본문 바로가기
KERD
논문KCI 등재

Catheter detection by transthoracic echocardiography during placement of peripherally inserted central catheters: a real-time method for eliminating misplacement

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: Although guidelines and protocols are available for central venous access, existing methods lack specificity and sensitivity, especially when placing peripherally inserted central catheters (PICCs). We evaluated the feasibility of catheter detection in the right atrial cavity using transthoracic echocardiography (TTE) during PICC placement.Methods: This single-center, retrospective study included consecutive patients who underwent PICC placement between January 2022 and March 2023. TTE was performed to detect the arrival of the catheter in the right atrial cavity. Catheter misplacement was defined as an aberrant catheter position on chest x-ray (CXR). The primary endpoint was predicting catheter misplacement based on catheter detection in the right atrial cavity. The secondary endpoint was optimizing catheter placement and examining catheter-associated complications.Results: Of the 110 patients identified, 10 were excluded because of poor echogenicity and vein access failure. The remaining 100 patients underwent PICC placement with TTE. The catheter was visualized in the right atrial cavity in 90 patients. CXR exams revealed catheter misplacement in seven cases. Eight patients with catheter misplacement underwent the same procedure in the other arm. In two patients, PICC placement failed due to anatomical reasons. Catheter misplacement was detected using TTE with sensitivity, specificity, positive predictive value, and negative predictive value of 97% confidence interval (CI, 91.31%–99.36%), 90% CI (55.50%–99.75%), 99%, and 75%, respectively.Conclusions: TTE is a reliable tool for detecting catheter misplacement and optimizing catheter tip positioning during PICC placement.

Abstract

Background: Although guidelines and protocols are available for central venous access, existing methods lack specificity and sensitivity, especially when placing peripherally inserted central catheters (PICCs). We evaluated the feasibility of catheter detection in the right atrial cavity using transthoracic echocardiography (TTE) during PICC placement.Methods: This single-center, retrospective study included consecutive patients who underwent PICC placement between January 2022 and March 2023. TTE was performed to detect the arrival of the catheter in the right atrial cavity. Catheter misplacement was defined as an aberrant catheter position on chest x-ray (CXR). The primary endpoint was predicting catheter misplacement based on catheter detection in the right atrial cavity. The secondary endpoint was optimizing catheter placement and examining catheter-associated complications.Results: Of the 110 patients identified, 10 were excluded because of poor echogenicity and vein access failure. The remaining 100 patients underwent PICC placement with TTE. The catheter was visualized in the right atrial cavity in 90 patients. CXR exams revealed catheter misplacement in seven cases. Eight patients with catheter misplacement underwent the same procedure in the other arm. In two patients, PICC placement failed due to anatomical reasons. Catheter misplacement was detected using TTE with sensitivity, specificity, positive predictive value, and negative predictive value of 97% confidence interval (CI, 91.31%–99.36%), 90% CI (55.50%–99.75%), 99%, and 75%, respectively.Conclusions: TTE is a reliable tool for detecting catheter misplacement and optimizing catheter tip positioning during PICC placement.

참고문헌 (17)

