580~584 (2009)
- Duran-Sindreu F, "Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator", J Endod, 38, pp.236~239 (2012)
- D’Assunção FL, "Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: a preliminary report", Int Endod J, 43, pp.269~274 (2010)
- McDonald NJ, "The electronic determination of working length", Dent Clin North Am, 36, pp.293~307 (1992)
- Jenkins JA, "An in vitro evaluation of the accuracy of the root ZX in the presence of various irrigants", J Endod, 27, pp.209~211 (2001)
- Kang JA, "Accuracies of seven different apex locators under various conditions", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 106, pp.e57~e62 (2008)
- Saito T, "Electronic determination of root canal length by newly developed measuring device. Influences of the diameter of apical foramen, the size of K-file and the root canal irrigants", Dent Jpn (Tokyo), 27, pp.65~72 (1990)
- Ebrahim AK, "An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals", Aust Dent J, 52, pp.193~197 (2007)
- Goldberg F, "In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption", J Endod, 28, pp.461~463 (2002)
- Dunlap CA, "An in vivo evaluation of an electronic apex locator that uses the ratio method in vital and necrotic canals", J Endod, 24, pp.48~50 (1998)
- Haffner C, "Accuracy of electronic apex locators in comparison to actual lengthan in vivo study", J Dent, 33, pp.619~625 (2005)
- Pilot TF, "Determination of impedance changes at varying frequencies in relation to root canal file position and irrigant", J Endod, 23, pp.719~724 (1997)