Polyethylene spinout in the Attune® Cruciate-Retaining Rotating-Platform (CR RP) total knee arthroplasty performed with a cruciate-sacrificing and measured-resection technique

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Introduction: Polyethylene (PE) spinout is a known but uncommon complication when using a mobile-bearing (MB) total knee arthroplasty (TKA) design. Sacrificing the posterior cruciate ligament (PCL) is within the manufacturer’s recommendations for the Attune® Cruciate-Retaining Rotating-Platform (CR RP) knee design. Aim: To discuss the potential aetiology and prevention of spinout in the Attune® CR RP knee. Methods: We used a retrospective radiological review from two centres reporting a higher rate of spinout in the Attune® CR RP knee using a cruciate-sacrificing and measured-resection technique when compared to a gapbalancing technique. Three hundred and thirty-two patients were evaluated over a 3-year period. Results: There were 8 out of 279 (2.86%) cases of spinout in our first cohort of patients using a measured-resection technique. There were 0 out of 53 cases of spinout in our second cohort of patients where a gap-balancing technique was used. One spinout was reduced closed, the other seven were initially revised to a thicker RP insert of the same design. Of these seven, three underwent a further revision TKA and one patient required a knee fusion/arthrodesis. Conclusions: This study reports a higher incidence of PE spinout in the Attune® CR RP TKA when a measuredresection technique in combination with PCL resection is performed. We recommend a gap-balancing technique with conservative soft-tissue release if the surgeon is planning to sacrifice the PCL in the Attune® CR RP.

Abstract

Introduction: Polyethylene (PE) spinout is a known but uncommon complication when using a mobile-bearing (MB) total knee arthroplasty (TKA) design. Sacrificing the posterior cruciate ligament (PCL) is within the manufacturer’s recommendations for the Attune® Cruciate-Retaining Rotating-Platform (CR RP) knee design. Aim: To discuss the potential aetiology and prevention of spinout in the Attune® CR RP knee. Methods: We used a retrospective radiological review from two centres reporting a higher rate of spinout in the Attune® CR RP knee using a cruciate-sacrificing and measured-resection technique when compared to a gapbalancing technique. Three hundred and thirty-two patients were evaluated over a 3-year period. Results: There were 8 out of 279 (2.86%) cases of spinout in our first cohort of patients using a measured-resection technique. There were 0 out of 53 cases of spinout in our second cohort of patients where a gap-balancing technique was used. One spinout was reduced closed, the other seven were initially revised to a thicker RP insert of the same design. Of these seven, three underwent a further revision TKA and one patient required a knee fusion/arthrodesis. Conclusions: This study reports a higher incidence of PE spinout in the Attune® CR RP TKA when a measuredresection technique in combination with PCL resection is performed. We recommend a gap-balancing technique with conservative soft-tissue release if the surgeon is planning to sacrifice the PCL in the Attune® CR RP.

참고문헌 (18)

