본문 바로가기
KERD
논문KCI 등재

A Study on Radiation Dose and Image Quality using High-Pitch Mode when Examining the Lumbar Region with Dual-Energy Computed Tomography

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류컴퓨터학

초록

The purpose of this study was to investigate whether the high pitch protocol method of dual-energy computed tomography can be successfully applied to a lumbar vertebrae CT scan and reduce the effect of exposure dose; the main indicators were the image quality and reduced scan time. This study was conducted using MDCT 256 row dual source, Somatom Definition Flash (Siemens medical system: Germany), on 69 patients who had lumbar vertebrae CT scans from March 1, 2021, to June 30, 2021. First, to determine the applicability of the high pitch protocol method, the phantom images quality was compared with that of the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol method with the AAPM phantom, respectively. In the phantom image, CT noise, uniformity, and spatial resolution were not significantly different between the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol. In addition, in images examined from clinical patients, single-sample ttest results of the comparison of the mean value of CT noise, uniformity, and spatial resolution did not show statistically significant differences, 0.081, 0.078, and 0.387, respectively. The results of the qualitative evaluation showed that all were also higher than 3 points, which is fair, and the radiation dose was reduced by 9.76% in the high pitch protocol. When examining the lumbar region with dual-energy computed tomography, the examination method using high-pitch mode reduced the radiation dose and shortened the examination time, with the same image quality as that of the existing lumbar vertebrae CT scan.

Abstract

The purpose of this study was to investigate whether the high pitch protocol method of dual-energy computed tomography can be successfully applied to a lumbar vertebrae CT scan and reduce the effect of exposure dose; the main indicators were the image quality and reduced scan time. This study was conducted using MDCT 256 row dual source, Somatom Definition Flash (Siemens medical system: Germany), on 69 patients who had lumbar vertebrae CT scans from March 1, 2021, to June 30, 2021. First, to determine the applicability of the high pitch protocol method, the phantom images quality was compared with that of the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol method with the AAPM phantom, respectively. In the phantom image, CT noise, uniformity, and spatial resolution were not significantly different between the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol. In addition, in images examined from clinical patients, single-sample ttest results of the comparison of the mean value of CT noise, uniformity, and spatial resolution did not show statistically significant differences, 0.081, 0.078, and 0.387, respectively. The results of the qualitative evaluation showed that all were also higher than 3 points, which is fair, and the radiation dose was reduced by 9.76% in the high pitch protocol. When examining the lumbar region with dual-energy computed tomography, the examination method using high-pitch mode reduced the radiation dose and shortened the examination time, with the same image quality as that of the existing lumbar vertebrae CT scan.

참고문헌 (17)

  1. M. Dewey, "Noninvasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to-head pilot investigation", Circulation, 120, pp.867~875 (2009)[DOI]
  2. S. Achenbach, "Dual-source cardiac computed tomography: image quality and dose considerations", European radiology, 18, pp.1188~1198 (2008)[DOI]
  3. 최남길, "관상동맥 전산화단층촬영에서 64 channel MDCT와 128 channel DSCT의 임상 유용성 평가", 한국산학기술학회논문지, 11(11), pp.4411~4417 (2010), 한국산학기술학회
  4. F. H. J. Elsholtz, "Ultra-low-dose periradicular infiltration of the lumbar spine: spot scanning and its potential for further dose reduction by replacing helical planning CT", La radiologia medica, 122, pp.705~712 (2017)[DOI]
  5. F. H. J. Elsholtz, "Periradicular infiltration of the lumbar spine: is iterative reconstruction software necessary to establish ultra-low-dose protocols? A quantitative and qualitative approach", La radiologia medica, 123, pp.827~832 (2018)[DOI]
  6. B. Tins, "Technical aspects of CT imaging of the spine", Insights into imaging, 1, pp.349~359 (2010)[DOI]
  7. J. H. Kwak, "Quality improvement of general anteroposterior radiographic image of vertebral body according to optimum angle of incidence", Journal of X-Ray Science and Technology, 29, pp.297~306 (2021)[DOI]
  8. 김성진, "이중 에너지 전산화 단층촬영ECG Gating High Pitch Chest Pain Protocol 모드를이용한 방사선량과 영상 품질에 관한 연구", 한국방사선학회논문지, 16(1), pp.7~13 (2022), 한국방사선학회[DOI]
  9. J. J. Chung, "Usefulness of a lead shielding device for reducing the radiation dose to tissues outside the primary beams during CT", La radiologia medica, 119, pp.951~957 (2014)[DOI]
  10. J. Valentin, "Radiation and your patient: A guide for medical practitioners: ICRP Supporting Guidance 2: Approved by ICRP Committee 3 in September 2001", pp.1~52 (2001), SAGE Publications
  11. Y. Fukushima, "Usefulness of model-based iterative reconstruction in low-dose lumbar spine CT after spine surgery with metal implant: a phantom study", Acta Radiologica, 62, pp.1333~1340 (2021)[DOI]
  12. 구현우, "Image Quality and Radiation Dose of High-Pitch Dual-Source Spiral Cardiothoracic Computed Tomography in Young Children with Congenital Heart Disease: Comparison of Non-Electrocardiography Synchronization and Prospective Electrocardiography Triggering", Korean Journal of Radiology, 19(6), pp.1031~1041 (2018), 대한영상의학회[DOI]
  13. J. Weinrich, "MDCT in suspected lumbar spine fracture: comparison of standard and reduced dose settings using iterative reconstruction", Clinical radiology, 73, pp.679~675 (2018)[DOI]
  14. J. Greffier, "Optimization of radiation dose for CT detection of lytic and sclerotic bone lesions: a phantom study", European radiology, 30, pp.1075~1078 (2020)
  15. B. Rasuli, "Radiation dose measurement for patients undergoing common spine medical x-ray examinations and proposed local diagnostic reference levels", Radiation Measurements, 87, pp.29~34 (2016)[DOI]
  16. S. H. Lee, "Diagnosis of lumbar spinal fractures in emergency department: lowdose versus standard-dose CT using model-based iterative reconstruction", Clinical imaging, 50, pp.216~222 (2018)[DOI]
  17. W. Tan, "Reduction of radiation dose in the spiral CT scan of the lumbar spine by the combined use of body mass index (BMI) and automatic exposure control (AEC)", Journal of X-ray science and technology, 22, pp.697~706 (2014)[DOI]

관련 논문AI추천