Digital Radiography 시스템을 사용하여 전 척추검사 시 차폐체 두께에 따른 유방피부선량 측정

Measurement of Breast Skin Dose According to Shield Thickness During Whole Spine Scanography Using Digital Radiography System

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Whole Spine Scanography (WSS) using the Digital Radiography (DR) system is an examination that requires whole body X-ray exposure, which involves more exposure to radiation for patients than other general radiographies. This can affect the occurrence of breast cancer. This research measured radiation dose when breasts were shield and not shield using the Auto Exposure Control (AEC) mode. The radiation dose without a shield was 1.540mGy, and that using a collimator was measured 0.506 mGy. Moreover, 0.733mGy was measured when 1 shield (0.3 mm) was used, and 0.523 μGy when 5 of them (1.5 mm) were used. The results showed that the radiation dose with 5 shields and the radiation dose with a collimator were similar. Moreover, 0.233mGy was measured when 8 shields (2.4 mm) were used. The standard deviation were 0.081 when using collimator and 0.014 when 5 shields were used. Also, when 8 shields were used, it was found to be 0.002. Most patients who go under a scoliosis test are children or young people who are highly sensitive to radiation. In the research results, the case where the organs sensitive to radiation, women’s breasts, were shielded showed more distinct differences compared to without shields. It is considered that using shields can provide more constant shield than using a collimator and lower the risk of breast cancer caused by exposure to radiation.

Abstract

Whole Spine Scanography (WSS) using the Digital Radiography (DR) system is an examination that requires whole body X-ray exposure, which involves more exposure to radiation for patients than other general radiographies. This can affect the occurrence of breast cancer. This research measured radiation dose when breasts were shield and not shield using the Auto Exposure Control (AEC) mode. The radiation dose without a shield was 1.540mGy, and that using a collimator was measured 0.506 mGy. Moreover, 0.733mGy was measured when 1 shield (0.3 mm) was used, and 0.523 μGy when 5 of them (1.5 mm) were used. The results showed that the radiation dose with 5 shields and the radiation dose with a collimator were similar. Moreover, 0.233mGy was measured when 8 shields (2.4 mm) were used. The standard deviation were 0.081 when using collimator and 0.014 when 5 shields were used. Also, when 8 shields were used, it was found to be 0.002. Most patients who go under a scoliosis test are children or young people who are highly sensitive to radiation. In the research results, the case where the organs sensitive to radiation, women’s breasts, were shielded showed more distinct differences compared to without shields. It is considered that using shields can provide more constant shield than using a collimator and lower the risk of breast cancer caused by exposure to radiation.

참고문헌 (23)

  1. Geijer H, "Digital radiography of scoliosis with a scanning method, radiation dose optimization", Eur Radiol, 13(3), pp.543~551 (2003)
  2. Nash CL, "Risks of exposure to X-rays in patients undergoing longterm treatment for scoliosis", J Bone Joint Surg Am, 61(3), pp.371~374 (1979)
  3. Induction of thyroid cancer by ionizing radiation, "National Council on Radiation Protection and Measurements", National Council on Radiation Protection and Measurements, ~, pp.Report No 80 (1985)
  4. "ICRP Publication 60 : 1990 Recommendations of the International Commission on Radiological Protection", Elsevier Health Sciences, ~ (1991)
  5. Hoffman DA, "Breast Cancer in Women With Scoliosis Exposed to Multiple Diagnostic X Rays", J of the National Cancer Institute, 81(17), pp.1307~1316 (1989)
  6. Ronckers CM, "Multiple Diagnostic X-rays for Spine Deformities and Risk of Breast Cancer", Cancer Epidemiol Biomarkers Prev, 17(3), pp.605~613 (2008)
  7. Anual report of cancer statistics in Korea in 2002, , National cancer center, Korea central cancer, ~ (2007)
  8. Anual report of cancer statistics in Korea in 2004, , National cancer center, Korea central cancer, ~ (2008)
  9. Anual report of cancer statistics in Korea in 20105, , National cancer center, Korea central cancer, ~ (2018)
  10. Land CE, "Incidence of female breast cancer among atomic bomb survivors, Hiroshima and Nagasaki, 1950–1990", Radiat Res, 160, pp.707~717 (2003)
  11. Doody MM, "Breast cancer mortality after diagnostic radiography: findings from the U.S. Scoliosis Cohort Study", Spine (Phila Pa 1976), 25, pp.2052~2063 (2000)
  12. Mekis N, "The effect of breast shielding during lumbar spine radiography", Radiol Oncol, 47(1), pp.26~31 (2013)
  13. No authors listed, "The 2007 recommendations of the International Commission on Radiological Protection", Ann ICRP, 1~332 (2007)
  14. Wrixon AD, "New ICRP recommendations", J Radiol Prot, 28, pp.161~168 (2008)
  15. Busch HP, "Image Quality And Dose Management In Digital Radiography: A New Paradigm For Optimisation", Radiation Protection Dosimetry, 117(1-3), pp.143~147 (2005)
  16. Busch HP, "Need for new optimization strategies in CR and direct Digital radiography", Radiation Protection Dosimetry, 90(1-2), pp.31~33 (2000)
  17. Kim JS, "Evaluation of the Patient Effective Dose in Whole Spine Scanography Based on the Automatic Image Pasting Method for Digital Radiography", Iran J Radiol, 13(1), pp.1~7 (2016)
  18. 황준호, "DR(Digital Radiography)에서 자동노출제어장치의 감도변화에 따른 검출기 성능평가 시 흡수선량과 히스토그램을 통한정량적 분석방법에 관한 연구", 한국콘텐츠학회 논문지, 18(1), pp.232~240 (2018), 한국콘텐츠학회[DOI]
  19. "Guidelines for patient dose recommendations for general radiology examination", Ministry of Food and Drug Safety, 1~47 (2012)
  20. Investgation of patient dose for Diagnostic Reference Levels(DRL) in radiographic examination:National Survey in Korea, , Ministry of Food and Drug Safety, 14~28 (2011)
  21. 김기원, "복부팬텀을 이용한 SID 변화와 부가필터 유무에 따른 피폭선량에 관한 연구", 방사선기술과학, 39(3), pp.407~414 (2016), 대한방사선과학회[DOI]
  22. Mekis N, "The effect of breast shielding during lumbar spine radiography", Radilo Oncol, 47(1), pp.26~31 (2013)
  23. 김정민, "Whole Spine Scanography의 검사방향에 따른 환자 선량 평가", 방사선기술과학, 38(1), pp.1~6 (2015), 대한방사선과학회[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (9건)