초록
Objective To assess the usefulness of a pressure algometer to measure pressure pain threshold (PPT) for diagnosis of myofascial pain syndrome (MPS) in the upper extremity and trunk muscles. Method A group of 221 desk workers complaining of upper body pain participated in this study. Five physiatrists made the diagnosis of MPS using physical examination and PPT measurements. PPT measurements were determined for several muscles in the back and upper extremities. Mean PPT data for gender, side, and dominant hand groups were analyzed. Sensitivity and specifi city of Fischer’s standard method were evaluated. PPT cut-off values for each muscle group were determined using an ROC curve. Results Cronbach’s alpha for each muscle was very high. The PPT in men was higher than in females, and the PPT in the left side was higher than in the right side for all muscles tested (p<0.05). Th ere was no signifi cant diff erence in PPT for all muscles between dominant and non-dominant hand groups. Diagnosis of MPS based on Fischer’s standard showed relatively high specifi city and poor sensitivity. Conclusion The digital pressure algometer showed high reliability. PPT might be a useful parameter for assessing a treatment’s effect, but not for use in diagnosis or even as a screening method.
Abstract
Objective To assess the usefulness of a pressure algometer to measure pressure pain threshold (PPT) for diagnosis of myofascial pain syndrome (MPS) in the upper extremity and trunk muscles. Method A group of 221 desk workers complaining of upper body pain participated in this study. Five physiatrists made the diagnosis of MPS using physical examination and PPT measurements. PPT measurements were determined for several muscles in the back and upper extremities. Mean PPT data for gender, side, and dominant hand groups were analyzed. Sensitivity and specifi city of Fischer’s standard method were evaluated. PPT cut-off values for each muscle group were determined using an ROC curve. Results Cronbach’s alpha for each muscle was very high. The PPT in men was higher than in females, and the PPT in the left side was higher than in the right side for all muscles tested (p<0.05). Th ere was no signifi cant diff erence in PPT for all muscles between dominant and non-dominant hand groups. Diagnosis of MPS based on Fischer’s standard showed relatively high specifi city and poor sensitivity. Conclusion The digital pressure algometer showed high reliability. PPT might be a useful parameter for assessing a treatment’s effect, but not for use in diagnosis or even as a screening method.
참고문헌 (21)
- Ohara H, "Occupational health hazards resulting from elevated work rate situations", J Human Ergol, 5, pp.173~182 (1976)
- Onishi N, "Shoulder muscle tenderness and physical features of female industrial workers", J Human Ergol, 5, pp.87~102 (1976)
- Bechtol CO, "Grip test; the use of a dynamometer with adjustable handle spacing", J Bone Joint Surg Am, 36, pp.820~824 (1954)
- Sola AE, "Myofascial pain syndromes. The management of pain", pp.352~ (1990), Lea & Febiger
- Hogeweg JA, "Algometry. Measuring pain threshold, method and characteristics in healthy subject", Scand J Rehabil Med, 24, pp.99~103 (1992)
- Fischer AA, "Pressure threshold meter: its use for quantifi cation of tender spots", Arch Phys Med Rehabil, 67, pp.836~838 (1986)
- Han TR, "Pain threshold using a pressure algometer in normal Korean young adults", J Korean Acad Rehabil Med, 18, pp.219~226 (1994)
- Lee GH, "Pressure threshold, grip and pinch strength in female telephone operators", J Korean Acad Rehab Med, 21, pp.589~593 (1997)
- Buchanan HM, "Evaluation of pain threshold using a simple pressure algometer", Clinical Rheumatology, 6, pp.510~517 (1987)
- Chesterton LS, "Gender diff erences in pressure pain threshold in healthy humans", Pain, 101, pp.259~266 (2003)
- Chung SC, "Evaluation of pressure pain threshold in head and neck muscles by electronic algometer: intrarater and interrater reliability", Cranio, 10, pp.28~34 (1992)
- Chesterton LS, "Interrater reliability of algometry in measuring pressure pain thresholds in healthy humans, using multiple raters", Clin J Pain, 23, pp.760~766 (2007)
- Dagtekin O, "Measuring pressure pain thresholds. Comparison of an electromechanically controlled algometer with established methods", Schmerz, 21, pp.439~444 (2007)
- Kinser AM, "Reliability and validity of a pressure algometer", J Strength Cond Res, 23, pp.312~314 (2009)
- Simons DG, "Myofascial pain and dys function: the trigger point mannual", pp.19~44 (1999), Williams & Wilkins
- 이중훈, "전자식 압력 통각계의 유용성", Annals of Rehabilitation Medicine, 32(6), pp.698~702 (2008), 대한재활의학회
- Kwon YE, "Pressure pain threshold measurement using a pressure algometer in myofascial pain syndromes", J Korean Assoc Pain Med, 14, pp.32~36 (2001)
- Reeves JL, "Graff-Radford SB. Reliability of the pressure algometer as a measure of myofascial trigger point sensitivity", Pain, 24, pp.313~321 (1986)
- Delancy GA, "Inter- and intra-rater reliability of the pressure threshold meter in measurement of myofascial trigger point sensitivity", Am J Med Rehabil, 72, pp.136~139 (1993)
- Vatine JJ, "Electronic pressure algometry of deep pain in healthy volunteers", Arch Phys Med Rehabil, 74, pp.526~530 (1993)
- Fischer AA, "Pressure algometry over normal muscles. Standard value, validity and reproducibility of pressure threshold", Pain, 30, pp.115~126 (1987)