순음청력검사기의 교정

Audiometric Calibration of Pure Tone Audiometers

2

조회

0

다운로드

4

인용

0

참고문헌

키워드

초록

Purpose: This review is to introduce the performance characteristics of pure tone audiometer specified in the standards and necessary basic calibration instrumentations, explain the electroacoustic calibration among calibration procedures in order to determine if the pure tone audiometer which is being used currently satisfy the standards. Methods: Audiometric calibration consists of three parts; measurement of audiologic equipment performance characteristics, comparison of the measurements to a standard specified in the international or national standards, and adjustment as needed. Calibration checks and adjustments are an essential process for reducing a measurement errors between audiologic equipment and improving the reliability and accuracy of audiologic test result. Results: Specific calibration items that should be checked from audiologic equipment are summarized in International Electrotechnical Commission, International Organization for Standardization, and American National Standards Insti-tute standards. Audiologists should use audiologic equipment which meets appropriate standards and carry out the calibration in order to determine whether the result of calibration check satisfies current standards and exists in the range of a specific tolerances. Reliable test results obtained with audiology equipments calibrated support best audiology practice leading to increased confidence of audiologists. Conclusion: Audiologists have a responsibility to either check its calibration personally or to arrange for periodic calibration of the equipment by an outside service in order to maintain the accuracy of audiologic equipment. For this, audiologists should have proper ability to check whether audiologic equipment meets to the standardized performance characteristics. However, the performance characteristics and calibration methods vary according to the type of audiologic equipment.

Abstract

Purpose: This review is to introduce the performance characteristics of pure tone audiometer specified in the standards and necessary basic calibration instrumentations, explain the electroacoustic calibration among calibration procedures in order to determine if the pure tone audiometer which is being used currently satisfy the standards. Methods: Audiometric calibration consists of three parts; measurement of audiologic equipment performance characteristics, comparison of the measurements to a standard specified in the international or national standards, and adjustment as needed. Calibration checks and adjustments are an essential process for reducing a measurement errors between audiologic equipment and improving the reliability and accuracy of audiologic test result. Results: Specific calibration items that should be checked from audiologic equipment are summarized in International Electrotechnical Commission, International Organization for Standardization, and American National Standards Insti-tute standards. Audiologists should use audiologic equipment which meets appropriate standards and carry out the calibration in order to determine whether the result of calibration check satisfies current standards and exists in the range of a specific tolerances. Reliable test results obtained with audiology equipments calibrated support best audiology practice leading to increased confidence of audiologists. Conclusion: Audiologists have a responsibility to either check its calibration personally or to arrange for periodic calibration of the equipment by an outside service in order to maintain the accuracy of audiologic equipment. For this, audiologists should have proper ability to check whether audiologic equipment meets to the standardized performance characteristics. However, the performance characteristics and calibration methods vary according to the type of audiologic equipment.

참고문헌 (46)

