This paper focused on establishing a closed-loop stimulation environment capable of accommodating multiple experimental animals. The key requirement for closed-loop stimulation is the ability to measure neural signals within the stimulation environment. To achieve simultaneous measurement of stimulation artifacts in the range of tens of mV and neural signals in the range of hundreds of 𝛍𝛍V, a recording system with a large dynamic range is essential to prevent saturation caused by the high-intensity stimulation artifacts. Furthermore, to transmit the acquired data externally, down-sampling is essential to reduce power consumption. In this work, delta-sigma (DS) modulation and digital prediction techniques are used to ensure a high dynamic range and prevent saturation in the recording stage. Additionally, to facilitate the simultaneous transmission of data from multiple animals without signal degradation, an internal cascaded integrator-comb (CIC) filter is utilized to encode the down-sampled data. The proposed approach enables the creation of a robust closed-loop stimulation recording system capable of handling multiple subjects while maintaining data integrity. We present 8 channel neural recording chip with a 65-nm CMOS process and the entire chip area is 1 mm2. Two signals were transmitted and received at a distance of 10 cm with a 1 Mbit data rate without an error.

Abstract

This paper focused on establishing a closed-loop stimulation environment capable of accommodating multiple experimental animals. The key requirement for closed-loop stimulation is the ability to measure neural signals within the stimulation environment. To achieve simultaneous measurement of stimulation artifacts in the range of tens of mV and neural signals in the range of hundreds of 𝛍𝛍V, a recording system with a large dynamic range is essential to prevent saturation caused by the high-intensity stimulation artifacts. Furthermore, to transmit the acquired data externally, down-sampling is essential to reduce power consumption. In this work, delta-sigma (DS) modulation and digital prediction techniques are used to ensure a high dynamic range and prevent saturation in the recording stage. Additionally, to facilitate the simultaneous transmission of data from multiple animals without signal degradation, an internal cascaded integrator-comb (CIC) filter is utilized to encode the down-sampled data. The proposed approach enables the creation of a robust closed-loop stimulation recording system capable of handling multiple subjects while maintaining data integrity. We present 8 channel neural recording chip with a 65-nm CMOS process and the entire chip area is 1 mm2. Two signals were transmitted and received at a distance of 10 cm with a 1 Mbit data rate without an error.

참고문헌 (17)

  1. Reitz, "Epidemiology of Alzheimer disease", Nature Reviews Neurology (2011)
  2. Benabid, Alim-Louis, "Combined(thalamotomy and stimulation)stereotactic surgery of the VIM thalamic nucleus for bilateral Parkinson disease", Stereotactic and functional neurosurgery (1987)
  3. Chang, Chun-Hung, "Brain stimulation in Alzheimer's disease", Frontiers in psychiatry, 9 (2018)
  4. Chen, K.-H. S., "Invasive and noninvasive brain stimulation in Parkinson’s disease : clinical effects and future perspectives", Clinical Pharmacology & Therapeutics, 106 (2019)
  5. Theodore, William H., "Brain stimulation for epilepsy", The Lancet Neurology, 3(2) (2004)
  6. Idogawa, Shinnosuke, "A lightweight, wireless Bluetooth-low-energy neuronal recording system for mice", Sensors and Actuators B: Chemical . Eng (2021)
  7. W. Li, "A 200Mb", s inductively coupled wireless transcranial transceiver achieving 5e-11 BER and 1.5pJ, IEEE International Solid - State Circuits Conference - (ISSCC), Eng, pp.2018[DOI]
  8. Mirbozorgi, S. Abdollah, "A single-chip full-duplex high speed transceiver for multi-site stimulating and recording neural implants", IEEE transactions on biomedical circuits and systems. Eng (2016)
  9. Jia, Yaoyao, "A trimodal wireless implantable neural interface system-on-chip", IEEE Transactions on Biomedical Circuits and Systems, Eng (2020)
  10. J. J. D. McKendry, "Visible-Light Communications Using a CMOS-Controlled Micro-Light-Emitting-Diode Array", Journal of Lightwave Technology, Eng (2012)
  11. de Groot, Andres, "NINscope, a versatile miniscope for multi-region circuit investigations", Elife, Eng (2020)
  12. C. Kim, "Sub-μW Vrms-Noise Sub-μW", Channel ADC-Direct Neural Recording With 200-mV, IEEE Journal of Solid-State Circuits, pp.2018, 53(11)[DOI]
  13. Hogenauer, Eugene, "An economical class of digital filters for decimation and interpolation", IEEE transactions on acoustics, speech, and signal processing, 29(2) (1981)
  14. Schreier, Richard, "Understanding delta-sigma data converters. Vol. 74" (2005), IEEE press
  15. H. Chandrakumar, "A 15.2-ENOB continuous-time ΔΣ ADC for a 200mVpp-linear-input-range neural recording front-end", IEEE ISSCC Dig. Tech. Papers, 15, pp.232~234 (2018)
  16. Yang, Xiaolin, "An AC-Coupled 1st-Order Δ-ΔΣ Readout IC for Area-Efficient Neural Signal Acquisition", IEEE Journal of Solid-State Circuits, 58(4) (2023)
  17. Oh, Sungjin, "Power-Efficient LFP-Adaptive Dynamic Zoom-and-Track Incremental ΔΣ Front-End for Dual-Band Subcortical Recordings", IEEE Transactions on Biomedical Circuits and Systems (2023)