초음파 영상용 플라스틱 기반의 팬텀제작 및 특성 분석

Analysis of Properties and Phantom Design Based on Plastic Hardener and Softener for Ultrasonic Imaging

0

조회

0

다운로드

6

인용

0

참고문헌

키워드

초록

lastic hardener and softener based ultrasound phantoms were made in various constitutions and their acoustic properties were measured. Speed of sound is approximately 1.4 mm/μsec in all the phantoms, which is about 7% less than that of in soft tissue. Attenuation coefficient is strongly dependent on the ratio between hardener and softener. In order to achieve the tissue level attenuation (0.5 dB/cm/MHz), 60% of hardener or less is required. The synthesized phantoms can be preserved for more than 6 months without structural degradation.

Abstract

lastic hardener and softener based ultrasound phantoms were made in various constitutions and their acoustic properties were measured. Speed of sound is approximately 1.4 mm/μsec in all the phantoms, which is about 7% less than that of in soft tissue. Attenuation coefficient is strongly dependent on the ratio between hardener and softener. In order to achieve the tissue level attenuation (0.5 dB/cm/MHz), 60% of hardener or less is required. The synthesized phantoms can be preserved for more than 6 months without structural degradation.

참고문헌 (8)

  1. E.L. Madsen, "Tissue mimicking materials for ultrasound phantoms", Medical Physics, 5(5), pp.391~394 (1978)
  2. M.M. Burlew, "A new ultrasound tissue-equivalent Material", Radiation Physics, 134, pp.517~520 (1980)
  3. T.L.Szabo, "Diagnostic Ultrasound Imaging" (2004)
  4. K . Takegami, "Polyacrylamide gel containing egg white as new model for irradiation experiments using focused ultrasound", Ultrasound in Medicine & Biology, 30(10), pp.1419~1422
  5. "Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment", AIUM, ~ (2004)
  6. R.C. Preston, "Calibration of Medical Ultrasound Equipment-Procedures and Accuracy Assessement", IEEE transactions on UFFC, 35(2), pp.110~121 (1988)
  7. R.F. Wagner, "Statistics of Speckle in Ultrasound B-Scans", IEEE transactions on sonics and ultrasonics, 30(3), pp.156~163 (1983)
  8. E.L. Madsen, "Interlaboratory Comparison of ultrasonic backscatter, attenuation, and speed measurements", J Ultrasound Med, 18, pp.615~631 (1999)