Distance and Angular Dependence of Polycrystalline Semiconductor Dosimeters Based on Mercury(II) Iodide (HgI2) and Tallium(I) Bromide(TlBr) for an Iridium-192 Source: Experimental Measurements and Monte Carlo Simulation Comparison

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study evaluated the feasibility of polycrystalline semiconductor dosimeters (PSDs) fabricated from mercury (II) iodide (HgI2) and thallium (I) bromide (TlBr) for quality assurance in high-dose-rate (HDR) brachytherapy with an Iridium-192 (Ir-192) source. The PSDs were manufactured using screen-printing and annealing processes, and their performance was evaluated with respect distance and angular dependence. Experimental measurements were conducted with varying source-to-detector distances (1–8 ㎝) and angular positions (0°–60°), and the results were validated using Monte Carlo (MC) simulations with the GATE platform. Both PSDs exhibited dose responses following the inverse square law, with the HgI2-based PSD showing a slope closer to the theoretical value of -2, indicating higher sensitivity. In angular dependence, the TlBr-based PSD demonstrated superior stability, with smaller deviations over different angles compared to HgI2. The experimental data showed good agreement with MC simulations, with differences mostly generally 3–9%. These findings confirm that HgI2 and TlBr PSDs provide reliable, real-time, and high-resolution dose measurements, highlighting their potential as alternative detectors for brachytherapy quality assurance. Further investigations into long-term stability and dose-rate dependence are warranted for clinical implementation.

Abstract

This study evaluated the feasibility of polycrystalline semiconductor dosimeters (PSDs) fabricated from mercury (II) iodide (HgI2) and thallium (I) bromide (TlBr) for quality assurance in high-dose-rate (HDR) brachytherapy with an Iridium-192 (Ir-192) source. The PSDs were manufactured using screen-printing and annealing processes, and their performance was evaluated with respect distance and angular dependence. Experimental measurements were conducted with varying source-to-detector distances (1–8 ㎝) and angular positions (0°–60°), and the results were validated using Monte Carlo (MC) simulations with the GATE platform. Both PSDs exhibited dose responses following the inverse square law, with the HgI2-based PSD showing a slope closer to the theoretical value of -2, indicating higher sensitivity. In angular dependence, the TlBr-based PSD demonstrated superior stability, with smaller deviations over different angles compared to HgI2. The experimental data showed good agreement with MC simulations, with differences mostly generally 3–9%. These findings confirm that HgI2 and TlBr PSDs provide reliable, real-time, and high-resolution dose measurements, highlighting their potential as alternative detectors for brachytherapy quality assurance. Further investigations into long-term stability and dose-rate dependence are warranted for clinical implementation.

관련 논문AI추천

Geant 4 Monte Carlo simulation for I-125 brachytherapyElsayed A.M.M., Liu Jie, Medhat M.E.한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제7호Monte Carlo simulation of absorbed dose in water for the INTRABEAM electronic brachytherapy unit using a micro-diamond detector modelMalidarreh Roya Boudaghi, Zakaly Hesham M.H.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제5호Dosimetric and energy deposition evaluation of the dual-microcavity capsule structure for glioma 125I brachytherapy using Monte Carlo simulationGao Weida, Guan Zhongbao, Liang Yu, Li Dongjie, Yao Gang한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제7호유기 섬광체-광섬유 방사선 센서를 이용한 근접 방사선원 Ir-192의상대 깊이 선량율 측정김신, 문주현, 박병기, 서정기, 신상훈, 유욱재, 이봉수, 장경원, 조동현한국센서학회 · 센서학회지2008 · 제17권 제6호Comparison of Treatment Planning on Dosimetric Differences Between 192Ir Sources for High-Dose Rate Brachytherapy권경태, 김대용, 신성수, 안우상, 양오남, 이상호, 임청환, 최원식대한방사선과학회 · Journal of Radiological Science and Technology2016 · 제39권 제2호Validation of electromagnetic physics models and electron range in Geant4 Brachytherapy applicationAbabneh E., Albqoor A., Okoor S., Zahran I.한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제1호

이 간행물의 다른 논문 (12건)