  1. "Peripherally inserted central catheter (PICC) line placement" (2024)[URL]
  2. Royce B, "Use of C-arm fluoroscopy by nurses for placement of PICC lines", J Assoc Vasc Access, 14, pp.138~141 (2009)
  3. Storm ES, "Radiation doses from venous access procedures", Radiology, 238, pp.1044~1050 (2006)
  4. Andreucci M, "Side effects of radiographic contrast media : pathogenesis, risk factors, and prevention", Biomed Res Int, 2014, pp.741018~ (2014)
  5. Bedel J, "Guidewire localization by transthoracic echocardiography during central venous catheter insertion : a periprocedural method to evaluate catheter placement", Intensive Care Med, 39, pp.1932~1937 (2013)
  6. Díaz-Gómez JL, "Point-of-care ultrasonography", N Engl J Med, 385, pp.1593~1602 (2021)
  7. Santos FK, "Peripherally inserted central catheter versus central venous catheter for intravenous access: a protocol for systematic review and meta-analysis", Medicine (Baltimore), 99, pp.e20352~ (2020)
  8. Johansson E, "Advantages and disadvantages of peripherally inserted central venous catheters(PICC)compared to other central venous lines : a systematic review of the literature", Acta Oncol, 52, pp.886~892 (2013)
  9. Greca A, "ECHOTIP : a structured protocol for ultrasound-based tip navigation and tip location during placement of central venous access devices in adult patients", J Vasc Access, 24, pp.535~544 (2023)
  10. Brescia F, "The SIP protocol update : Eight strategies, incorporating Rapid Peripheral Vein Assessment(RaPeVA), to minimize complications associated with peripherally inserted central catheter insertion", J Vasc Access, 25, pp.5~13 (2024)
  11. Emoli A, "The ISP(Safe Insertion of PICCs)protocol : a bundle of 8 recommendations to minimize the complications related to the peripherally inserted central venous catheters(PICC)", Assist Inferm Ric, 33, pp.82~89 (2014)
  12. Tomaszewski KJ, "Time and resources of peripherally inserted central catheter insertion procedures : a comparison between blind insertion", chest X-ray and a real time tip navigation and confirmation system, 2017, pp.9, 115~125
  13. Gullo G, "Appropriateness of replacing fluoroscopic guidance with ECG-electromagnetic guidance for PICC insertion : a randomized controlled trial", AJR Am J Roentgenol, 216, pp.981~988 (2021)
  14. Kwon S, "Outcomes of bedside peripherally inserted central catheter placement : a retrospective study at a single institution", Acute Crit Care, 35, pp.31~37 (2020)
  15. Ploton G, "A STROBE cohort study of 755 deep and superficial upper-extremity vein thrombosis", Medicine (Baltimore), 99, pp.e18996~ (2020)
  16. Moran J, "Screening for novel risk factors related to peripherally inserted central catheter-associated complications", J Hosp Med, 9, pp.481~489 (2014)
  17. Sharifkazemi M, "Three simple but interesting transthoracic echocardiographic road maps for proximal superior vena cava visualisation in healthy young adults", Int J Cardiol Heart Vasc, 39, pp.101004~ (2022)

관련 논문AI추천

Outcomes of bedside peripherally inserted central catheter placement: a retrospective study at a single institution권수경, 김장용, 김지일, 김향경, 문인성, 손수미, 이슬희, Joung Hee Kim대한중환자의학회 · Acute and Critical Care2020 · 제35권 제1호중심정맥도관의 우심방 내 위치 분류 및 예측배진건, 이창재, 이태림, 정태녕한국보건정보통계학회 · 보건정보통계학회지2015 · 제40권 제3호중심정맥도관삽입술 시 위치이상의 빈도를 감소시키는 방법: 실시간 초음파 유도하 재위치유승, 김건동, 김승환, 안홍준, 유연호, 유인술, 이진웅, 정원준, 조용철, 조별님희대한중환자의학회 · Acute and Critical Care2013 · 제28권 제4호Direction of the J-Tip of the Guidewire to Decrease the Malposition Rate of an Internal Jugular Vein Catheter유인술, Byeong jun Ahn, Jin Woong Lee, Seung Ryu, Sung Uk Cho, Won Joon Jeong, Yeon Ho You, Yong chul Cho대한중환자의학회 · Acute and Critical Care2015 · 제30권 제4호Guide Wire Entrapment during Central Venous Catheterization김경우, 김경태, 김정원, 김준현, 김지연, 박세혁, 박장수, 이상일, 최원주대한중환자의학회 · Acute and Critical Care2014 · 제29권 제2호오른쪽 빗장밑정맥을 이용한 중심정맥도관술에서 환자의 키와 중심정맥카테터의 깊이의 관계김기태, 이동준, 임의재대한중환자의학회 · Acute and Critical Care2011 · 제26권 제3호

이 간행물의 다른 논문 (16건)