Using a spatial compound imaging technique in a medical ultrasound imaging system, the average speed of sound in a medium of interest is measured, and imaging of its distribution is implemented. When the brightness reaches the highest level in an ultrasonic image obtained as the speed of sound used in focusing is varied, it turns out that the focusing has been accomplished satisfactorily and that the speed of sound which has been adopted becomes the sought-after average speed of sound. Because spatial compound imaging provides many different views of the same object, the adverse effect of erroneous speed-of-sound estimation tends to be more severe in compound imaging than in plain B-mode imaging. Thus, in compound imaging, the average speed of sound even in the case of speckled images can be accurately estimated by observing the brightness change due to different speeds of sound employed. Using this new method that offers spatial diversity, we can construct an image of the speed of sound distribution in a phantom embedded with a 10-mm diameter plastic cylinder whose speed of sound is different from that of the background. The speed of sound in the cylinder is found to be different from that of the surrounding medium

Abstract

Using a spatial compound imaging technique in a medical ultrasound imaging system, the average speed of sound in a medium of interest is measured, and imaging of its distribution is implemented. When the brightness reaches the highest level in an ultrasonic image obtained as the speed of sound used in focusing is varied, it turns out that the focusing has been accomplished satisfactorily and that the speed of sound which has been adopted becomes the sought-after average speed of sound. Because spatial compound imaging provides many different views of the same object, the adverse effect of erroneous speed-of-sound estimation tends to be more severe in compound imaging than in plain B-mode imaging. Thus, in compound imaging, the average speed of sound even in the case of speckled images can be accurately estimated by observing the brightness change due to different speeds of sound employed. Using this new method that offers spatial diversity, we can construct an image of the speed of sound distribution in a phantom embedded with a 10-mm diameter plastic cylinder whose speed of sound is different from that of the background. The speed of sound in the cylinder is found to be different from that of the surrounding medium

참고문헌 (15)

  1. A. Goss, "Comprehensive compilation of empirical ultrasonic properties of mammalian tissues", J. Acoust. Soc. Am, 64(2), pp.423~457 (1978)
  2. H. Hiroyuki, "Relationship between speed of sound in and density of normal and diseased rat livers", Jpn. J. Appl. Phys, 33(5B), pp.3130~3133 (1994)
  3. C. H. Chang, "Reconstruction of ultrasonic sound velocity and attenuation coefficient using linear arrays: Clinical assessment", Ultrasound Med. Biol, 33(11), pp.1681~1687 (2007)
  4. J. H. Kim, "A pipelined sampled delay focusing in ultrasonic imaging systems", Ultrason. Imaging, 9, pp.75~91 (1987)
  5. K. Jeon, "An efficient real time focusing delay calculation in ultrasonic imaging systems", Ultrason. Imaging, 16(4), pp.231~248 (1994)
  6. T. Kontonassios, "Variance reduction of speed of sound estimation in tissues using the beam tracking method", IEEE Trans. Ultrason. Ferroelect. Freq. Contr, 34(5), pp.624~530 (1987)
  7. J. Ophir, "Optimization of speed-of-sound estimation from noisy ultrasonic signals", IEEE Trans. Ultrason. Ferroelect. Freq. Contr, 36(1), pp.16~24 (1989)
  8. D. Napolitano, "Sound speed correction in ultrasound imaging", Ultrasonics, 44(1), pp.e43~e46 (2006)
  9. M. K. Jeong, "An efficient method for estimating average speed of sound in ultrasound medical imaging", J. Biomed. Eng. Res, 19(3), pp.321~325 (1998)
  10. J. H. Kim, "Study on velocity in-homogeneous effect in fat and its correction in ultrasound imaging system", J. Biomed. Eng. Res, 19(1), pp.9~17 (1998)
  11. M. E. Anderson, "The impact of sound speed errors on medical ultrasound imaging", J. Acoust. Soc. Am, 107(6), pp.3540~3548 (2000)
  12. D. Zaho, "Comparison of image quality factors for phase aberration correction with diffuse and point targets: Theory and experiments", IEEE Trans. Ultrason. Ferroelect. Freq. Contr, 38(2), pp.125~132 (1991)
  13. J. E. Wilhjelm, "Some imaging strategies in multi- angle spatial compounding", Proc. IEEE Ultrason. Symp, pp.1615~1618 (2000)
  14. U. Bencharit, "Frequency and spatial compounding techniques for improved ultrasonic imaging", Proc. IEEE Ultrason. Symp, pp.1021~1026 (1986)
  15. B. K. Seo, "Sonographic evaluation of breast nodules: Comparison of conventional, real-time compound, and pulse-inversion harmonic images", Korean J. Radiol, 3(1), pp.38~44 (2002)