Proposal of Modified HIL-indices for Determining Hemolysis, Icterus and Lipemia Interference on the Beckman Coulter AU5800 Automated Platform

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: The amount of interference due to hemolysis, bilirubin, and lipemia can be measured on the AU5800 autoanalyzer (Beckman Coulter, USA) by spectrophotometry. This is reported as semi-quantitative indices, specifically H-index, I-index, and L-index, respectively. In this study, we evaluated the impact of interference using chemistry assays and established the concentration of interfering substances and HIL-index above which analytically significant interference exists, according to CLSI guidelines C56-A and EP7-A2. Methods: Pooled sera including different concentrations of analytes were prepared and mixed with hemoglobin, bilirubin, or Intralipid. These samples were then tested for 35 clinical chemistry analytes by AU5800 and the bias based on interferent concentrations was computed. The interferent concentration above which significant interference exists was calculated from the 50% within-subject biological variation (desirable analytic goal), and the corresponding index was assigned. Results: Among 35 items evaluated, interference was detected for 12 analytes by hemoglobin, 7 analytes by bilirubin, and 12 analytes by Intralipid. We proposed HIL-index1 and HIL-index2 for each analyte according to 2 different medical decision levels. HIL-index1 and HIL-index2 were considered more reasonable criteria than the HIL-index from the manufacturer’s technical document (HIL-indexTD). This is because HIL-indexTD was empirically set to 5% or 10%, and had a wide tolerance range, which was not sufficient to reflect the presence of interference, compared to HIL-index1 and HIL-index2. Conclusions: We have demonstrated hemoglobin, bilirubin, and Intralipid interferences according to CLSI guidelines using the desirable analytic goal. Our results provide applicable information for Beckman Coulter automated chemistry analyzers.

Abstract

Background: The amount of interference due to hemolysis, bilirubin, and lipemia can be measured on the AU5800 autoanalyzer (Beckman Coulter, USA) by spectrophotometry. This is reported as semi-quantitative indices, specifically H-index, I-index, and L-index, respectively. In this study, we evaluated the impact of interference using chemistry assays and established the concentration of interfering substances and HIL-index above which analytically significant interference exists, according to CLSI guidelines C56-A and EP7-A2. Methods: Pooled sera including different concentrations of analytes were prepared and mixed with hemoglobin, bilirubin, or Intralipid. These samples were then tested for 35 clinical chemistry analytes by AU5800 and the bias based on interferent concentrations was computed. The interferent concentration above which significant interference exists was calculated from the 50% within-subject biological variation (desirable analytic goal), and the corresponding index was assigned. Results: Among 35 items evaluated, interference was detected for 12 analytes by hemoglobin, 7 analytes by bilirubin, and 12 analytes by Intralipid. We proposed HIL-index1 and HIL-index2 for each analyte according to 2 different medical decision levels. HIL-index1 and HIL-index2 were considered more reasonable criteria than the HIL-index from the manufacturer’s technical document (HIL-indexTD). This is because HIL-indexTD was empirically set to 5% or 10%, and had a wide tolerance range, which was not sufficient to reflect the presence of interference, compared to HIL-index1 and HIL-index2. Conclusions: We have demonstrated hemoglobin, bilirubin, and Intralipid interferences according to CLSI guidelines using the desirable analytic goal. Our results provide applicable information for Beckman Coulter automated chemistry analyzers.

참고문헌 (20)

