The purpose of this study is to obtain brain MRI images through Heavenly T2 FLAIR and DWItechniques to find out strengths and weaknesses of each image. Data were analyzed on 13 normal peopleand 17 brain tumor patients. Philips Ingenia 3.0TCX was used as the equipment used for the inspection,and 32 Channel Head Coil was used to acquire data. Using Image J and Infinity PACS Data, 3 mm2 of graymatter, white matter, cerebellum, basal ganglia, and tumor areas were set and measured. Quantitativeanalysis measured SNR and CNR as an analysis method, and qualitative analysis evaluated overallimage quality, lesion conspicuity, image distortion, susceptibility artifact and ghost artifact on a 5-pointscale. The statistical significance of data analysis was that Wilcox-on Signed Rank Test and Paired t-testwere executed, and the statistical program used was SPSS ver.22.0 and the p value was less than 0.05. In quantitative analysis, the SNR of gray matter, white matter, cerebellum, basal ganglia, and tumorof Heavily T2 FLAIR is 41.45±0.13, 40.52±0.45, 41.44±0.51, 40.96±0.09, 35.28±0.46 and the CNR is15.24±0.13, 16.75±0.23, 16.28±0.41, 15.83±0.17, 16.63±0.51. In DWI, SNR is 32.58±0.22, 36.75±0.17,30.21±0.19, 35.83±0.11, 43.29±0.08, and CNR is 13.14±0.63, 14.21±0.31, 12.95±0.32, 11.73±0.09,17.56±0.52. In normal tissues, Heavenly T2 FLAIR obtained high results, but in disease evaluation, highresults were obtained at DWI, b=1000 (p<0.05). In addition, in the qualitative analysis, overall imagequality, lesion conspicuity, image distortion, susceptibility artifact and ghost artifact aspects of theHeavily T2 FLAIR were evaluated, and 3.75±0.28, 2.29±0.24, 3.86±0.23, 4.08±0.21, 3.79±0.22 valueswere found, respectively, and 2.53±0.39, 4.13±0.29, 1.90±0.20, 1.81±0.21, 1.52±0.45 in DWI. As a resultof qualitative analysis, overall image quality, image distortion, susceptibility artifact and ghost artifact wererated higher than DWI. However, DWI was evaluated higher in lesion conspicuity (p<0.05). In normaltissues, the level of Heavenly T2 FLAIR was higher, but the DWI technique was higher in the evaluation ofthe disease (tumor). The two results were necessary techniques depending on the normal site and thelocation of the disease. In conclusion, statistically significant results were obtained from the two techniques. In quantitative and qualitative analysis, the two techniques had advantages and disadvantages, andin normal and disease evaluation, the two techniques produced useful results. These results are believedto be educational data for clinical basic evaluation and MRI in the future.
Abstract
The purpose of this study is to obtain brain MRI images through Heavenly T2 FLAIR and DWItechniques to find out strengths and weaknesses of each image. Data were analyzed on 13 normal peopleand 17 brain tumor patients. Philips Ingenia 3.0TCX was used as the equipment used for the inspection,and 32 Channel Head Coil was used to acquire data. Using Image J and Infinity PACS Data, 3 mm2 of graymatter, white matter, cerebellum, basal ganglia, and tumor areas were set and measured. Quantitativeanalysis measured SNR and CNR as an analysis method, and qualitative analysis evaluated overallimage quality, lesion conspicuity, image distortion, susceptibility artifact and ghost artifact on a 5-pointscale. The statistical significance of data analysis was that Wilcox-on Signed Rank Test and Paired t-testwere executed, and the statistical program used was SPSS ver.22.0 and the p value was less than 0.05. In quantitative analysis, the SNR of gray matter, white matter, cerebellum, basal ganglia, and tumorof Heavily T2 FLAIR is 41.45±0.13, 40.52±0.45, 41.44±0.51, 40.96±0.09, 35.28±0.46 and the CNR is15.24±0.13, 16.75±0.23, 16.28±0.41, 15.83±0.17, 16.63±0.51. In DWI, SNR is 32.58±0.22, 36.75±0.17,30.21±0.19, 35.83±0.11, 43.29±0.08, and CNR is 13.14±0.63, 14.21±0.31, 12.95±0.32, 11.73±0.09,17.56±0.52. In normal tissues, Heavenly T2 FLAIR obtained high results, but in disease evaluation, highresults were obtained at DWI, b=1000 (p<0.05). In addition, in the qualitative analysis, overall imagequality, lesion conspicuity, image distortion, susceptibility artifact and ghost artifact aspects of theHeavily T2 FLAIR were evaluated, and 3.75±0.28, 2.29±0.24, 3.86±0.23, 4.08±0.21, 3.79±0.22 valueswere found, respectively, and 2.53±0.39, 4.13±0.29, 1.90±0.20, 1.81±0.21, 1.52±0.45 in DWI. As a resultof qualitative analysis, overall image quality, image distortion, susceptibility artifact and ghost artifact wererated higher than DWI. However, DWI was evaluated higher in lesion conspicuity (p<0.05). In normaltissues, the level of Heavenly T2 FLAIR was higher, but the DWI technique was higher in the evaluation ofthe disease (tumor). The two results were necessary techniques depending on the normal site and thelocation of the disease. In conclusion, statistically significant results were obtained from the two techniques. In quantitative and qualitative analysis, the two techniques had advantages and disadvantages, andin normal and disease evaluation, the two techniques produced useful results. These results are believedto be educational data for clinical basic evaluation and MRI in the future.
