Measurements and comparisons of PDDs using ion chamber and fiber-optic dosimeter irradiated by high energy photon beam
0
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
In this study, we have fabricated a fiber-optic dosimeter using an organic scintillator and a plastic optical fiber for measuring percentage depth dose with high energy X-ray beam. The scintillating light generated in organic sensor probe embedded in a solid water are guided by 20 m plastic optical fiber to the light-measuring device such as a photodiode- amplifier system. Using a fiber-optic dosimeter and an ion chamber, percentage depth dose curves are measured with 6 and 15 MV energies of X-ray beam whose field sizes are 2 cm × 2 cm and 10 cm × 10 cm.
Abstract
In this study, we have fabricated a fiber-optic dosimeter using an organic scintillator and a plastic optical fiber for measuring percentage depth dose with high energy X-ray beam. The scintillating light generated in organic sensor probe embedded in a solid water are guided by 20 m plastic optical fiber to the light-measuring device such as a photodiode- amplifier system. Using a fiber-optic dosimeter and an ion chamber, percentage depth dose curves are measured with 6 and 15 MV energies of X-ray beam whose field sizes are 2 cm × 2 cm and 10 cm × 10 cm.
참고문헌 (14)
- B. M. Rogina, "Application of optical fiber sensors for radiation dosimetry", Radiation Measurements, 26(4), pp.599~602 (1996)
- T. Aoyama, "Depth-dose measuring device using a multichannel scintillating fiber array for electron beam therapy", pp.1235~1239 (1997), Nagoya University College of Medical Technology
- K. W. Jang, "Development of two-dimensional fiber-optic radiation sensor for high energy photon beam therapy dosimetry", J. Nuclear Science and Technology, 5, pp.466~469 (2008)
- D. H. Cho, "Performance evaluation of one-dimensional fiber-optic radiation sensor for measuring high energy electron beam using a cahrge-coupled device", J. Nuclear Science and Technology, 5, pp.477~480 (2008)
- I. Kwan, "Comparison of the New MOSkin Detector and Fiber Optic Dosimetry System for Radiotherapy", J. Nuclear Science and Technology, 5, pp.518~521 (2008)
- A. S. Beddar, "A miniature “scintillator-fiberoptic-PMT” detector system for the dosimetry of small fields in streotactic radiosurgery", IEEE Trans. Nucl. Sci., 48(3), pp.924~928 (2001)
- L. Archambault, "Water-equivalent dosimeter array for small-field external beam radiotherapy", Med. Phys, 34, pp.1583~ (2007)
- A. S. Beddar, "plastic scintillation dosimetry and its application to radiotherapy", Radiation Measurements, pp.S124~ (2007)
- 황영묵, "무기 섬광체를 이용한 플라스틱 광섬유 방사선 센서부 제작 및 특성평가", 센서학회지, 14(4), pp.244~249 (2005), 한국센서학회
- 조동현, "의료용 선형가속기의 X-선 분포도 측정을 위한 1차원 광섬유 방사선 센서의 제작 및 성능평가", 센서학회지, 16(1), pp.33~38 (2007), 한국센서학회
- W. R. Hendee, "Radiation therapy physics, 3rd ed", pp.130~ (2005), John Wiley & Sons, Inc.
- A. S. Beddar, "Water-equivallent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical consideration", Phys. Med. Biol., 37, pp.1883~1900 (1992)
- A. S. Beddar, "Water-equivallent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements", Phys. Med. Biol., 37, pp.1901~1913 (1992)
- Faiz M. Khan, "The physics of radiation therapy, 2nd ed", pp.183~ (1994), Williams & Wilkins