민감도 향상과 센서 소형화를 위한 자기 및 용량형 안압센서의 개발
Development of Inductive and Capacitive Type Intraocular Pressure (IOP) Sensor to Improve Sensitivity and Minimize Size
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
We had presented an inductive type intraocular pressure sensor (L-sensor) in previous work. The distance between a micro coil and a ferrite on the membrane was modulated by pressure, and as a result the inductance and resonant frequency were changed. However, L-sensor has some problems to implant in eyes. First problem is low sensitivity. When L-sensor was implanted in rabbit’ eyes, resonant frequency of L-sensor was very hard to detect. Second problem is biocompatibility. Size of L-sensor is 6×7×1.2mm. When L-sensor was implanted in the eyes, it caused the inflammation. Therefore, this study suggests an inductive and capacitive type IOP sensor (LCsensor). The sensitivity of the LC-sensor 27.3 kHz/mmHg under 60mmHg. It is much larger than 14 kHz/mmHg of the L-sensor. And the size of LC-sensor is 47% smaller than L-sensor. After 2 weeks from the implantation of LC-sensor into rabbit eyes, we measured the changes of resonant frequency of LC-sensor according to increased IOP by Balanced Salt Solution (BSS) injection. As a result, the sensitivity of LC-sensor in in vivo test is 25 kHz/mmHg. That is similar to the sensitivity of in vitro test.
Abstract
We had presented an inductive type intraocular pressure sensor (L-sensor) in previous work. The distance between a micro coil and a ferrite on the membrane was modulated by pressure, and as a result the inductance and resonant frequency were changed. However, L-sensor has some problems to implant in eyes. First problem is low sensitivity. When L-sensor was implanted in rabbit’ eyes, resonant frequency of L-sensor was very hard to detect. Second problem is biocompatibility. Size of L-sensor is 6×7×1.2mm. When L-sensor was implanted in the eyes, it caused the inflammation. Therefore, this study suggests an inductive and capacitive type IOP sensor (LCsensor). The sensitivity of the LC-sensor 27.3 kHz/mmHg under 60mmHg. It is much larger than 14 kHz/mmHg of the L-sensor. And the size of LC-sensor is 47% smaller than L-sensor. After 2 weeks from the implantation of LC-sensor into rabbit eyes, we measured the changes of resonant frequency of LC-sensor according to increased IOP by Balanced Salt Solution (BSS) injection. As a result, the sensitivity of LC-sensor in in vivo test is 25 kHz/mmHg. That is similar to the sensitivity of in vitro test.
참고문헌 (23)
- M. Sherwood, "Ophthalmology, 2nd ed", pp.1413~1473 (2004), Mosby
- H. A. Q. a. A. T. Broman, "The number of people with glaucoma worldwide in 2010 and 2020", Br. J. Ophthalmol, 90(3), pp.262~267 (2006)
- R. E. P. Frenkel, "Comparison of the tono-pen to the goldmann applanation tonometer", Arch Ophthalmol, pp.750~753 (1988)
- E. Schneider, "Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry", J Glaucoma, 15, pp.2~6 (2006)
- F. A. a. J. M. D. Albert D M, "Principles and Practice of Ophthalmology" (1994), W.B. Saunders Company
- R. W. S., "Tonography in the clinical management of glaucoma", Trans Am Ophthalmol Soc, 53, pp.265~299 (1955)
- R. Puers, "Electrodeposited copper inductors for intraocular pressure telemetry", J Micromech Microeng, 10, pp.124~ (2000)
- G. Z. Chen, "Capacitive contact lens sensor for continuous non-invasive intraocular pressure monitoring", Sens. Actuator A-Phys., 203, pp.112~118 (2013)
- T. Chien-Kai, "Design and fabricate a contact lens sensor with a micro-inductor embedded for intraocular pressure monitoring", Sensors, 2012 IEEE, pp.1~4 (2012)
- M. Leonardi, "Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes", Acta Ophthalmologica, 87, pp.433~437 (2009)
- H. Yu-Chieh, "A contact lens sensor system with a micro-capacitor for wireless intraocular pressure monitoring", Sensors, 2013 IEEE, pp.1~4 (2013)
- G. Chitnis, "A minimally invasive implantable wireless pressure sensor for continuous IOP monitoring", IEEE T Bio-Med Eng, 60, pp.250~256 (2013)
- C. Po-Jui, "Implantable micromechanical parylenebased pressure sensors for unpowered intraocular pressure sensing", J Micromech Microeng, 17, pp.1931~ (2007)
- C. Po-Jui, "Implantable parylene-based wireless intraocular pressure sensor", Micro Electro Mechanical Systems, 2008. MEMS 2008. IEEE 21st International Conference on, pp.58~61 (2008)
- C. Po-Jui, "Wireless intraocular pressure sensing using microfabricated minimally invasive flexible-coiled LC sensor implant", J. Microelectromech. Syst, 19, pp.721~734 (2010)
- S. H. Cho, "A wireless powered fully integrated SU-8-based implantable LC transponder", MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 16권(호), pp.1657~1663 (2010), SPRINGER
- X. Ning, "A SU-8-based microfabricated implantable inductively coupled passive RF wireless intraocular pressure sensor", J. Microelectromech. Syst, 21, pp.1338~1346 (2012)
- 강대실, "마이크로 머시닝으로 제작한 기계적 가이드를 갖는 정전용량 선형 인코더", 센서학회지, 21(6), pp.440~445 (2012), 한국센서학회[DOI]
- 이인광, "의복착용형 무선 호흡모니터 시스템", 센서학회지, 17(2), pp.133~142 (2008), 한국센서학회
- 강병주, "A Wireless Intraocular Pressure Sensor with Variable Inductance Using a Ferrite Material", JOURNAL OF SEMICONDUCTOR TECHNOLOGY AND SCIENCE, 13(4), pp.355~360 (2013), 대한전자공학회
- J. C. H. Lin, "High quality factor parylene-based intraocular pressure sensor", Nano, 2012, 137~140
- O. Akar, "A wireless batch sealed absolute capacitive pressure sensor", Sens. Actuator APhys., 95, pp.29~38 (2001)
- T. H. Lee, "The Design of CMOS Radio-frequency Integrated Circuits, 2nd ed" (2004), Cambridge University Press