초록

Titanium and titanium alloys are widely used for orthopedic and dental implants for their superior mechanical properties, low modulus, excellent corrosion resistance and good biocompatibility. In this study, corrosion behaviors of nanotube formed on the bone plate of Ti-6Al-4V alloy for dental use have been investigated. TiO2 nanotubes were formed on the dental bone plates by anodization in H3PO4 containing 0.6 wt % NaF solution at 25 ℃. Electrochemical experiments were performed using a conventional three-electrode configuration with a platinum counter electrode and a saturated calomel reference electrode. Anodization was carried out using a scanning potentiostat(EG&G Co, Model 263A USA), and all experiments were conducted at room temperature. The surface morphology were observed using field emission scanning electron microscopy(FE-SEM) and energy dispersive x-ray spectroscopy(EDS). The corrosion behavior of the dental bone plates was examined using potentiodynamic test(potential range of -1500 ~ 2000 mV) in a 0.9 % NaCl solution by potentiostat(EG&G Co, PARSTAT 2273. USA). The inner diameter of nanotube was about 150 ~ 180 nm with wall thickness of about 20 nm. Tie interspace of nanotube to nanotube was 50 nm. The passive region of the nanotube formed bone plates showed the broad range compared to non-nanotube formed bone plates. The corrosion surface of samples were covered with corrosion products.

Abstract

Titanium and titanium alloys are widely used for orthopedic and dental implants for their superior mechanical properties, low modulus, excellent corrosion resistance and good biocompatibility. In this study, corrosion behaviors of nanotube formed on the bone plate of Ti-6Al-4V alloy for dental use have been investigated. TiO2 nanotubes were formed on the dental bone plates by anodization in H3PO4 containing 0.6 wt % NaF solution at 25 ℃. Electrochemical experiments were performed using a conventional three-electrode configuration with a platinum counter electrode and a saturated calomel reference electrode. Anodization was carried out using a scanning potentiostat(EG&G Co, Model 263A USA), and all experiments were conducted at room temperature. The surface morphology were observed using field emission scanning electron microscopy(FE-SEM) and energy dispersive x-ray spectroscopy(EDS). The corrosion behavior of the dental bone plates was examined using potentiodynamic test(potential range of -1500 ~ 2000 mV) in a 0.9 % NaCl solution by potentiostat(EG&G Co, PARSTAT 2273. USA). The inner diameter of nanotube was about 150 ~ 180 nm with wall thickness of about 20 nm. Tie interspace of nanotube to nanotube was 50 nm. The passive region of the nanotube formed bone plates showed the broad range compared to non-nanotube formed bone plates. The corrosion surface of samples were covered with corrosion products.

참고문헌 (35)

  1. D. M Brunette, "Titanium in Medicine", pp.317~ (2001), Springer
  2. H. C. Choe, "Materials Science Forum", Materials Science Forum, pp.475~479 (2005)
  3. M. I. Jones, , Diamond and Related Materials, 8, pp.457~ (1999)
  4. Viswanathan S. Saji, "A Comparative Study on Corrosion Behavior of Ti-35Nb-5Ta-7Zr, Ti-6Al-4V and CP-Ti in 0.9 wt% NaCl", CORROSION SCIENCE AND TECHNOLOGY, 8(4), pp.139~142 (2009), 한국부식방식학회
  5. 김영운, "TiN", Ti 다층막 코팅된 생체용 Ti-30Ta-xZr 합금의 부식특성, 2009, pp.8(4), 162~169[DOI]
  6. 정용훈, "HA", Ti 복합층 코팅한 Ti-30Ta-(3~15)Nb 합금의 전기화학적 특성, 2008, pp.41(2), 57~62[DOI]
  7. 김원기, "골아세포가 배양된 치과 임플란트용 Ti-Ta합금의 전기화학적 특성", 한국표면공학회지, 41(2), pp.69~75 (2008), 한국표면공학회
  8. 김태한, "다양한 유사체액과 인공타액에서 치과용 임플란트의 전기화학적 특성", 한국표면공학회지, 41(5), pp.226~231 (2008), 한국표면공학회
  9. G. Gianluca, , Biomaterials, 25, pp.5583~ (2004)
  10. J. W. Thomas, , Biomaterials, 25, pp.4731~ (2004)
  11. T. O. Paulo, , Biomaterials, 25, pp.403~ (2004)
  12. J. W. Thomas, , Biomaterials, 25, pp.4731~ (2004)
  13. T. O. Paulo, , Biomaterials, 25, pp.403~ (2004)
  14. J. D. Bronzino, "Biomedical Enginnering Handbook", pp.274~ (1995), CRC’ Press
  15. J. Zhoa, , Solid. State. Communication, 134, pp.705~ (2005)
  16. J. M. Macak, , Angew. Chem. Int. Ed., 44, pp.2100~ (2005)
  17. J. M. K. Sirotna, , Electrochimica Acta, 50, pp.3679~ (2005)
  18. R. Beranek, , Electrochemical Solid-State Letters, 6, pp.B12~ (2003)
  19. H. Tsuchiya, , Electrochemistry Communications, 7, pp.576~ (2005)
  20. G. K. Mor, , J. Mater. Res., 18, pp.2588~ (2003)
  21. S. Kaneco, , Scripta Materialia, 56, pp.373~ (2007)
  22. B. Yang, , Biomaterials, 25, pp.1003~ (2004)
  23. D. Gong, , J. Maer. Res., 16, pp.3331~ (2001)
  24. W. G. Kim, , NSTI-Nanotech, 1, pp.462~ (2008)
  25. 김현승, "양극 산화법에 의해 형성된 나노 튜브 타이타늄 산화 피막의 표면 특성", 대한치과기재학회지, 33(2), pp.163~170 (2006), 대한치과기재학회
  26. W. G. Kim, , Thin Solid Films, 517, pp.5033~ (2009)
  27. 김원기, "Electrochemical Behaviors of a TiN-Coated", Nanotube-formed Ti-Zr Alloy, 2009, pp.54(3), 1036~1041[DOI]
  28. 김현승, "양극산화에 의해 CP Ti와 Ti-10Ta-10Nb 합금 표면에 형성된 산화 피막의 형상 및 표면 특성", 한국재료학회지, 17(1), pp.6~10 (2007), 한국재료학회
  29. S. H. Jang, , Thin Solid Films, 517, pp.5038~ (2009)
  30. V. S. Saji, , Corrosion Science, 51, pp.1658~ (2009)
  31. V. S. Saji, , Acta Biomateirialia, 5, pp.2303~ (2009)
  32. Y. H. Jeong, , Thin Solid Films, 517, pp.5365~ (2009)
  33. P. P. Binon, , Quintessence, ~ (1992)
  34. A. F. Recum, , Biomaterials, 9, pp.297~ (1990)
  35. 정해만, "양극산화처리된 임플란트 고정체의 전기화학적 특성", 대한치과기재학회지, 35(4), pp.369~380 (2008), 대한치과기재학회

관련 논문AI추천

이 간행물의 다른 논문 (7건)