  1. ANSI standard, "ANSI", ASA S3.13-1972, Method for Coupler Calibration of Earphones, 1972
  2. ANSI standard, "ANSI", ASA S1.4-1983 (R2006), Specification For Sound Level Meters, 2006
  3. ANSI standard, "ANSI", ASA S3. 7-1995(R2008), Method for Coupler Calibration of Earphones, 2008
  4. ANSI standard, "ANSI", ASA S1. 11-2004(R2009), Specification For Octave-Band And Fractional-Octave-Band Analog And Digital Filters, 2009
  5. ANSI standard, "ANSI", ASA S3. 6-2010, Specification for Audiometers, 2010
  6. ANSI standard, "ANSI", ASA S1. 40-2006(R2011), Specifications and Verification Procedures for Sound Calibrators, 2011
  7. Burkhard, M. D, "The response of earphones in ears and couplers", The Journal of the Acoustical Society of America, 26, pp.679~685 (1954)
  8. Corliss, E. L. R, "A probe tube method for the transfer of threshold standards between audiometer earphones", The Journal of the Acoustical Society of America, 25, pp.990~993 (1953)
  9. Dirks, D. D, "Bone conduction calibration : Current status", The Journal of Speech and Hearing Disorders, 44(2), pp.143~155 (1979)
  10. Frank, T, "Influence of contralateral masking on bone-conduction thresholds", Ear and Hearing, 3(6), pp.314~319 (1982)
  11. Frank, T, "Influence of temperature on the output of a mechanical coupler", Ear and Hearing, 6(4), pp.206~210 (1985)
  12. Hawkins, D. B, "Comparisons among SPLs in real ears, 2 cm3 and 6 cm3 couplers", Journal of the American Academy of Audiology, 1(3), pp.154~161 (1990)
  13. "IEC 60318-1, IEC 60318-1:2009 Electroacoustics - Simulators of human head and ear - Part 1: Ear simulator for the measurement of supra-aural and circumaural earphones", ~ (2009)
  14. "IEC 60318-3, Electroacoustics - Simulators of human head and ear -Part 3: Acoustic coupler for the calibration of supra-aural earphones used in audiometry", ~ (2014)
  15. "IEC 60318-4, Electroacoustics - Simulators of human head and ear -Part 4: Occluded-ear simulator for the measurement of earphones coupled to the ear by means of ear inserts", ~ (2010)
  16. "IEC 60318-6, Electroacoustics - Simulators of human head and ear -Part 6: Mechanical coupler for the measurement on bone vibrators", ~ (2007)
  17. "IEC 60645-1, Electroacoustics - Audiometric equipment - Part 1:Equipment for pure-tone audiometry", ~ (2012)
  18. "ISO 389-1, Acoustics -- Reference zero for the calibration of audiometric equipment -- Part 1: Reference equivalent threshold sound pressure levels for pure tones and supra-aural earphones", ~ (1998)
  19. "ISO 389-2, Acoustics -- Reference zero for the calibration of audiometric equipment -- Part 2: Reference equivalent threshold sound pressure levels for pure tones and insert earphones", ~ (1994)
  20. "ISO 389-3, Acoustics -- Reference zero for the calibration of audiometric equipment -- Part 3: Reference equivalent threshold force levels for pure tones and bone vibrators", ~ (1994)
  21. "ISO 389-4, Acoustics -- Reference zero for the calibration of audiometric equipment -- Part 4: Reference levels for narrow-band masking noise", ~ (1994)
  22. "ISO 389-5, Acoustics - Reference zero for the calibration of audiometric equipment. Part 5 - Reference equivalent threshold sound pressure levels for pure tones in the frequency range 8 kHz to 16 kHz", ~ (2006)
  23. "ISO 389-7, Acoustics - Reference zero for the calibration of audiometric equipment. Part 7 - Reference threshold of of hearing under free-field and diffuse-field listening conditions", ~ (2005)
  24. "ISO 389-8, Acoustics - Reference zero for the calibration of audiometric equipment. Part 8 - Reference equivalent threshold sound pressure levels for pure tones and circumaural earphones", ~ (2004)
  25. "ISO 8253-1, Acoustics -- Audiometric test methods -- Part 1: Puretone air and bone conduction audiometry", ~ (2010)
  26. "ISO 8253-2, Acoustics -- Audiometric test methods -- Part 2: Sound field audiometry with pure-tone and narrow-band test signals", ~ (2009)
  27. 강덕훈, "자동 ABLB 검사 소프트웨어 개발", 한국산학기술학회논문지, 11(12), pp.5120~5126 (2010), 한국산학기술학회
  28. 강덕훈, "PC 기반의 SISI 검사 소프트웨어 개발", 한국산학기술학회논문지, 11(4), pp.1325~1332 (2010), 한국산학기술학회
  29. 강덕훈, "PC 기반의 Békésy 청력검사 시스템 개발", 한국컴퓨터정보학회논문지, 16(7), pp.77~84 (2011), 한국컴퓨터정보학회
  30. Killion, M. C, "Revised estimate of minimum audible pressure : Where is the"missing 6 dB"?", The Journal of the Acoustical Society of America, 63(5), pp.1501~1508 (1978)
  31. Kim, H. W, "Current medical examination practices for the determination of occupational noise induced hearing losses", Korean Journal of Occupational And Environmental Medicine, 6(2), pp.276~288 (1994)
  32. 김진동, "PC 기반 순음청력검사기를 위한 자동보정시스템 개발", 한국산학기술학회논문지, 11(7), pp.2586~2594 (2010), 한국산학기술학회
  33. 김진동, "다중모드 지원 자동차폐 순음청력검사 시스템 개발", 한국산학기술학회논문지, 10(6), pp.1229~1236 (2009), 한국산학기술학회
  34. Kim, K. S, "Development of Korean audiometric quality control program for the special health examination", Audiology, 3(1), pp.9~16 (2007)
  35. Kim, K. S, "Audiometric quality assurance program in Korea", Korean Journal of Audiology, 13, pp.91~107 (2009)
  36. 김규상, "특수건강진단기관 청력검사기의 음향보정", Annals of Occupational and Environmental Medicine, 16(4), pp.381~390 (2004), 대한직업환경의학회
  37. "KS C IEC 60645-1, Audiometers - Part 1: Pure-tone audiometers", Korean Standards Association, ~ (2005)
  38. "KS I ISO 8253-1, Acoustics - Audiometric test methods - Part 1: Basic pure tone air and bone conduction threshold audiometry", Korean Standards Association, ~ (2014)
  39. "Larson Davis, Audiometer calibration & electroacoustic test systems" (2013)[URL]
  40. National Institute for Occupational Safety and Health, "Criteria for a Recommended Standard Occupational Noise Exposure:Revised Criteria 1998.U.S", Department of Health and Human Services,Public Health Service, Centers for Disease Control and Prevention,National Institute for OccupationalSafety and Health, DHHS(NIOSH), ~ (1998)
  41. Occupational Safety and Health Administration, "OccupationalSafety and Health Standards, Occupational Safety and HealthStandards,Occupational Health and Environmental Control, Occupationalnoise exposure, Hearing Conservation Amendment. OSHARegulations(Standards 29 CFR 1910.95 App E)-Acoustic calibrationof audiometers", Federal Register, United States GovernmentPrinting Office, ~ (1996)
  42. Rudmose, W, "Concerning the problem of calibrating TDH-39 earphones at 6 kHz with a 9A coupler", The Journal of the Acoustical Society of America, 36(5), pp.1049~ (1964)
  43. 송복득, "다수 피검자를 동시 검사할 수 있는 PC 기반 클라이언트", 서버 순음청력검사 시스템 개발, 2010, pp.11(6), 2225~2233[DOI]
  44. Zwicker, E, "Analytical expressions for critical band rate and critical bandwidth as a function of frequency", The Journal of the Acoustical Society of America, 68(5), pp.1523~1525 (1980)
  45. An acoustic coupler for earphone calibration, "Zwislocki, J. J", Laboratory of Sensory Communication, ~, pp.Special Report LSC-S-7 (1970)
  46. An ear-like coupler for earphone calibration, "Zwislocki, J. J", Laboratory of Sensory Communication, ~, pp.Special Report LSC-S-9 (1971)

관련 논문AI추천

이 간행물의 다른 논문 (6건)