The purpose of this paper is to recognize the usefulness of the Phantom produced with 3D printing technology by reproducing the original phantom with 3D printing technology. Using CT, we obtained information from the original phantom. The acquired file was printed by the SLA method of ABS materials. For inspection, SPECT/CT was used to obtain images. We filled the both Phantom with a solution mixed with 99mTcO4 1 mCi i n 1 liter of water a nd acq uired images in accordance with the standard protocol. Using Image J, the SNR for each slice of the image was obtained. As a reference images, AC images were used. For the analysis of acquired images, ROI was set in the White mater and Gray mater sections of each image, and the average Intensity Value within the ROI were compared. According to the results of this study, 3D printed phantom's SNR is about 0.1 higher than the conventional phantom. And the ratio of Intensity Value was shown in the original 1 : 3.4, and the printed phantom was shown to be 1 : 3.2. Therefore, if Calibration Value is applied, It is assumed that it can be used as an alternative to the original.

Abstract

The purpose of this paper is to recognize the usefulness of the Phantom produced with 3D printing technology by reproducing the original phantom with 3D printing technology. Using CT, we obtained information from the original phantom. The acquired file was printed by the SLA method of ABS materials. For inspection, SPECT/CT was used to obtain images. We filled the both Phantom with a solution mixed with 99mTcO4 1 mCi i n 1 liter of water a nd acq uired images in accordance with the standard protocol. Using Image J, the SNR for each slice of the image was obtained. As a reference images, AC images were used. For the analysis of acquired images, ROI was set in the White mater and Gray mater sections of each image, and the average Intensity Value within the ROI were compared. According to the results of this study, 3D printed phantom's SNR is about 0.1 higher than the conventional phantom. And the ratio of Intensity Value was shown in the original 1 : 3.4, and the printed phantom was shown to be 1 : 3.2. Therefore, if Calibration Value is applied, It is assumed that it can be used as an alternative to the original.

참고문헌 (13)

  1. Christensen A, "Maintaining safety and efficacy for 3D printing in medicine", 3D Print Med, 3, pp.1~ (2017)
  2. George F, "Measuring and establishing the accuracy & reproducibility of 3D-printed medical models", Radiographics, 37(5), pp.1424~1450 (2017)
  3. Epley BL, "Computer-generated patient models for reconstruction of cranial and facial deformities", 9(6), pp.548~566 (1998)
  4. Salmia M, "Accuracy of medical models made by additive manufacturing (rapid manufacturing)", Journal of Cranio-Maxillofacial Surgery, 41(7), pp.603~609 (2013)
  5. Charles W, "Chuck Hull: Apparatus for production of three-dimensional objects by stereolithography", US 4575330 A, ~ (1984)
  6. S. Scott Crump, "Apparatus and method for creating three-dimensional objects", US 5121329 A, ~ (1989)
  7. 강현대, "3D 프린팅 기술을 활용한 가구디자인 사례 분석 연구", 한국콘텐츠학회 논문지, 15(2), pp.177~186 (2015), 한국콘텐츠학회[DOI]
  8. Lim YS, "A study on the influence and potential of a 3D printer in the design and construction processes of the architectural industry", Doctorate, pp.2015
  9. "Printing Technology sparks Fly as NASA Pushes the Limits of 3-D" (2014)[URL]
  10. 한경진, "A 3D-printing Bone Model for Surgical Planning of Total Hip Replacement after Failed Triple Pelvic Osteotomy", 한국임상수의학회지, 34(6), pp.463~466 (2017), 한국임상수의학회[DOI]
  11. Veneziani GR, "Attenuation coefficient determination of printed ABS and PLA samples in diagnostic radiology standard beams", Journal of Physics : Conference Series, 733, pp.1~ (2016)
  12. Park HH, "Evaluation of nuclear medicine applications of 3D printing phantom -Gamma camera centrically", The Korean Journal of Nuclear Medicine Technology, 21(2), pp.65~73 (2017)
  13. Lee BI, "Quality Assurance and performance evaluation of PET", CT, 2008, pp.42(2), 137~144