  1. Malinzak RA, "The effect of rotating platform TKA on strain distribution and torque transmission on the proximal tibia", J Arthroplast, 29, pp.541~547 (2014)
  2. Hopley CD, "Long-term clinical outcomes and survivorship after total knee arthroplasty using a rotating platform knee prosthesis : a meta-analysis", J Arthroplast, 28(1), pp.68~77 (2013)
  3. Baker PN, "Patient satisfaction with total knee replacement cannot be predicted from pre-operative variables alone: a cohort study from the National Joint Registry for England and Wales", Bone Joint J, 95-B(10), pp.1359~1365 (2013)
  4. Bourne RB, "Comparing patient outcomes after THA and TKA: is there a difference?", Clin Orthop Relat Res, 468(2), pp.542~546 (2010)
  5. "15th Annual Report" (2018)[URL]
  6. "Attune® knee system surgical technique guide"[URL]
  7. Diamond OJ, "Spinout", dislocation in mobile-bearing total knee arthroplasty : a report of 26 cases, 2018, pp.33(2), 537~543
  8. Daines BK, "Gap balancing vs. measured resection technique in total knee arthroplasty", Clin Orthop Surg, 6(1), pp.1~8 (2014)
  9. Pagoti R, "Soft tissue balancing without ligament release in complex total knee replacement" (2017), Thieme Medical and Scientific Publishers Pvt
  10. Pagoti R, "Unconstrained total knee arthroplasty in significant valgus deformity : a modified surgical technique to balance the knee and avoid instability", Knee Surg Sports Traumatol Arthrosc, 25(9), pp.2825~2834 (2017)
  11. Lebel B, "Total knee arthroplasty dislocation: incidence, etiology, and management", J Bone Jt Surg Br, 92-B(Suppl II) (2010)
  12. Yoshiya S, "In vivo kinematic comparison of posterior cruciate-retaining and posterior stabilized total knee arthroplasties under passive and weightbearing conditions", J Arthroplast, 20(6), pp.777~783 (2005)
  13. Dennis DA, "Gap balancing versus measured resection technique for total knee arthroplasty", CORR, 468(1), pp.102~107 (2010)
  14. Katz MA, "Determining femoral rotational alignment in total knee arthroplasty : reliability of techniques", J Arthroplasty, 16(3), pp.301~305 (2001)
  15. Kim CW, "The effects of surgical technique in total knee arthroplasty for Varus osteoarthritic knee on the rotational alignment of femoral component: gap balancing technique versus measured resection technique", J Knee Surg, 33(2), pp.144~151 (2020)
  16. Huang T, "Meta-analysis of gap balancing versus measured resection techniques in total knee arthroplasty", Bone Joint J, 99-B(2), pp.151~158 (2017)
  17. Harvey RA, "Femoral component rotation in the Attune TKR, balanced technique and measured resection instrumentation", Poster presentation. BASK 2017
  18. "ATTUNE® CR Knee and ATTUNE® PS Knee NJR Implant Summary Report August 2018"[URL]

관련 논문AI추천

Higher rates of fully preserved posterior cruciate ligament in total knee arthroplasty using a double tibial cut: a prospective randomized controlled trialBeatrice Perciballi, Francesca Romana Ripani, Gianluca Cinotti, Giuseppe Giannicola, Giuseppe La Torre대한슬관절학회 · Knee Surgery and Related Research2024 · 제36권 Short-Term Results of Total Knee Arthroplasty with Anterior-Posterior Glide LCS Mobile-Bearing System왕립, 김경택, 박진우, 상민수, 임영훈대한슬관절학회 · Knee Surgery and Related Research2014 · 제26권 제3호Early functional outcomes after condylar-stabilizing (deep-dish) versus standard bearing surface for cruciate-retaining total knee arthroplastyD. MacDonald, G. J. Macpherson, J. T. Patton, N. D. Clement, P. Stirling, R. Burnett대한슬관절학회 · Knee Surgery and Related Research2019 · 제31권 제1호Comparison of Clinical Results and Risk of Patellar Injury between Attune and PFC Sigma Knee Systems송상준, Cheol Hee Park, Dae Kyung Bae, Se Gu Kang대한슬관절학회 · Knee Surgery and Related Research2018 · 제30권 제4호Outcome of Unicompartmental Knee Arthroplasty: A Systematic Review of Comparative Studies between Fixed and Mobile Bearings Focusing on ComplicationsKwang-Jun Oh, Manan Ramesh Gujarathi, Young-Bong Ko대한슬관절학회 · Knee Surgery and Related Research2015 · 제27권 제3호A Comparison of the Clinical and Radiographic Results of Press Fit Condylar Rotating-Platform High-Flexion and Low Contact Stress Mobile Bearing Prosthesis in Total Knee Arthroplasty: Short term Results이범구, 이용석, Ji Hoon Kwak, Nam Ki Kim, Shin Woo Nam대한슬관절학회 · Knee Surgery and Related Research2012 · 제24권 제1호

이 간행물의 다른 논문 (63건)