참고문헌 (22)
- Kim Jihyun, "자궁체부의 양성 및 악성 종양의 자기공명영상 소견과 감별 진단", 대한영상의학회지, 82(5), pp.1103~1123 (2021), 대한영상의학회[DOI]
- 강은주, "FLAIR 영상에서 영유아 뇌백질 신호강도의 시기적 변화", 대한영상의학회지, 62(5), pp.411~419 (2010), 대한영상의학회
- Lee YM, "Brain Diffusion-Weighted MRI : Comparison of Tetrahedral Versus Orthogonal Diffusion Brain", KSMRT, 22(1), pp.73~74 (2012)
- 김경민, "뇌 확산강조검사에서 기존 High b-value 확산강조영상과의 비교를 통한 Computed High b-value 확산강조영상의 진단적 유용성 평가", 대한자기공명기술학회지, 29(1), pp.1~10 (2019), 대한자기공명기술학회[DOI]
- 박종빈, "뇌줄기 확산강조영상에서 자화 감수성 인공물 감소를 위한 SENSE 인자에 관한 연구", 대한자기공명기술학회지, 33(1), pp.9~26 (2023), 대한자기공명기술학회[DOI]
- Holdsworth SJ, "Diffusion-weighted imaging with dual-echo echo-planar imaging for better sensitivity to acute stroke", AJNR Am. J. Neuroradiol, 35(7), pp.1293~1302 (2014)[DOI]
- Tsubouchi Y, "Use of high b value diffusion-weighted magnetic resonance imaging in acute encephalopathy", encephalitis during childhood, 2018, pp.40(2), 116~125
- 남궁인, "초기 확산강조 MR 영상에서 음성을 보인 급성기 뇌경색 환자의 임상 양상", 대한응급의학회지, 24(4), pp.396~402 (2013), 대한응급의학회
- 홍인기, "다발성 경화증 질환의 자기공명 T2 강조영상에서 단면 두께 변화에 따른 잡음 평가", 방사선기술과학, 44(4), pp.327~333 (2021), 대한방사선과학회
- 손순룡, "T-spine exhalation technique과 자연스런 호흡 상태에서 촬영하는 T-spine breathing technique의 정량적 평가", 한국산학기술학회논문지, 14(9), pp.4429~4436 (2013), 한국산학기술학회[DOI]
- 박지은, "뇌종양 영상의 현재와 미래", 대한영상의학회지, 81(3), pp.467~487 (2020), 대한영상의학회[DOI]
- Byun JH, "Patients with brain metastases the usefulness of contrast-enhanced FLAIR images after delay", KJDIM, 16(1), pp.13~19 (2014)
- 조은, "Isolated Acute Nontraumatic Cortical Subarachnoid Hemorrhage: Etiologies Based on MRI Findings", 대한영상의학회지, 74(6), pp.380~388 (2016), 대한영상의학회[DOI]
- 이원정, "흉부 CT촬영에서 저선량 프로토콜의 선량과 화질: 표준선량 프로토콜과 비교", 방사선방어학회지, 37(2), pp.84~89 (2012), 테스트학회[DOI]
- 노태관, "급성 허혈성 뇌경색 환자의 자기공명영상 검사 시 Echo Planar Image T2 FLAIR 기법의 유용성에 관한 연구", 한국방사선학회논문지, 14(7), pp.957~964 (2020), 한국방사선학회[DOI]
- Kim JT, "Usefulness of the High B-value DWI in Brain Tumors", KJDIM, 17(1), pp.33~41 (2015)
- Sant’Ovaia H, "Magnetic susceptibility and isothermal remanent magnetization in human tissues : a study case", Biometals, 28(6), pp.951~958 (2015)[DOI]
- Zhao Y, "Self-adapting multi-peak water-fat reconstruction for the removal of lipid artifacts in chemicalexchange saturation transfer(CEST)imaging", Magn. Reson. Med, 82(5), pp.1700~1712 (2019)[DOI]
- 구은회, "자기공명확산강조영상에서 SE-EPI 와 SSH-TSE 기법을 이용한 영상의 질 평가", 한국방사선학회논문지, 15(7), pp.991~998 (2021), 한국방사선학회[DOI]
- Kirchgesner T, "Fat suppression at 2D MR imaging of the hands : Dixon method versus CHESS technique and STIR sequence", Eur. J. Radiol, 89, pp.40~46 (2017)[DOI]
- 구은회, "뇌 확산강조 자기공명영상에 대한 정량적, 성적 평가: 1.5 T와 3.0 T 기기 비교", 방사선산업학회지, 10(4), pp.227~230 (2016), (사)한국방사선산업학회[DOI]
- Zeng Q, "Apparent diffusion coefficient maps obtained from high b value diffusion-weighted imaging in the preoperative evaluation of gliomas at 3T : comparison with standard b value diffusion-weighted imaging", Eur. Radiol, 27(12), pp.5309~5315 (2017)[DOI]