본문 바로가기
KERD
논문KCI 등재

금 나노입자를 이용한 새로운 당화혈색소의 검출 기술

Novel Detection Technology for Glycated Hemoglobin using Gold Nanoparticles

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류공학일반

초록

We report a novel detection technology for glycated hemoglobin (HbA1c) that is measured primarily to identify the three-month average plasma glucose concentration. In enzymatic measuring of glycated hemoglobin, the generated hydrogen peroxide was then used as a reducing agent of gold (III) for the synthesis of gold (0). Gold nanoparticles obtained from this novel approach were measured by optical and piezoelectric methods. In optical method, we have developed polymer based film-type sensor cartridge filled with all the reagents for glycated hemoglobin analysis and the cartridge worked very well having the detection limit of 0.53% of glycated hemoglobin. On the other hand, quartz crystal microbalance (QCM) sensors also have been developed to determine the abilities of surface modified QCM sensors at various levels of the concentration of glycated hemoglobin to bind gold nanoparticles and limit of detection was 0.90%. Finally, despite of relatively lower sensitivities of QCM sensor and film-type optical sensor than well-plate based optical detection, these two sensors were available to measure the glycated hemoglobin level for diabetes patients and normal person

Abstract

We report a novel detection technology for glycated hemoglobin (HbA1c) that is measured primarily to identify the three-month average plasma glucose concentration. In enzymatic measuring of glycated hemoglobin, the generated hydrogen peroxide was then used as a reducing agent of gold (III) for the synthesis of gold (0). Gold nanoparticles obtained from this novel approach were measured by optical and piezoelectric methods. In optical method, we have developed polymer based film-type sensor cartridge filled with all the reagents for glycated hemoglobin analysis and the cartridge worked very well having the detection limit of 0.53% of glycated hemoglobin. On the other hand, quartz crystal microbalance (QCM) sensors also have been developed to determine the abilities of surface modified QCM sensors at various levels of the concentration of glycated hemoglobin to bind gold nanoparticles and limit of detection was 0.90%. Finally, despite of relatively lower sensitivities of QCM sensor and film-type optical sensor than well-plate based optical detection, these two sensors were available to measure the glycated hemoglobin level for diabetes patients and normal person

참고문헌 (15)

  1. J. W. Baynes, "Role of oxidative stress in diabetic complications : a new perspective on an old paradigm", Diabetes, 48(1), pp.1~9 (1999)
  2. K. P. Peterson, "What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry", Clin. Chem, 44(9), pp.1951~1958 (1998)
  3. F. B. Araujo, "Evaluation of oxidative stress in patients with hyperlipidemia", Atherosclerosis, 117(1), pp.61~71 (1995)
  4. P. K. Singal, "The role of oxidative stress in the genesis of heart disease", Cardiovasc. Res, 40(3), pp.426~432 (1998)
  5. L. Liu, "Direct enzymatic assay for %HbA1c in human whole blood samples", Clin. Biochem, 41(7), pp.576~583 (2008)
  6. L. Liu, "Boronic acid-based electrochemical sensors for detection of Biomolecules", Int. J. Electrochem. Sci, 8(9), pp.11161~11174 (2013)
  7. V. V. Shumyantseva, "Electrochemical methods for detection of post-translational modifications of proteins", Biosens. Bioelectron, 61, pp.131~139 (2014)
  8. C. S. Pundir, "Determination of glycated hemoglobin with special emphasis on biosensing methods", Anal. Biochem, 444, pp.47~56 (2014)
  9. S. H. Ang, "Current aspects in hemoglobin A1c detection : A review", Clin. Chim. Acta, 439, pp.202~211 (2015)
  10. B. Wang, "Recent Progress in Electro-chemical HbA1c Sensors : A Review", Materials, 8, pp.1187~1203 (2015)
  11. C. E. L. Myhre, "Optical properties in the UV and visible spectral region of organic acids relevant to tropospheric aerosols", Atmos. Chem. Phys, 4, pp.1759~1769 (2004)
  12. K. A. Marx, "Quartz crystal microbalance : a useful tool for studying thin polymer films and complex biomolecular systems at the solution-surface interface", Biomacromolecules, 4(5), pp.1099~1120 (2003)
  13. D. S. Ballantine, "Acoustic Wave Sensors", pp.80~82 (1997), Academic Press
  14. D. Kim, "Microarray-based multiplexed scanometric immunoassay for protein cancer markers using gold nanoparticle probes", Anal. Chem, 81(21), pp.9183~9187 (2009)
  15. A. Shrivastava, "Methods for the determination of limit of detection and limit of quantitation of the analytical methods", Chronicles of Young Scientists, 2(1), pp.21~25 (2011)

관련 논문AI추천

이 간행물의 다른 논문 (11건)