본문 바로가기
KERD
논문KCI 등재

유기 섬광체-광섬유 방사선 센서를 이용한 근접 방사선원 Ir-192의상대 깊이 선량율 측정

Measurements of relative depth dose rates for a brachytherapy Ir-192 sourceusing an organic scintillator fiber-optic radiation sensor

1

조회

0

다운로드

4

인용

0

참고문헌

키워드

주제분류공학일반

초록

In this study, we have fabricated a fiber-optic radiation sensor using an organic scintillator and plastic optical fiber for brachytherapy dosimetry. Also, we have measured relative depth dose rates of Ir-192 source using a fiber-optic sensor and compared them with the results obtained using a conventional EBT film. Cerenkov lights which can be a noise in measuring scintillating light with a fiber-optic sensor are measured and eliminated by using of a background optical fiber. It is expected that a fiber-optic radiation sensor can be used in brachytherapy dosimetry due to its advantages such as a low cost, simple usage and a small volume.

Abstract

In this study, we have fabricated a fiber-optic radiation sensor using an organic scintillator and plastic optical fiber for brachytherapy dosimetry. Also, we have measured relative depth dose rates of Ir-192 source using a fiber-optic sensor and compared them with the results obtained using a conventional EBT film. Cerenkov lights which can be a noise in measuring scintillating light with a fiber-optic sensor are measured and eliminated by using of a background optical fiber. It is expected that a fiber-optic radiation sensor can be used in brachytherapy dosimetry due to its advantages such as a low cost, simple usage and a small volume.

참고문헌 (14)

  1. K. H. Beck, "A multichannel dosimeter based on scintillating fibers for medical application", Nucl. Instru. and Meth. in Phys. Res, 454(1), pp.147~151 (2000)
  2. C. Dilcher, "Dose mapping porcine coronary arteries using an optical fiber dosimeter", Cardiovascular Revascularization Medicine, 6(4), pp.163~169 (2005)
  3. T. Nakano, "High dose-rate brachytherapy source localization: positional resolution using a diamond detector", Phys. Med. Biol, 48(14), pp.2133~2146 (2003)
  4. B. Lee, "Polymer-polymer miscibility study for plastic gradient index optical fiber", Polymer Eng. and Sci, 40(9), pp.1996~1999 (2000)
  5. 황영묵, "무기 섬광체를 이용한 플라스틱 광섬유 방사선 센서부 제작 및 특성평가", 센서학회지, 14(4), pp.244~249 (2005), 한국센서학회
  6. J. Lambert, "A plastic scintillation dosimeter for high dose rate brachytherapy", Phys. Med. Biol, 51, pp.5505~5516 (2006)
  7. A. S. Beddar, "Water equivalent plastic scintillation detectors for high-energy beam dosimetry-. Properties and measurements", Phys. Med. Biol, 37(10), pp.1901~1913 (1992)
  8. 강세식, "방사선 치료학", pp.348~350 (2001), 도서출판 대학서림
  9. Khan. Faiz M., "The Physics of Radiation Therapy", pp.434~ (1994), Williams & Wilkins, Baltimore, Maryland, USA
  10. C. Mori (et al.), "Measurement of neutron and γ-ray intensity distributions with an optical fiber-scintillator detector", Nucl. Inst. and Meth. in Phys. Res, 422, pp.129~132 (1999)
  11. 장경원, "고에너지 전자선 측정을 위한 광섬유 방사선 센서에서의체렌코프 빛 측정 및 분석", 센서학회지, 15(3), pp.186~191 (2006), 한국센서학회
  12. A. S. Kirov, "An estimate of the perturbing effect of cerenkov radiation to the signal from a liquid scintillator volume irradiated by a megavoltage photon beam", Proc. of the 22nd Annual EMBS International Conf.,, pp.3148~3151 (2000), chicago, USA
  13. Frelin (et al.), "Spectral discrimination of Derenkov radiation in scintillating dosimeters", Med. Phys, 32(9), pp.3000~3006 (2005)
  14. 강세식, "Gafchromic EBT 필름을 이용한 선량분포의 평가", 방사선기술과학, 30(2), pp.139~145 (2007)

관련 논문AI추천

고 에너지 X-선 조사에 의한 광섬유 방사선량계와 이온 전리함의 심부선량 백분율 측정 및 비교문주현, 박병기, 서정기, 유욱재, 이봉수, 장경원, 조동현, 조영호, 허지연한국센서학회 · 센서학회지2009 · 제18권 제2호광섬유 방사선량계를 이용한 Co-60 방사선원으로부터 조사되는 상대선량측정김신, 박병기, 서정기, 신상훈, 유욱재, 이봉수, 장경원, 조동현, 허지연한국센서학회 · 센서학회지2010 · 제19권 제1호광섬유 방사선량계를 이용한 선량보강 영역에서의 심부선량 백분율과 피부 선량률 측정서정기, 유욱재, 이봉수, 장경원, 조동현, 조영호, 허지연한국광학회 · 한국광학회지2010 · 제21권 제1호광학 필터를 이용한 광섬유 방사선 센서의 체렌코프 빛 제거김신, 김형식, 이봉수, 이정한, 이정환, 장경원, 조동현, 조효성한국광학회 · 한국광학회지2006 · 제17권 제4호고 에너지 광자선 계측용 2차원 광섬유 방사선 센서의 제작 및 특성분석김신, 김형식, 신상훈, 이봉수, 이정한, 장경원, 조동현, 조효성한국광학회 · 한국광학회지2007 · 제18권 제4호1차원 광섬유 팬텀선량계를 이용한 치료용 광자선의 피부 및 선량보강영역에서 상대선량 측정문진수, 박장연, 서정기, 유욱재, 이봉수, 장경원, 조영호한국센서학회 · 센서학회지2011 · 제20권 제2호

이 간행물의 다른 논문 (9건)