Empirical study of alginate impression materials by customized proportioning system
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
PURPOSE. Alginate mixers available in the market do not have the automatic proportioning unit. In this study, an automatic proportioning unit for the alginate mixer and controller software were designed and produced for a new automatic proportioning unit. With this device, it was ensured that proportioning operation could arrange weight-based alginate impression materials. MATERIALS AND METHODS. The variation of coefficient in the tested groups was compared with the manual proportioning. Compression tension and tear tests were conducted to determine the mechanical properties of alginate impression materials. The experimental data were statistically analyzed using one way ANOVA and Tukey test at the 0.05 level of significance. RESULTS. No statistically significant differences in modulus of elastisity (P>0.3), tensional/compresional strength (P>0.3), resilience (P>0.2), strain in failure (P>0.4), and tear energy (P>0.7) of alginate impression materials were seen. However, a decrease in the standard deviation of tested groups was observed when the customized machine was used. To verify the efficiency of the system, powder and powder/water mixing were weighed and significant decrease was observed. CONCLUSION. It was possible to obtain more mechanically stable alginate impression materials by using the custom-made proportioning unit.
Abstract
PURPOSE. Alginate mixers available in the market do not have the automatic proportioning unit. In this study, an automatic proportioning unit for the alginate mixer and controller software were designed and produced for a new automatic proportioning unit. With this device, it was ensured that proportioning operation could arrange weight-based alginate impression materials. MATERIALS AND METHODS. The variation of coefficient in the tested groups was compared with the manual proportioning. Compression tension and tear tests were conducted to determine the mechanical properties of alginate impression materials. The experimental data were statistically analyzed using one way ANOVA and Tukey test at the 0.05 level of significance. RESULTS. No statistically significant differences in modulus of elastisity (P>0.3), tensional/compresional strength (P>0.3), resilience (P>0.2), strain in failure (P>0.4), and tear energy (P>0.7) of alginate impression materials were seen. However, a decrease in the standard deviation of tested groups was observed when the customized machine was used. To verify the efficiency of the system, powder and powder/water mixing were weighed and significant decrease was observed. CONCLUSION. It was possible to obtain more mechanically stable alginate impression materials by using the custom-made proportioning unit.
참고문헌 (20)
- Satheesh B. Haralur, "Effect of alginate chemical disinfection on bacterial count over gypsum cast", The Journal of Advanced of Prosthodontics, 4(2), pp.84~88 (2012), 대한치과보철학회[DOI]
- Doubleday B, "Impression materials", Br J Orthod, 25, pp.133~140 (1998)
- Cook W, "Alginate dental impression materials: chemistry, structure, and properties", J Biomed Mater Res, 20, pp.1~24 (1986)
- Inoue K, "Effect of mixing method on rheological properties of alginate impression materials", J Oral Rehabil, 29, pp.615~619 (2002)
- Re D, "Mechanical properties of elastomeric impression materials: An in vitro comparison", Int J Dent, 2015, pp.428286~ (2015)
- "American National Standard", American Dental Association Specification No. 18. Alginate Impression Materials, 1992
- Vrijhoef MM, "Tear energy of impression materials", J Dent, 14, pp.175~177 (1986)
- Wilson HJ, "Some properties of alginate impression materials relevant to clinical practice", Br Dent J, 121, pp.463~467 (1966)
- Jörgensen KD, "A new method of recording the elastic recovery of dental impression materials", Scand J Dent Res, 84, pp.175~182 (1976)
- Frey G, "Effect of mixing methods on mechanical properties of alginate impression materials", J Prosthodont, 14, pp.221~225 (2005)
- Caswell CW, "Volumetric proportioning techniques for irreversible hydrocolloids: a comparative study", J Am Dent Assoc, 112, pp.859~861 (1986)
- Tüfekci K, "Effects of gamma radiation sterilization and strain rate on compressive behavior of equine cortical bone", J Mech Behav Biomed Mater, 34, pp.231~242 (2014)
- Sneed WD, "Tear strength of ten elastomeric impression materials", J Prosthet Dent, 49, pp.511~513 (1983)
- Hondrum SO, "Tear and energy properties of three impression materials", Int J Prosthodont, 7, pp.517~521 (1994)
- Chai J, "Clinically relevant mechanical properties of elastomeric impression materials", Int J Prosthodont, 11, pp.219~223 (1998)
- Webber RL, "The determination of tear energy of extensible materials of dental interest", J Biomed Mater Res, 2, pp.281~296 (1968)
- ADA, "Specification no. 19 for dental elastomeric impression material. Revised ANSI", ADA Specification no. 18-2004
- Lee YK, "Effect of fluoride addition on the properties of dental alginate impression materials", J Mater Sci Mater Med, 15, pp.219~224 (2004)
- Nallamuthu NA, "Some aspects of the formulation of alginate dental impression materials-setting characteristics and mechanical properties", Dent Mater, 28, pp.756~762 (2012)
- Gent AN, "On the Relation Between Indentation Hardness and Young’s Modulus", Transactions of The Institute of the Rubber Industry, 34, pp.47~57 (1958)