  1. Kroll MH, "Interference with clinical laboratory analyses", Clin Chem, 40, pp.1996~2005 (1994)
  2. Dimeski G, "Interference testing", Clin Biochem Rev, 29(S1), pp.S43~S48 (2008)
  3. Glick MR, "Unreliable visual estimation of the incidence and amount of turbidity, hemolysis, and icterus in serum from hospitalized patients", Clin Chem, 35, pp.837~839 (1989)
  4. Steen G, "Multicenter evaluation of the interference of hemoglobin, bilirubin and lipids on Synchron LX-20 assays", Clin Chem Lab Med, 44, pp.413~419 (2006)
  5. Ji JZ, "Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assays", Clin Chim Acta, 412, pp.1550~1553 (2011)
  6. Steen G, "Evaluation of the interference due to haemoglobin, bilirubin and lipids on Immulite 2500 assays: a practical approach", Ann Clin Biochem, 48, pp.170~175 (2011)
  7. turbidity indices as indicators of interference in clinical laboratory analysis; approved guideline. C56-A, "Hemolysis, icterus, and lipemia", pp.2012, Clinical and Laboratory Standards Institute
  8. Clinical and Laboratory Standards Institute, "Interference testing in clinical chemistry; approved guideline" (2005), Clinical and Laboratory Standards Institute
  9. Kroll MH, "Evaluating interference caused by lipemia", Clin Chem, 50, pp.1968~1969 (2004)
  10. Fraser CG, "Test result variation and the quality of evidence-based clinical guidelines", Clin Chim Acta, 346, pp.19~24 (2004)
  11. Fraser CG, "Proposals for setting generally applicable quality goals solely based on biology", Ann Clin Biochem, 34(Pt 1), pp.8~12 (1997)
  12. Nikolac N., "Lipemia: causes, interference mechanisms, detection and management", Biochem Med (Zagreb), 24, pp.57~67 (2014)
  13. Petersen PH, "Strategies to set global analytical quality specifications in laboratory medicine: 10 years on from the Stockholm consensus conference", Accred. Qual. Assur, 15, pp.323~330 (2010)
  14. Ricos C, "Within-subject biological variation in disease : collated data and clinical consequences", Ann Clin Biochem, 44, pp.343~352 (2007)
  15. Young DS, "Effects of drugs on clinical laboratory tests", Ann Clin Biochem, 34(Pt 6), pp.579~581 (1997)
  16. Sandberg S, "Defining analytical performance specifications : Consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine", Clin Chem Lab Med, 53, pp.833~835 (2015)
  17. 우길성, "Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens", Annals of Laboratory Medicine, 35(3), pp.341~347 (2015), 대한진단검사의학회
  18. Bornhorst JA, "Assay-specific differences in lipemic interference in native and intralipid-supplemented samples", Clin Chem, 50, pp.2197~2201 (2004)
  19. Huang YC, "Evaluation of two homogeneous methods for measuring high-density lipoprotein cholesterol", Clin Chem, 43, pp.1048~1055 (1997)

관련 논문AI추천

용혈, 황달 및 지질혈증: 자동화 임상화학분석기를 이용한 검사 결과 간섭 평가Jeong Ina, Jun Kyung Ran, Kim Won-Joo, Lee Sun Min, Oh Seung Hwan, Park Chul Min, Yoo Dong-Won대한진단검사의학회 · Laboratory Medicine Online2026 · 제16권 제1호Development of an Integrated Reporting System for Verifying Hemolysis, Icterus, and Lipemia in Clinical Chemistry Results어영, 김갑성, 김주원, 서동민, 신동훈, 윤갑준, 윤광로, 이만희, 이세일, 장재윤대한진단검사의학회 · Annals of Laboratory Medicine2014 · 제34권 제4호Impact of Individualized Hemolysis Management Based on Biological Variation Cut-offs in a Clinical LaboratoryJung David Hansoe Heredero, Marques-Garcia Fernando, Pérez Sandra Elena대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제2호화학검사 결과의 검증을 위한 용혈 지수 일치화 도구 개발남현수, 이승관, 이혜경, 차영종, 표상신대한임상검사과학회 · 대한임상검사과학회지2017 · 제49권 제4호Development of a Novel Quality Improvement Indicator Based on the Hemolysis Index강희정, 김미영, 김한성, 박민정, 이영경, 이은진대한진단검사의학회 · Annals of Laboratory Medicine2016 · 제36권 제6호재채혈된 검체를 이용한 허용 Hemolysis Index에 대한 연구Dong Il Won, Hong Bum KIM, Jin Man KIM, Kyoung Ae SON, Yu Jin WOO대한임상검사과학회 · 대한임상검사과학회지2024 · 제56권 제1호

이 간행물의 다른 논문 (5건)