Wear, microleakage and plastic deformation of an implantsupported chair-side bar system

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

PURPOSE. This in-vitro study was designed to evaluate retention forces, microleakage and plastic deformation of a prefabricated 2-implant bar attachment system (SFI-Bar, Cendres+Métaux, Switzerland). MATERIALS AND METHODS. Two SFI implant-adapters were torqued with 35 Ncm into two implant analogues. Before the tube bars were finally sealed, the inner cavity of the tube bar was filled with liquid red dye to evaluate microleakage. As tube bar sealing agents three different materials were used (AGC Cem (AGC, resin based), Cervitec Plus (CP; varnish) and Gapseal (GS; silicone based). Four groups with eight specimens each were tested (GS, GS+AGC, AGC, CP). For cyclic loading, the attachment system was assembled parallel to the female counterparts in a chewing simulator. The mean retention forces of the initial and final ten cycles were statistically evaluated (ANOVA, α≤.05). RESULTS. All groups showed a significant loss of retention forces. Their means differed between 30-39 N initially and 22-28 N after 50,000 loading cycles. No significant statistical differences could be found between the groups at the beginning (P=.224), at the end (P=.257) or between the loss of retention forces (P=.288). Microleakage occurred initially only in some groups but after 10,000 loading cycles all groupsexhibited microleakage. CONCLUSION. Long-term retention forces of the SFI-Bar remained above 20 N which can be considered clinically sufficient. The sealing agents in this study are not suitable to prevent microleakage.

Abstract

PURPOSE. This in-vitro study was designed to evaluate retention forces, microleakage and plastic deformation of a prefabricated 2-implant bar attachment system (SFI-Bar, Cendres+Métaux, Switzerland). MATERIALS AND METHODS. Two SFI implant-adapters were torqued with 35 Ncm into two implant analogues. Before the tube bars were finally sealed, the inner cavity of the tube bar was filled with liquid red dye to evaluate microleakage. As tube bar sealing agents three different materials were used (AGC Cem (AGC, resin based), Cervitec Plus (CP; varnish) and Gapseal (GS; silicone based). Four groups with eight specimens each were tested (GS, GS+AGC, AGC, CP). For cyclic loading, the attachment system was assembled parallel to the female counterparts in a chewing simulator. The mean retention forces of the initial and final ten cycles were statistically evaluated (ANOVA, α≤.05). RESULTS. All groups showed a significant loss of retention forces. Their means differed between 30-39 N initially and 22-28 N after 50,000 loading cycles. No significant statistical differences could be found between the groups at the beginning (P=.224), at the end (P=.257) or between the loss of retention forces (P=.288). Microleakage occurred initially only in some groups but after 10,000 loading cycles all groupsexhibited microleakage. CONCLUSION. Long-term retention forces of the SFI-Bar remained above 20 N which can be considered clinically sufficient. The sealing agents in this study are not suitable to prevent microleakage.

참고문헌 (22)

