본문 바로가기
KERD
논문KCI 등재

In vivo assessment of accuracy of Propex II, Root ZX II, and radiographic measurements for location of the major foramen

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Objectives: The aim of this in vivo study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF). Materials and Methods: Thirty-two premolars with single canals that required extraction were included. Following anesthesia, access, and initial canal preparation with size 10 and 15 K-flex files and SX and S1 rotary ProTaper files, the canals were irrigated with 2.5% sodium hypochlorite. The length of the root canal was verified 3 times for each tooth using the 2 apex locators and once using the radiographic technique. Teeth were extracted and the actual WL was determined using size 15 K-files under a × 25 magnification. The Biostat 4.0 program (AnalystSoft Inc.) was used for comparing the direct measurements with those obtained using radiographic technique and the apex locators. Pearson's correlation analysis and analysis of variance (ANOVA) were used for statistical analyses. Results: The measurements obtained using the visual method exhibited the strongest correlation with Root ZX II (r = 0.94), followed by Propex II (r = 0.90) and Ingle's technique (r = 0.81; p < 0.001). Descriptive statistics using ANOVA (Tukey's post hoc test) revealed significant differences between the radiographic measurements and both EALs measurements (p < 0.05). Conclusions: Both EALs presented similar accuracy that was higher than that of the radiographic measurements obtained with Ingle's technique. Our results suggest that the use of these EALs for MF location is more accurate than the use of radiographic measurements.

Abstract

Objectives: The aim of this in vivo study was to assess the accuracy of 2 third-generation electronic apex locators (EALs), Propex II (Dentsply Maillefer) and Root ZX II (J. Morita), and radiographic technique for locating the major foramen (MF). Materials and Methods: Thirty-two premolars with single canals that required extraction were included. Following anesthesia, access, and initial canal preparation with size 10 and 15 K-flex files and SX and S1 rotary ProTaper files, the canals were irrigated with 2.5% sodium hypochlorite. The length of the root canal was verified 3 times for each tooth using the 2 apex locators and once using the radiographic technique. Teeth were extracted and the actual WL was determined using size 15 K-files under a × 25 magnification. The Biostat 4.0 program (AnalystSoft Inc.) was used for comparing the direct measurements with those obtained using radiographic technique and the apex locators. Pearson's correlation analysis and analysis of variance (ANOVA) were used for statistical analyses. Results: The measurements obtained using the visual method exhibited the strongest correlation with Root ZX II (r = 0.94), followed by Propex II (r = 0.90) and Ingle's technique (r = 0.81; p < 0.001). Descriptive statistics using ANOVA (Tukey's post hoc test) revealed significant differences between the radiographic measurements and both EALs measurements (p < 0.05). Conclusions: Both EALs presented similar accuracy that was higher than that of the radiographic measurements obtained with Ingle's technique. Our results suggest that the use of these EALs for MF location is more accurate than the use of radiographic measurements.

참고문헌 (22)

  1. Tsesis I, "The precision of electronic apex locators in working length determination: a systematic review and meta-analysis of the literature", J Endod, 41, pp.1818~1823 (2015)
  2. Ricucci D, "Apical limit of root canal instrumentation and obturation, part 1. literature review", Int Endod J, 31, pp.384~393 (1998)
  3. Siqueira JF Jr, "Incidence of postoperative pain after intracanal procedures based on an antimicrobial strategy", J Endod, 28, pp.457~460 (2002)
  4. Kuttler Y, "Microscopic investigation of root apexes", J Am Dent Assoc, 50, pp.544~552 (1955)
  5. Ponce EH, "The cemento-dentino-canal junction, the apical foramen, and the apical constriction: evaluation by optical microscopy", J Endod, 29, pp.214~219 (2003)
  6. Dummer PM, "The position and topography of the apical canal constriction and apical foramen", Int Endod J, 17, pp.192~198 (1984)
  7. Martins JN, "Clinical efficacy of electronic apex locators: systematic review", J Endod, 40, pp.759~777 (2014)
  8. Kobayashi C, "New electronic canal measuring device based on the ratio method", J Endod, 20, pp.111~114 (1994)
  9. Kara Tuncer A, "Effect of working length measurement by electronic apex locator or digital radiography on postoperative pain: a randomized clinical trial", J Endod, 40, pp.38~41 (2014)
  10. de Vasconcelos BC, "Ex vivo evaluation of the accuracy of electronic foramen locators in root canals with an obstructed apical foramen", J Endod, 41, pp.1551~1554 (2015)
  11. Schneider SW, "A comparison of canal preparations in straight and curved root canals", Oral Surg Oral Med Oral Pathol, 32, pp.271~275 (1971)
  12. Ingle JI, "Endodontics", pp.179~ (1976), Lea & Febiger
  13. Guise GM, "In vitro comparison of three electronic apex locators", J Endod, 36, pp.279~281 (2010)
  14. Khandewal D, "Comparative evaluation of accuracy of 2 electronic Apex locators with conventional radiography: an ex vivo study", J Endod, 41, pp.201~204 (2015)
  15. Pagavino G, "A SEM study of in vivo accuracy of the Root ZX electronic apex locator", J Endod, 24, pp.438~441 (1998)
  16. Haffner C, "Accuracy of electronic apex locators in comparison to actual length--an in vivo study", J Dent, 33, pp.619~625 (2005)
  17. Vasconcelos BC, "Accuracy of five electronic foramen locators with different operating systems: an ex vivo study", J Appl Oral Sci, 21, pp.132~137 (2013)
  18. Real DG, "Accuracy of working length determination using 3 electronic apex locators and direct digital radiography", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 111, pp.e44~e49 (2011)
  19. Kim E, "An in vivo comparison of working length determination by only root-ZX apex locator versus combining root-ZX apex locator with radiographs using a new impression technique", Oral Surg Oral Med Oral Pathol Oral Radiol Endod, 105, pp.e79~e83 (2008)
  20. Plotino G, "Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and ProPex", Int Endod J, 39, pp.408~414 (2006)
  21. Parente LA, "Efficacy of electronic foramen locators in controlling root canal working length during rotary instrumentation", Braz Dent J, 26, pp.547~551 (2015)
  22. Somma F, "In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II", Int Endod J, 45, pp.552~556 (2012)

관련 논문AI추천

이 간행물의 다른 논문 (9건)