Evaluation of Matrix Effect in Body Fluid Chemistry on Roche Cobas 8000 c702 System

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: Analysis of body fluids aids in diagnosis and monitoring disease. However, only a few testing platforms and reagents have been validated for a range of body fluids or analytes. In this study, we evaluated a testing system, which has been approved for blood samples, in analyzing body fluid specimens upon matrix mixing. Methods: Serum and body fluid samples, including cerebrospinal fluid (CSF), ascites, pleural fluid, amniotic fluid, and synovial fluid, were mixed, then the matrix effect and linearity for major analytes, namely amylase, chloride, glucose, LDH, and protein were evaluated (N = 30 serum-body fluid pairs) on the Cobas 8000 c702. The obtained data was compared with that of open reagents evaluated on the Architect c16000. Results: For all analyte-body fluid pairs, the mean percent recovery ranged from 98.4% to 101.7%, and this was within the acceptable range for matrix effect. In the linearity test, maximum non-linearity for each analyte-body fluid pair ranged from -5.0% to +4.2%. In interference test, proteins showed positive hemolytic, icteric, and lipemic interference in CSF and hemolytic interference in amniotic fluid. There was no significant interference in the other analyte-body fluid pairs. Results were highly correlated between the Cobas 8000 c702 and the Architect c16000 system. Conclusions: Our findings revealed that the matrix effect of major analytes in body fluid specimens can be excluded and they also validated the linearity of the analytes in the body fluid specimens. Therefore, reagents specified for blood samples can be readily adopted for the analysis of body fluids.

Abstract

Background: Analysis of body fluids aids in diagnosis and monitoring disease. However, only a few testing platforms and reagents have been validated for a range of body fluids or analytes. In this study, we evaluated a testing system, which has been approved for blood samples, in analyzing body fluid specimens upon matrix mixing. Methods: Serum and body fluid samples, including cerebrospinal fluid (CSF), ascites, pleural fluid, amniotic fluid, and synovial fluid, were mixed, then the matrix effect and linearity for major analytes, namely amylase, chloride, glucose, LDH, and protein were evaluated (N = 30 serum-body fluid pairs) on the Cobas 8000 c702. The obtained data was compared with that of open reagents evaluated on the Architect c16000. Results: For all analyte-body fluid pairs, the mean percent recovery ranged from 98.4% to 101.7%, and this was within the acceptable range for matrix effect. In the linearity test, maximum non-linearity for each analyte-body fluid pair ranged from -5.0% to +4.2%. In interference test, proteins showed positive hemolytic, icteric, and lipemic interference in CSF and hemolytic interference in amniotic fluid. There was no significant interference in the other analyte-body fluid pairs. Results were highly correlated between the Cobas 8000 c702 and the Architect c16000 system. Conclusions: Our findings revealed that the matrix effect of major analytes in body fluid specimens can be excluded and they also validated the linearity of the analytes in the body fluid specimens. Therefore, reagents specified for blood samples can be readily adopted for the analysis of body fluids.

참고문헌 (12)

  1. American Society of Clinical Pathologists, "Handbook of clinical pathology" (1999), American Society of Clinical Pathologists
  2. Wians FH Jr, "Clinical laboratory tests: Which, why, and what do the results mean?", Labmedicine, 40, pp.105~113 (2009)
  3. Block DR, "Body fluid analysis: clinical utility and applicability of published studies to guide interpretation of today’s laboratory testing in serous fluids", Crit Rev Clin Lab Sci, 50, pp.107~124 (2013)
  4. Owen WE, "Body fluid matrix evaluation on a Roche cobas 8000 system", Clin Biochem, 48, pp.911~914 (2015)
  5. Block DR, "An approach to analytical validation and testing of body fluid assays for the automated clinical laboratory", Clin Biochem, 58, pp.44~52 (2018)
  6. Adams A, "Peritoneal and pleural fluid chemistry measurements performed on three chemistry platforms", Lab Med, 50, pp.145~149 (2019)
  7. Clinical and Laboratory Standards Institute, "Analysis of body fluids in clinical chemistry: Approved guideline. CLSI document C49-A", Clinical and Laboratory Standards Institute, ~ (2007)
  8. Swift CP, "Body fluid testing at John F. Kennedy Medical Center in Liberia", Am J Clin Pathol, 152, pp.86~90 (2019)
  9. "Desirable biological variation database specifications" (2014)[URL]
  10. Clinical and Laboratory Standards Institute, "Evaluation of the linearity of quantitative measurement procedures: a statistical approach; Approved guideline. CLSI document EP06-A", Clinical and Laboratory Standards Institute, ~ (2003)
  11. "All common checklist: COM.40620 Body fluid analysis" (2017)[URL]
  12. Lippi G, "Opportunities and drawbacks of nonstandard body fluid analysis", Clin Chem Lab Med, 55, pp.907~909 (2017)

관련 논문AI추천

Hitachi LABOSPECT 008에서의 Wako 및 Sekisui 임상화학시약의 분석성능김명신, 김용구, 박해일, 이제훈, 채효진대한진단검사의학회 · Laboratory Medicine Online2013 · 제3권 제4호Performance Evaluation of the JEOL BioMajesty JCA-BM6010/C Automated Clinical Chemistry Analyzer김형년, 문희원, 윤여민, 지미숙, 허미나대한진단검사의학회 · Laboratory Medicine Online2017 · 제7권 제3호Proposal of Modified HIL-indices for Determining Hemolysis, Icterus and Lipemia Interference on the Beckman Coulter AU5800 Automated Platform임용관, 차영주대한진단검사의학회 · Laboratory Medicine Online2017 · 제7권 제2호Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular AnalyzersJaewoo Song, Jeong-HoKim, Joowon Oh, Jooyoung Cho, Sang-Guk Lee, You-Hee Lee대한진단검사의학회 · Annals of Laboratory Medicine2020 · 제40권 제2호Periodic Comparability Verification and Within-Laboratory Harmonization of Clinical Chemistry Laboratory Results at a Large Healthcare Center With Multiple InstrumentsJun Sun-Hee, Kim Sung Min, Lee Joon Hee, Lee Kyunghoon, Nam Youngwon, Song Junghan, Song Sang Hoon대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제2호Roche-Hitachi cobas 8000 c702 자동화학분석기의 성능 평가김소영, 민원기, 이우창, 전사일, 정태동대한진단검사의학회 · Laboratory Medicine Online2014 · 제4권 제3호

이 간행물의 다른 논문 (11건)