  1. Thompson GW, "The impact of the demographics of aging and the edentulous condition on dental care services", J Prosthet Dent, 79, pp.56~59 (1998)
  2. van Waas MA, "Treatment results with immediate overdentures: an evaluation of 4.5years", J Prosthet Dent, 76, pp.153~157 (1996)
  3. Bergendal T, "Implant-supported overdentures: a longitudinal prospective study", Int J Oral Maxillofac Implants, 13, pp.253~262 (1998)
  4. Timmerman R, "An eight-year follow-up to a randomized clinical trial of participant satisfaction with three types of mandibular implant-retained overdentures", J Dent Res, 83, pp.630~633 (2004)
  5. MacEntee MI, "A clinical trial of patient satisfaction and prosthodontic needs with ball and bar attachments for implant-retained complete overdentures: three-year results", J Prosthet Dent, 93, pp.28~37 (2005)
  6. Zou D, "A 3-year prospective clinical study of telescopic crown, bar, and locator attachments for removable four implant-supported maxillary overdentures", Int J Prosthodont, 26, pp.566~573 (2013)
  7. Albrecht D, "Space requirement of a prefabricated bar on two interforaminal implants: a prospective clinical study", Clin Oral Implants Res, 26, pp.143~148 (2015)
  8. Petropoulos VC, "Comparison of retention and release periods for implant overdenture attachments", Int J Oral Maxillofac Implants, 12, pp.176~185 (1997)
  9. Leung T, "Retention profiles of stud-type precision attachments", Int J Prosthodont, 4, pp.175~179 (1991)
  10. Dudic A, "Retention mechanisms and prosthetic complications of implant-supported mandibular overdentures:long-term results", Clin Implant Dent Relat Res, 4, pp.212~219 (2002)
  11. Naert I, "Prosthetic aspects and patient satisfaction with two-implant-retained mandibular overdentures:a 10-year randomized clinical study", Int J Prosthodont, 17, pp.401~410 (2004)
  12. Naert I, "A 10-year randomized clinical trial on the influence of splinted and unsplinted oral implants retaining mandibular overdentures:peri-implant outcome", Int J Oral Maxillofac Implants, 19, pp.695~702 (2004)
  13. Ender A, "Accuracy of complete-arch dental impressions:a new method of measuring trueness and precision", J Prosthet Dent, 109, pp.121~128 (2013)
  14. Wee AG, "Strategies to achieve fit in implant prosthodontics: a review of the literature", Int J Prosthodont, 12, pp.167~178 (1999)
  15. Deslis A, "Numerical investigations of the loading behaviour of a prefabricated non-rigid bar system", Ann Anat, 194, pp.538~544 (2012)
  16. Kobayashi M, "Effects of in vitro cyclic dislodging on retentive force and removal torque of three overdenture attachment systems", Clin Oral Implants Res, 25, pp.426~434 (2014)
  17. 하승룡, "Implant-supported overdenture with prefabricated bar attachment system in edentulous mandibular patient", The Journal of Advanced of Prosthodontics, 4(4), pp.254~258 (2012), 대한치과보철학회[DOI]
  18. 김하영, "Immediate loading on mandibular edentulous patient with SFI Bar� overdenture", The Journal of Advanced of Prosthodontics, 3(1), pp.47~50 (2011), 대한치과보철학회
  19. Bayer S, "Investigation of the wear of prefabricated attachments--an in vitro study of retention forces and fitting tolerances", Quintessence Int, 38, pp.e229~e237 (2007)
  20. Wolf K, "Analysis of retention and wear of ball attachments", Quintessence Int, 40, pp.405~412 (2009)
  21. Harder S, "Molecular leakage at implant-abutment connectionin vitro investigation of tightness of internal conical implantabutment connections against endotoxin penetration", Clin Oral Investig, 14, pp.427~432 (2010)
  22. Fritzemeier C, "Prophylaxis of periimplantitis by sealing of the implant interiors with GapSeal®", Implantol, 15, pp.71~79 (2007)

관련 논문AI추천

Comparison of changes in retentive force of three stud attachments for implant overdentures김수민, 윤미정, 이소현, 전영찬, 정창모, 최재원, 허중보대한치과보철학회 · The Journal of Advanced of Prosthodontics2015 · 제7권 제4호Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro studyAna Cristina Braga, Antonio Sergio Silva, Carlos Aroso, Esther Berastegui, Jose Manuel Mendes, Raul Ustrell, Tomas Escuin대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제1호Pull-off resistance of a screwless implant-abutment connection and surface evaluation after cyclic loadingAhmed Othman, Constantin von See, Gergo Mitov, Richard Mosch, Vasilios Alevizakos대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제3호Stress analysis of mandibular implant overdenture with locator and bar/clip attachment: Comparative study with differences in the denture base length권긍록, 노관태, 백장현, 유진숙, 이현종대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제3호Effect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis김석규, 손미경, 정재헌대한치과보철학회 · The Journal of Advanced of Prosthodontics2015 · 제7권 제3호Immediate loading on mandibular edentulous patient with SFI Bar® overdenture김진영, 김하영, 신상완, 정창모, 허중보, Sarah Qadeer대한치과보철학회 · The Journal of Advanced of Prosthodontics2011 · 제3권 제1호

이 간행물의 다른 논문 (9건)