Removal of Dental Implants using Electrocautery: A Preliminary Report

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Purpose: We introduce a technique using a monopolar electrocautery unit to eliminate failed dental implant fixtures. The goal of this research was to establish a protocol of electrocautery settings when removing dental implants. Materials and Methods: The study was conducted on fresh pork ribs. Implantium implants sized Ø4.5 × 10 mm were used. They (n = 45) were categorized into three groups according to the monopolar electric cautery power (coagulation mode) and adapted time as follows: 30 watt and 20 second (n = 15), 40 watt and 40 second (n = 15), and control group (n = 15). The torque in the removal of the implant fixtures, after monopolar electro cauterization directly inside the implant fixture, was also measured. Results: The mean torque [unit; N·mm] during removal of the dental implant fixtures with an Elcomed implant engine, after monopolar electrocauterization directly inside the implant fixture, for each group was 296 in the 30 watt and 20 second group and 184 in the 40 watt and 40 second group. The mean torque of removal of the dental implant fixtures was 279 N·mm in the control group. The mean torque was significantly lower in the 40 watt and 40 second group than in the other groups (p = 0.006). Conclusion: This method employs the principle of thermo-necrosis of the surrounding bone and weakening of the bone-implant interface. Monopolar electrocautery can be considered as an alternative method for the removal of failed or poorly positioned dental implants without the need of grinding the bony tissue surrounding the implant fixture.

Abstract

Purpose: We introduce a technique using a monopolar electrocautery unit to eliminate failed dental implant fixtures. The goal of this research was to establish a protocol of electrocautery settings when removing dental implants. Materials and Methods: The study was conducted on fresh pork ribs. Implantium implants sized Ø4.5 × 10 mm were used. They (n = 45) were categorized into three groups according to the monopolar electric cautery power (coagulation mode) and adapted time as follows: 30 watt and 20 second (n = 15), 40 watt and 40 second (n = 15), and control group (n = 15). The torque in the removal of the implant fixtures, after monopolar electro cauterization directly inside the implant fixture, was also measured. Results: The mean torque [unit; N·mm] during removal of the dental implant fixtures with an Elcomed implant engine, after monopolar electrocauterization directly inside the implant fixture, for each group was 296 in the 30 watt and 20 second group and 184 in the 40 watt and 40 second group. The mean torque of removal of the dental implant fixtures was 279 N·mm in the control group. The mean torque was significantly lower in the 40 watt and 40 second group than in the other groups (p = 0.006). Conclusion: This method employs the principle of thermo-necrosis of the surrounding bone and weakening of the bone-implant interface. Monopolar electrocautery can be considered as an alternative method for the removal of failed or poorly positioned dental implants without the need of grinding the bony tissue surrounding the implant fixture.

참고문헌 (16)

  1. Moraschini V, "Evaluation of survival and success rates of dental implants reported in longitudinal studies with a follow-up period of at least 10 years: a systematic review", Int J Oral Maxillofac Surg, 44, pp.377~388 (2015)
  2. Howe MS, "Long-term (10-year) dental implant survival: a systematic review and sensitivity meta-analysis", J Dent, 84, pp.9~21 (2019)
  3. Stajčić Z, "Removal of dental implants: review of five different techniques", Int J Oral Maxillofac Surg, 45, pp.641~648 (2016)
  4. Cunliffe J, "Removal of a dental implant: an unusual case report", Journal of Dental Implants, 1, pp.22~25 (2011)
  5. Eriksson AR, "Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit", J Prosthet Dent, 50, pp.101~107 (1983)
  6. Chee W, "Failures in implant dentistry", Br Dent J, 202, pp.123~129 (2007)
  7. Krejci RF, "Effects of electrosurgery on dog pulps under cervical metallic restorations", Oral Surg Oral Med Oral Pathol, 54, pp.575~582 (1982)
  8. Zach L, "Pulp response to externally applied heat", Oral Surg Oral Med Oral Pathol, 19, pp.515~530 (1965)
  9. Baldissara P, "Clinical and histological evaluation of thermal injury thresholds in human teeth: a preliminary study", J Oral Rehabil, 24, pp.791~801 (1997)
  10. Kate M, "Implant failure: a dentist's nightmare", J Dent Implant, 6, pp.51~56 (2016)
  11. Kotsovilis S, "Therapy of peri-implantitis: a systematic review", J Clin Periodontol, 35, pp.621~629 (2008)
  12. Smith DE, "Criteria for success of osseointegrated endosseous implants", J Prosthet Dent, 62, pp.567~572 (1989)
  13. Branemark PI, "Osseointegration and its experimental background", J Prosthet Dent, 50, pp.399~410 (1983)
  14. Lin G, "A retrospective study of 30,959 implants: risk factors associated with early and late implant loss", J Clin Periodontol, 45, pp.733~743 (2018)
  15. Javed F, "Role of primary stability for successful osseointegration of dental implants: factors of influence and evaluation", Interv Med Appl Sci, 5, pp.162~167 (2013)
  16. Wilcox CW, "Use of electrosurgery and lasers in the presence of dental implants", Int J Oral Maxillofac Implants, 16, pp.578~582 (2001)

관련 논문AI추천

이 간행물의 다른 논문 (4건)