본문 바로가기
KERD
논문KCI 등재

차트 각도를 이용한 해상력 특성 평가

Evaluation of Resolution Characteristics by Using Chart Device Angle

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study aim was quantitative assessment of MTFs of spectrum of the square wave chart images and Coltman chart images for 0o, 1.7o, 2.2o, 2.9o, 4.1o by using chart method. In general device was AccuRay-650 (DK Medical System, Korea) used, indirect flat panel detector(FPD) Aero (Konica, Japan) used and MATLAB R2019a (MathWorks, USA) used. The result of comparison for each angle of MTF the edge image was highest quantitatively value for MTF finding of showed the best value of 0.1 based on the frequency of 3.5 mm-1, value of 0.1 based on the square wave was frequency of 3.0 mm-1 and value of 0.1 based on the Coltman transform was frequency of 2.4 mm-1. In this study it was significant that the methodology of the international Electro-technical Commission was applied mutandis by using the Fujita method within 2∼3o.

Abstract

This study aim was quantitative assessment of MTFs of spectrum of the square wave chart images and Coltman chart images for 0o, 1.7o, 2.2o, 2.9o, 4.1o by using chart method. In general device was AccuRay-650 (DK Medical System, Korea) used, indirect flat panel detector(FPD) Aero (Konica, Japan) used and MATLAB R2019a (MathWorks, USA) used. The result of comparison for each angle of MTF the edge image was highest quantitatively value for MTF finding of showed the best value of 0.1 based on the frequency of 3.5 mm-1, value of 0.1 based on the square wave was frequency of 3.0 mm-1 and value of 0.1 based on the Coltman transform was frequency of 2.4 mm-1. In this study it was significant that the methodology of the international Electro-technical Commission was applied mutandis by using the Fujita method within 2∼3o.

참고문헌 (18)

  1. 민정환, "3.0T 자기공명영상을 이용한 해상력 특성에 대한 연구", 방사선기술과학, 43(4), pp.251~257 (2020), 대한방사선과학회
  2. 정회원, "CsI(TI) Indirect Flat Panel Detector의 선질에 따른 물리적 영상 평가", 방사선기술과학, 35(2), pp.109~117 (2012), 대한방사선과학회[DOI]
  3. Miyati T, "Measurements of MTF and SNR(f)using a subtraction method im MRI", Physics in Medicine and Biology, 47, pp.2961~2972 (2002)
  4. 손순룡, "Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy", 방사선기술과학, 38(4), pp.451~461 (2015), 대한방사선과학회[DOI]
  5. 손순룡, "Measurement of Image Quality According to the Time of Computed Radiography System", 방사선기술과학, 38(4), pp.365~374 (2015), 대한방사선과학회[DOI]
  6. 김기원, "유방영상에서 엣지를 이용한 물리적 영상 평가", 방사선기술과학, 39(3), pp.415~420 (2016), 대한방사선과학회[DOI]
  7. 김기원, "컴퓨터단층검사에서 AAPM Phantom을 이용한 변조전달함수 평가", 방사선기술과학, 39(2), pp.193~198 (2016), 대한방사선과학회[DOI]
  8. Sawant A, "Slit design for efficient and accurate MTF measurement at megavoltage X-ray energies", Medical Physics, 34, pp.1535~1545 (2007)
  9. Kim JM, "Correction method of slit modulation transfer function on digital medical imaging system", Journal of Radiological Science and Technology, 29(3), pp.133~139 (2006)
  10. 민정환, "간접평판형 검출기에서 변조전달함수 측정 시 Edge 각도에 따른 비교 연구", 방사선기술과학, 42(4), pp.259~263 (2019), 대한방사선과학회
  11. 민정환, "변조전달함수 방법에서 엣지 장치 설정에 대한 각도 최적화 평가", 방사선기술과학, 43(1), pp.15~21 (2020), 대한방사선과학회
  12. Ichikawa K, "New method of measuring presampling MTF by fourier transform of square wave chart image", Japanese Journal of Radiological Technology, 58(9), pp.1261~1267 (2002)
  13. Kunitomo H, "Modulation transfer function-presampled MTF measurements by fourier analysis using square wave chart", Japanese Journal of Radiological Technology, 65(11), pp.1561~1567 (2009)
  14. Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Medical Physics, 27, pp.2048~ (2000)
  15. Fujita H, "A simple method for determining the modulation transfer function in digital radiography", IEEE Trans Med Imaging, 11(1), pp.34~39 (1992)
  16. Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Medical Physics, 25, pp.102~ (1998)
  17. "IEC (International Electro-technical Commission)62220-1. Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency", ~ (2003)
  18. 민정환, "3.0T 자기공명영상을 이용한 잡음전력스펙트럼 특성 평가", 방사선기술과학, 44(1), pp.31~37 (2021), 대한방사선과학회

관련 논문AI추천