Purpose: To propose a new infusion rate monitoring technique based on the 2D image marker tacking to improve patient safety by preventing syringe pump-related medication accidents due to decreased infusion rate control accuracy. Materials and Methods: The infusion rate of the syringe pump and drug residue in the pump-equipped syringe were monitored in real time by tracking the movement of the 2D image markers attached to the syringe pump. Results: The error rate between the set and the estimated infusion rates was 1.03, 0.66, 1.95, 0.23, and 1.05% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. In addition, the error rate between the actual and the estimated drug residues was 1.04, 0.47, 0.60, 3.66, and 0.00% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. Conclusion: Experimental results demonstrated that the proposed technique can increase the efficiency of the safety management system for seriously ill inpatients by decreasing a possibility of syringe pump- related medication accidents in hospitals.

Abstract

Purpose: To propose a new infusion rate monitoring technique based on the 2D image marker tacking to improve patient safety by preventing syringe pump-related medication accidents due to decreased infusion rate control accuracy. Materials and Methods: The infusion rate of the syringe pump and drug residue in the pump-equipped syringe were monitored in real time by tracking the movement of the 2D image markers attached to the syringe pump. Results: The error rate between the set and the estimated infusion rates was 1.03, 0.66, 1.95, 0.23, and 1.05% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. In addition, the error rate between the actual and the estimated drug residues was 1.04, 0.47, 0.60, 3.66, and 0.00% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. Conclusion: Experimental results demonstrated that the proposed technique can increase the efficiency of the safety management system for seriously ill inpatients by decreasing a possibility of syringe pump- related medication accidents in hospitals.

참고문헌 (16)

  1. Brocx S, "When and how to use a syringe driver in palliative care", Best Practice J, 48, pp.28~35 (2012)
  2. Kim GH, "A new injection rate estimation technique for on-site screening test of medication infusion pump by nurses", P I Mech Eng H, 234(4), pp.370~376 (2020)
  3. Batliner M, "Evaluation of a novel flow-controlled infusion pump for precise and continuous drug delivery at low flow rates : a laboratory study", Anaesthesia, 74(11), pp.1425~1431 (2019)
  4. "Health insurance treatment statistics for inpatients in intensive care units" (2022)[URL]
  5. Lin L, "Patient safety, potential adverse drug events, and medical device design : a human factors engineering approach", J Biomed Inform, 34(4), pp.274~284 (2001)
  6. Vicente KJ, "Programming errors contribute to death from patient-controlled analgesia : case report and estimate of probability", Can J Anaesth, 50(4), pp.328~332 (2003)
  7. Syed S, "Morphine overdose from error propagation on an acute pain service", Can J Anaesth, 53(6), pp.586~590 (2006)
  8. 박호준, "위험관리기반의 성능관리 의료기기 선정 절차 수립 및 시험 항목 도출", 의공학회지, 40(1), pp.20~31 (2019), 대한의용생체공학회
  9. International Electrotechnical Commission, "IEC 60601-2-24:Medical electrical equipment-Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers", 17~34 (2012)
  10. Batliner M, "Evaluation of a novel flow-controlled syringe infusion pump for precise and continuous drug delivery at low flow rates : a laboratory study", Anaesthesia, 74(11), pp.1425~1431 (2019)
  11. AnJani AG, "Application of IoT using nodeMCU ESP8266 on the Syringe Pump Device to Increase Patient Safety", Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics, 4(1), pp.23~27 (2022)
  12. Das S, "A Pencil Drawn Capacitive Sensor used for Liquid Drug Volume Measurement in Syringe Pump", 2020IEEE Applied Signal Processing Conference (ASPCON), pp.85~88 (2020)
  13. Hagihghi R, "A miniaturized piezoresistive flow sensor for real-time monitoring of intravenous infusion", Journal of Biomedical Materials Research Part B : Applied Biomaterials, 108(2), pp.568~576 (2020)
  14. Shen Z, "Biomimetic flow sensors for biomedical flow sensing in intravenous tubes", 2016 IEEE SENSORS, 1~3 (2016)
  15. Islam MR, "Design and Implementation of Low Cost Smart Syringe Pump for Telemedicine and Healthcare", 2019 International Conference on Robotics, Electrical and Signal Processing Techniques (ICREST), pp.440~444 (2019)
  16. Rajasekwaran S, "Iot based low cost syringe pump for telemedicine and health care", International Research Journal of Modernization in Engineering Technology and Science, 4(6), pp.4222~4229 (2022)

관련 논문AI추천

이 간행물의 다른 논문 (11건)