A compact biosignal monitoring device was developed. Electrodes for electrocardiogram (ECG) and a LED and silicondetector for photoplethysmogram (PPG) were used. A lead II type was arranged for ECG measurement and reflected lightmeasurement protocol and processed measured signals. PPG and ECG had a sampling rate of 300 Hz with 8-bit resolution.The maximum power consumption was 100 mW. The microprocesor computed pulse transit time (PTT) between the R-wave of ECG and the peak of PPG. To increase the resolution of PT, analog peak detectors obtained the peaks of ECGand PPG whose interval was calculated using an internal clock cycle of 921.6 kHz. The device was designed to be operatedby 3-volt battery. Biosignals can be measured for 2~3 days continuously without the external interuptions and data isstored to an on-board memory. Our system was succesfuly tested with human subjects.

참고문헌 (9)

  1. "Design of Pulse Oximeters Institute of Physics Publishing Ltd", pp.40~55 (1997)
  2. "Respiratory variations in the reflection mode photoplethysmographic signal. Relationships to peripheral venous pressure", 41, pp.249~254 (2003)
  3. "Medical Instrumentation", pp.139~151 (1998)
  4. "광전용적맥파 센서를 이용한 맥파전달시간의 측정", 13(5), pp.383~391 (2004)
  5. "Continuous measurement of systolic blood pressure using the PTT and other parameters", pp.6153~6156 (2005)
  6. "Textbook of medical physiology" (1986), W. B. Saunders Company
  7. "Mobile monitoring with wearable photoplethysmographic biosensors Engineering in Medicine and Biology Magazine", 22(3), pp.28~40 (2003)
  8. "A novel mobile transtelephonic system with synthesized 12-lead ECG", 8(4), pp.428~438 (2004)
  9. "압력맥파 및 광전용적맥파 검출용 일체형 센서의 구현 및 평가", 15(2), pp.106~111 (2006)