본문 바로가기
KERD
논문KCI 등재

바이오센서를 위한 전하분할 방식 감지회로의 CMOS 구현

Development of PC-based and portable high speed impedance analyzer for biosensor

1

조회

0

다운로드

0

인용

0

참고문헌

주제분류공학일반

초록

Abstract

For more convenient electrode-electrolyte interface impedance analysis in biosensor, a stand-alone impedance measurement system is required. In our study, we developed a PC-based portable system to analyze impedance of the electrochemical cell using microprocessor. The devised system consists of signal generator, programmable amplifiers, A/D converter, low pass filter, potentiostat, I/V converter, microprocessor, and PC interface. As a microprocessor, PIC16F877 which has the processing speed of 5 MIPS was used. For data acquisition, the sampling rate at 40k samples/sec, resolution of 12 bit is used. RS-232 with 115.2 kbps speed is used for the PC communication. The square wave was used as stimuli signal for impedance analysis and voltage-controlled current measurement method of three-electrode-method were adopted. Acquired voltage and current data are calculated to multifrequency impedance signal after Fourier transform. To evaluate the implemented system, we set up the dummy cell as equivalent circuit of which was composed of resistor, parallel circuit of capacitor and resistor connected in parallel and measured the impedance of the dummy cell; the result showed that there exist accuracy within 5% errors and reproduction within 1% errors compared to output of Hioki LCR tester and HP impedance analyzer as a standard product. These results imply that it is possible to analyze electrode-electrolyte interface impedance quantitatively in biosensor and to implement the more portable high speed impedance analysis system compared to existing systems.

참고문헌 (9)

  1. "Some thoughts regarding electrical bioimpedance and its clinical use", 18, pp.221~3 (1993)
  2. "푸리에 변환을 이용한 전기화학식 임피던스 측정시스템 제작", 제6호, pp.1991~
  3. "Real time impedance plots with arbitrary frequency components", 20, pp.103~114 (1999)
  4. "An Electrochemical Impedance Measurement Technique Employing Fourier Transform", 72, pp.2035~2041 (2000)
  5. "Fourier transform Faradic admittance measurements-on the use of high-precision data for characterization of very rapid electrode preocess kinetic parameter", 45, pp.2401~3 (1973)
  6. "계면과 전극과정의 과학, 기술 전기화학" (2001), 청문각
  7. "Effect of skin temperature on multifrequency bioelectrical impedance analysis", 1, pp.1996~
  8. "재료전기화학개론" (2001), 시그마프레스
  9. "Bioimpedance and Bioelectricity Basic" (2000), Academic Press

관련 논문AI추천

이 간행물의 다른 논문 (20건)